Bardex® Lubricath® Foley Catheter Councill Model 0196L16Recalls & reported problems
C. R. Bard, Inc. · Bardex® Lubricath® Councill Model Foley Catheter Red Latex Radiopaque 5cc Balloon Lubricious Coated · Class 2
All models of Bardex Lubricath Foley Catheter Councill Model (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00801741017926), received between Dec 4, 2017 and Sep 10, 2026 — classified as 3 injuries, 20 malfunctions.
The most frequently reported specific device problem is “Material Puncture/Hole” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole8
- Difficult to Remove3
- Deflation Problem2
- Failure to Infuse2
- Material Rupture2
- Appropriate Device Problem Term/Code Not Available1
- Burst Container or Vessel1
- Component Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Discomfort1
- No Information1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Material Puncture/Hole
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50722283 ↗MalfunctionSep 4, 2026Material Puncture/Hole
It was reported that Customer had experienced with at least three of our Bard Council Tip catheters. Instead of the balloon inflating at the tip of the catheter within the patient's bladder, it inflated at the fork where the syringe connection and foley bag attachment sites were located. For today¿s issue, the REF# is (b)(4) and the LOT# is NGLR3780. Customer also had issues where there was a leak at the Y junction and the fluid in the balloon port leaked out. They had two patient's council catheters fell out after surgery.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50620017 ↗MalfunctionSep 4, 2026Material Puncture/Hole
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50612422 ↗InjuryMar 4, 2026Material Puncture/Hole; Device Dislodged or Dislocated
IT WAS REPORTED THAT WHEN THE WRITER WENT IN TO ASSESS THE PATIENT, NO DRAINAGE WAS NOTED IN THE FOLEY BAG. AT THE START OF THE SHIFT, A SMALL AMOUNT OF DRAINAGE HAD BEEN OBSERVED IN THE TUBING. THE WRITER ASKED THE PATIENT IF THE BAG HAD BEEN EMPTIED, AND THE PATIENT STATED THAT IT HAD NOT. THE FOLEY TUBING HAD BEEN ROUTED THROUGH THE PATIENT¿S PANTS, MAKING IT IMPOSSIBLE TO SEE WHETHER IT WAS SECURED IN THE LEG STRAP. A BLADDER SCAN SHOWED 800 ML. THE PATIENT REPORTED BEING COMFORTABLE AND DID NOT FEEL AN URGE TO VOID. WHEN THE PATIENT REMOVED THEIR PANTS, THE FOLEY CATHETER WAS FOUND TO BE OUT OF PLACE WITH THE BALLOON DEFLATED. NO TRAUMA WAS NOTED TO THE PENIS, AND THE PATIENT STATED THEY DID NOT FEEL ANY PAIN OR A ¿TUG¿ AND WERE UNSURE HOW IT HAPPENED. THE FOLEY CATHETER WAS PLACED IN A SPECIMEN BAG AND PROVIDED TO THE CNE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24506147 ↗MalfunctionJan 30, 2026Difficult to Remove
IT WAS REPORTED THAT THE CUSTOMER WAS UNABLE TO DEFLATE FOLEY CATHETER BALLOON TO REMOVE FROM PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24223863 ↗MalfunctionDec 22, 2025Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT FOLEY CATHETER WOULD NOT DEFLATE FOR REMOVAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23871904 ↗MalfunctionApr 15, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE PATIENT THAT THE FOLEY CATHETERS ARE LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21839163 ↗MalfunctionOct 24, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT 60-YEAR-OLD MALE WITH HISTORY OF VRE (VANCOMYCIN-RESISTANT ENTEROCOCCI), SACRAL ULCER, LEFT FOOD ULCER, CAD (CORONARY ARTERY DISEASE), CKD (CHRONIC KIDNEY DISEASE), UNDERWENT SUPRAPUBIC FOLEY CATHETER PLACEMENT. WHILE PLACING CATHETER IN BLADDER VIA 20FR PEEL AWAY. FILLED BALLOON WITH STERILE WATER AND CONTRAST 50:50. UPON REMOVING PEEL AWAY, RECHECKED IMAGE, THE BALLOON LEAKED AND DEFLATED. HAD TO GET NEW PEEL AWAY AND COUNCIL-TIP CATHETER AND REPLACE THE PREVIOUS. NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20523816 ↗MalfunctionAug 27, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE FOLEY CATHETER IN THE KIT HAD A SMALL HOLE AND WAS UNABLE TO FILL THE BALLOON. THE WATER KEPT COMING OUT AND THE CATHETER FELL OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20075269 ↗MalfunctionJul 18, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT CUSTOMER WAS NOT CLEAR AND DID NOT KNOW IF IT WAS A NON-DEFLATOR OR IF IT LEAKED BUT NURSES BELIEVED THERE WAS A HOLE IN THE FOLEY CATHETER BALLOON. IT WAS SAID THAT THE CATHETER FELL OUT OF THE PATIENT, IT WAS NOT CLEAR WHAT EXACTLY HAPPENED, BUT THE NURSE BELIEVED THERE TO BE A HOLE IN THE BALLOON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19772459 ↗MalfunctionApr 11, 2024Burst Container or Vessel; Material Rupture; Burst Container or Vessel; Material Rupture
IT WAS REPORTED THAT THE FOLEY CATHETER BALLOON POPPED WHILE IN PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19087279 ↗MalfunctionDec 22, 2023Fracture; Fracture
IT WAS REPORTED THAT WATER LEAKED FROM THE FUNNEL OF THE FOLEY CATHETER WHEN STERILE DISTILLED WATER WAS INJECTED THROUGH THE VALVE USING A SYRINGE. IT WAS NOTED THAT THE THEY FOUND A CRACK IN THE FUNNEL.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18382948 ↗MalfunctionFeb 8, 2023Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE FOLEY CATHETER BALLOON FAILED DUE TO LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16317809 ↗MalfunctionAug 9, 2022Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THE CUSTOMER HAD AN ISSUE WITH THE FOLEY CATHETER. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15190636 ↗InjuryMay 24, 2022Deflation Problem; Deflation Problem
IT WAS REPORTED THAT THE SUPRAPUBIC TUBE BALLOON WOULD NOT BE DEFLATED WHILE PHYSICIAN ATTEMPTED SUPRAPUBIC TUBE EXCHANGE AND IT WAS UNSUCCESSFUL. NURSE ATTEMPTED FLEXIBLE CYSTOSCOPY WITH MULTIPLE GRASPERS AND FOLEY CATHETER BALLOON WAS UNABLE TO PUNCTURE AND WAS ALSO UNABLE TO PLACE SPINAL NEEDLE TO PUNCTURE THE FOLEY BALLOON. STATED THAT THE PATIENT COULD NOT TOLERATE ANY MORE MANIPULATION AND WAS SENT TO THE OPERATING ROOM FOR ENDOSCOPIC BALLOON DEFLATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14488734 ↗MalfunctionMay 4, 2022Failure to Deflate
PHYSICIAN ATTEMPTED SUPRAPUBIC TUBE EXCHANGE. EXCHANGE WAS UNSUCCESSFUL. THE SUPRAPUBIC TUBE BALLOON WOULD NOT DEFLATE. ATTEMPTED FLEXIBLE CYSTOSCOPY WITH MULTIPLE GRASPERS AND KUMAR CATHETER UNABLE TO PUNCTURE THE BALLOON. UNABLE TO PLACE SPINAL NEEDLE TO PUNCTURE BALLOON. PATIENT COULD NOT TOLERATE ANYMORE MANIPULATION AND WAS SENT TO THE O.R. FOR ENDOSCOPIC BALLOON DEFLATION.
