Skip to content

Bard® InLay Optima® Ureteral Stent 788414Recalls & reported problems

C. R. Bard, Inc. · INLAY OPTIMA URETERAL STENT W/O GUIDEWIRE · Class 2 · Implant

All models of Bard InLay Optima Ureteral Stent (21)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741015649), received between Jun 2, 2021 and Jan 23, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation6
  • Tear, Rip or Hole in Device Packaging2
  • Device Contamination with Chemical or Other Material1
  • Difficult to Advance1
  • Flaked1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Abrasion1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Material Fragmentation

    IT WAS REPORTED THAT, DURING URETERAL STENT PLACEMENT, THE URETERAL STENT WAS OPENED AND THE BLADDER LOOP SUTURE REMOVED, REVEALING A FRACTURE AT THE SMALL HOLE OF THE BLADDER LOOP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24173012 ↗
  • MalfunctionNov 6, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT, AFTER STONE SURGERY, UPON OPENING THE URETERAL STENT, FOREIGN MATTER WAS DISCOVERED IN THE SAC. REPLACED THE STENT WITH A NEW ONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23485359 ↗
  • MalfunctionSep 25, 2025Flaked; Patient Device Interaction Problem; Flaked; Patient Device Interaction Problem

    IT WAS REPORTED THAT, THE URETERAL STENTS EXHIBIT A HIGH RATE OF SCABBING WHEN USED ON PATIENTS FOR APPROXIMATELY TWO WEEKS. WITHIN ONE MONTH, SCABBING HAD OCCURRED SEVEN OR EIGHT TIMES.

    Patient problems: Abrasion; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23149382 ↗
  • MalfunctionSep 2, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT, WHEN USING THE URETERAL STENT, THE KIDNEY TRANSPLANT RECIPIENT REMOVED IT FROM THE PACKAGING, FIRST SNAPPED THE SUTURE THREAD, THEN STRAIGHTENED THE STENT, AND APPLIED A SIMPLE LATERAL PULL, CAUSING IT TO BREAK. AFTER REPLACING THE STENT, THE ISSUE DID NOT RECUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22937280 ↗
  • MalfunctionAug 28, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT WHEN THE KIDNEY TRANSPLANT PATIENT USES THE URETERAL STENT, THEY TAKE OUT THE PACKAGING BOX AND FIRST BREAK THE SILK THREAD, THEN STRAIGHTEN THE STENT AND GIVE IT A SIMPLE HORIZONTAL PULL, CAUSING IT TO BREAK. AND REPLACED THE STENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22913981 ↗
  • MalfunctionJul 1, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT WHEN OPENED THE OUTER PACKAGING OF THE URETERAL STENT, IT WAS FOUND THAT THE INNER PACKAGING WAS DAMAGED, SO OPENED A NEW URETERAL STENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22374348 ↗
  • MalfunctionApr 3, 2025Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT THE STENT WAS UNABLE TO ADVANCE ON THE GUIDE WIRE. IT MAY BE A PROBLEM WITH AN EXCESSIVE RESISTANCE DURING STENT DELIVERY. PER ADDITIONAL INFORMATION RECEIVED VIA IBC ON 19MAR2025, STATED THAT THEY TRIED ABOUT 3 TIMES TO ADVANCE THE STENT OVER THE GUIDEWIRE, BUT THEY FAILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21762760 ↗
  • MalfunctionAug 30, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING THE STONE SURGERY, THE PIGTAIL RING AT THE HEAD OF THE URETERAL STENT WAS FOUND TO BE BROKEN AFTER THE STENT PACKAGING WAS OPENED, SO IT WAS STOPPED FROM USE. REPLACED WITH A NEW STENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20108373 ↗
  • MalfunctionSep 12, 2023Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE SEAL WAS FOUND DAMAGED UPON OPENING OF THE URETERAL STENT OUTER PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17728365 ↗
  • MalfunctionNov 9, 2021Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE DRAINAGE HOLE AT THE TIP OF THE URETERAL STENT WAS FOUND BROKEN AFTER THE PACKAGE WAS OPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12781011 ↗
  • MalfunctionJun 2, 2021Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE STENT RIPPED DURING THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11921926 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page