Skip to content

Bard® InLay® Ureteral Stent 778424Recalls & reported problems

C. R. Bard, Inc. · INLAY URETERAL STENT w/o GUIDEWIRE · Class 2 · Implant

All models of Bard InLay Ureteral Stent (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741014758), received between Jun 29, 2017 and Aug 6, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2
  • Break1
  • Calcified1
  • Detachment Of Device Component1
  • Migration1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • Foreign Body In Patient1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2025Calcified; Calcified

    IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT URETEROSCOPIC LITHOTRIPSY FOR KIDNEY STONES AND NEEDED A STENT REMOVAL. THE STENT HAD BEEN IN PLACE FOR 20 DAYS. DURING THE PROCEDURE TO REMOVE THE URETERAL STENT, SEVERE ENCRUSTATION OF THE STENT WAS OBSERVED, MAKING IT DIFFICULT TO REMOVE. AFTER CLINICAL EVALUATION, SURGICAL INTERVENTION WAS PERFORMED TO REMOVE THE STENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED, AND THE STENT WAS REMOVED. INCREASING THE DURATION AND COST OF THE PROCEDURE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22726537 ↗
  • MalfunctionApr 9, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE ON (B)(6) 2025, THE PATIENT UNDERWENT URETEROSCOPY LITHOTRIPSY FOR KIDNEY STONES. WHEN THE PROCEDURE ENTERED THE CATHETERIZATION STAGE, THE URETERAL STENT WAS INSERTED. AFTER THE STENT WAS OPENED, A SMALL GAP WAS FOUND AT THE HEAD OF THE PIGTAIL OF THE STENT WHEN THE STENT WAS INSERTED (ACTUAL SAMPLES AND PHOTOS CAN BE SEEN). AFTER CLINICAL JUDGMENT, A NEW PACKAGE WAS CHOSEN. THE PRODUCT COMPLETED THE SUBSEQUENT OPERATIONS SMOOTHLY WITHOUT ANY PROBLEMS, AND THE OPERATION WAS COMPLETED SMOOTHLY. THE OPERATION WAS DELAYED, AND THERE WERE NO OTHER IMPACTS. PER FOLLOW UP INFORMATION RECEIVED VIA IBC ON 03APR2025, STATED THAT THE PATIENT DID NOT USE THE PRODUCT, AND THE QUALITY ISSUE WITH THE STENT WAS DISCOVERED BEFORE THE STENT WAS INSERTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21805056 ↗
  • MalfunctionDec 28, 2021Migration; Migration

    IT WAS REPORTED THAT THE STENT WAS FOUND IN THE ABDOMEN OUTSIDE THE URETER. THERE WAS A PIGTAIL CATHETER ADJACENT TO THE URETER. THIS WAS EXTERNAL TO THE URETER AND NOT IN BOWEL. CLINICAL CORRELATION WAS RECOMMENDED AND THIS HAS BEEN PLACED IN THE INTERVAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13102512 ↗
  • MalfunctionApr 15, 2020Material Fragmentation

    IT WAS REPORTED THAT UPON INSERTING URETERAL STENT PUSHER INTO NIPPLE ON 22F SHEATH, ORANGE SHAVINGS FROM PUSHER WAS OBSERVED AND REMOVED FROM BLADDER. IT WAS LATER REPORTED FROM THE SALES REP. VIA EMAIL ON 31MAR2020, THE DOCTOR LEFT THE SHEATH IN AND LET THE PIECES/SHAVINGS DRAIN OUT.

    Patient problems: Foreign Body In Patient; No Known Impact Or Consequence To Patient; Patient Problem/Medical Problem

    MAUDE report 9962308 ↗
  • MalfunctionJun 29, 2017Break; Detachment Of Device Component

    IT WAS REPORTED THAT PRIOR TO USE, THE SUTURE LOOP BROKE AWAY FROM THE DEVICE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6676615 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →

Sources and updates