Patient problems: Discomfort
MAUDE report 14282264 ↗MalfunctionJan 17, 2022Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THAT THE WATER LEAKED OUT LITTLE BY LITTLE FROM THE FOLEY CATHETER EVEN AFTER INJECTING THE STERILIZED WATER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13271562 ↗InjuryJul 2, 2020Material Rupture
IT WAS REPORTED THAT THE CATHETER BALLOON SPONTANEOUSLY RUPTURED INSIDE THE PATIENT. THE PATIENT WAS HOSPITALIZED. PER ADDITIONAL INFORMATION RECEIVED 18JUN2020, THE CATHETER WAS PLACED IN A PATIENT THAT WAS ALREADY HOSPITALIZED. THE RUPTURED CATHETER WAS REMOVED FROM THE PATIENT AND ANOTHER CATHETER INSERTED PER UROLOGY. THE PATIENT LATER HAD A CYSTOSCOPY TO IDENTIFY AND REMOVE THE RETAINED FRAGMENT FROM THE INITIAL CATHETER BALLOON BURST. ONE PIECE WAS REMOVED.
Patient problems: No Known Impact Or Consequence To Patient; Patient Problem/Medical Problem
MAUDE report 10223760 ↗MalfunctionJun 19, 2018Inflation Problem; Failure to Infuse; Failure to Infuse
I WAS REPORTED THAT THE USER WAS UNABLE TO INFLATE THE BALLOON OF THE CATHETER.
Patient problems: No Consequences Or Impact To Patient; No Information
MAUDE report 7613524 ↗MalfunctionJun 18, 2018Deflation Problem; Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT IN SPITE SEVERING THE VALVE, THE BALLOON COULD NOT BE DEFLATED. THE SHAFT NEAR THE FUNNEL WAS CUTE AND THE BALLOON STILL COULD NOT BE DEFLATED. ON THE SAME DAY, THE SHAFT 4CM DISTAL FROM THE FUNNEL WAS SEVERED, AND WATER WAS ABLE TO BE WITHDRAWN AND THE CATHETER WAS REMOVED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7610301 ↗MalfunctionFeb 23, 2018Misassembled; Material Discolored; Difficult to Insert; Misassembled
IT WAS REPORTED THAT THE DRAINAGE EYE WAS NOT IN ITS STANDARD POSITION AND MADE IT DIFFICULT TO INSERT INTO THE PATIENT. PER THE PHOTO SAMPLE, THE CATHETER HAD BECOME DISCOLORED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7292082 ↗MalfunctionJan 22, 2018Failure to Infuse
IT WAS REPORTED THAT THE BALLOON OF THE CATHETER WOULD NOT INFUSE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7209546 ↗MalfunctionDec 4, 2017Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE FOLEY CATHETER SEEMED THICKER THAN USUAL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7080368 ↗
Other CATHETER, COUDE devices
- Bardex® Lubricath® Foley Catheter 0165L14C. R. Bard, Inc. · CATHETER, COUDE562 reports
- Bardex® Lubricath® Foley Catheter 0165L16C. R. Bard, Inc. · CATHETER, COUDE539 reports
- Bardia® Foley Catheter Silicone Coated 123516AC. R. Bard, Inc. · CATHETER, COUDE381 reports
- Bardia® Foley Catheter Silicone Coated 123514AC. R. Bard, Inc. · CATHETER, COUDE195 reports
- SureStep™ Foley Tray System Bardex® Lubricath® Foley Catheter A899916C. R. Bard, Inc. · CATHETER, COUDE3 recalls19 reports
- Bardia® Foley Catheter Silicone Coated 123518AC. R. Bard, Inc. · CATHETER, COUDE153 reports
- SureStep™ Foley Tray System Bardex® Lubricath® Foley Catheter A902916C. R. Bard, Inc. · CATHETER, COUDE2 recalls50 reports
- Bard® Lubricath® Foley Catheter 365716C. R. Bard, Inc. · CATHETER, COUDE117 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.