Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741017612), received between Oct 2, 2018 and Aug 18, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Puncture/Hole” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Puncture/Hole2
  • Device Dislodged or Dislocated1
  • Failure to Infuse1
  • Fluid/Blood Leak1
  • Inflation Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2
  • Cramp(s) /Muscle Spasm(s)1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50311244 ↗
  • MalfunctionMar 18, 2026Material Puncture/Hole

    IT WAS REPORTED THAT THE CATHETER FELL OUT OF THE PATIENT SHORTLY AFTER PLACEMENT, AND A HOLE IN THE BALLOON WAS IDENTIFIED, PREVENTING INFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24630889 ↗
  • MalfunctionAug 8, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE FOLEYS ARE NOT STAYING IN THE PATIENT AND THE NURSING TEAM IS QUESTIONING THE INTEGRITY OF THE PRODUCT. PER THE CUSTOMER COMPLAINT, THEY DID NOT HAVE ALL THE DETAILS BUT THE PICTURES OF FOLEYS THAT THE NURSING TEAM HAD ISSUES WITH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22749791 ↗
  • MalfunctionMar 4, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE HOME NURSE HAD A PATIENT WHO HAD BEEN USING LUBRICATH FOLEY TRAY, 0166L22, FOR 4-5 YEARS. THEY PLACED A NEW FOLEY AND PATIENT WAS LEAKING AND HAVING BLADDER SPASMS AND ASKED IF THERE WAS ANY CHANGE THE LATEX USED IN THE FOLEYS. REPRESENTATIVE DISCUSSED INFLATION VOLUMES AND NURSE MENTIONED THEY WERE ONLY INFLATING BALLOON WITH 30CC, EXPLAINED OVER/UNDER INFLATION AND HOW THE BALLOON SHOULD BE INFLATED WITH 35CC, AND SUGGESTED THE PATIENT TO REACH OUT TO UROLOGIST TO SEE IF THERE WAS ANYTHING ELSE GOING ON AND WHAT THEY SHOULD DO NEXT.

    Patient problems: Cramp(s) /Muscle Spasm(s)

    MAUDE report 18825811 ↗
  • MalfunctionMay 7, 2019Inflation Problem; Failure to Infuse; Failure to Infuse

    IT WAS REPORTED THAT WHILE ATTEMPTING TO PLACE THE CATHETER INTO THE PATIENT'S CERVIX, UPON INFLATION OF THE BALLOON, THE CATHETER ATTACHED TO THE FLUID PORT AND WAS FILLING WITH FLUID. THE FLUID WAS NOT GOING INTO THE BALLOON. THE CATHETER WAS REMOVED AND A NEW ONE WAS PLACED. IT WAS LATER REPORTED PER ADDITIONAL INFORMATION FROM THE COMPLAINANT VIA EMAIL ON (B)(6)2019: THE FLUID WAS GOING INTO THE SHAFT OF THE CATHETER INSTEAD OF THE BALLOON OF THE CATHETER.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8589203 ↗
  • MalfunctionOct 2, 2018Leak/Splash

    IT WAS REPORTED THAT THE BALLOON OF THE CATHETER ONLY INFLATED TO ONE SIDE, CAUSING LEAKAGE DURING USE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7927786 ↗