Bardex® Lubricath® Foley Catheter Councill Model 0196L22Recalls & reported problems
C. R. Bard, Inc. · Bardex® Lubricath® Councill Model Foley Catheter Red Latex Radiopaque 5cc Balloon Lubricious Coated · Class 2
All models of Bardex Lubricath Foley Catheter Councill Model (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741017957), received between Jul 12, 2017 and Feb 3, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Fell” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Fell2
- Expiration Date Error2
- Material Split, Cut or Torn2
- Deflation Problem1
- Inadequate Instructions for Healthcare Professional1
- Material Fragmentation1
- Material Puncture/Hole1
Patient problems
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2026Material Puncture/Hole
IT WAS REPORTED THAT THE LATEX FOLEYS WERE NOT STAYING INFLATED AND WERE FALLING OUT OF PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24250907 ↗MalfunctionJan 31, 2020Expiration Date Error; Material Split, Cut or Torn; Device Fell; Expiration Date Error
IT WAS REPORTED THAT THE CATHETER FELL OUT OF THE PATIENT WITH A LARGE SLIT FOUND ON THE BALLOON. PER SAMPLE EVALUATION, IT WAS FOUND THAT THE CATHETER DATECODE APPEARED TO HAVE A PRINTING ERROR, THE CATHETER PACKAGING WAS MISSING AN EXPIRATION DATE, AND ACCORDING TO THE NOGALES JDE SEARCH, THE CATHETER WAS EXPIRED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9653530 ↗MalfunctionJan 31, 2020Inadequate Instructions for Healthcare Professional; Expiration Date Error; Material Split, Cut or Torn; Device Fell; Inadequate Instructions for Healthcare Professional
IT WAS REPORTED THAT THE CATHETER FELL OUT OF THE PATIENT WITH A LARGE SLIT FOUND ON THE BALLOON. PER SAMPLE EVALUATION, IT WAS FOUND THAT THE CATHETER DATECODE APPEARED TO HAVE A PRINTING ERROR, THE CATHETER PACKAGING WAS MISSING AN EXPIRATION DATE, AND ACCORDING TO THE NOGALES JDE SEARCH, THE CATHETER WAS EXPIRED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9653528 ↗MalfunctionNov 5, 2019Deflation Problem; Deflation Problem
IT WAS REPORTED THAT THE CATHETER FELL OUT OF THE PATIENT WITH A LARGE SLIT FOUND ON THE BALLOON. ADDITIONAL INFORMATION: 21OCT2019 - CUSTOMER UPDATED: THIS ITEM WAS USED ON A PATIENT. THE PATIENT RETURNED STATING THE CATHETER FELL OUT, AND THE BALLOON WAS NO LONGER DEFLATED. I NOTICED UPON CLOSER INSPECTION A CUT IN THE BALLOON. THERE WAS NO PATIENT INJURY, AND THE PATIENT WAS RE-CATHETERIZED WITH NO ISSUES. THERE WAS NO LIFE THREATENING STATE, NO PERMANENT IMPAIRMENT AND THE ONLY INTERVENTION WAS TO PLACE A NEW CATHETER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9280101 ↗MalfunctionJul 12, 2017Material Fragmentation
IT WAS REPORTED THAT WHEN THE CATHETER WAS REMOVED, THE BALLOON WAS NOTED TO BE IN FRAGMENTS. IT WAS LATER REPORTED FRAGMENTS WERE LEFT IN THE PATIENT, HOWEVER THE PATIENT'S FAMILY CHOSE NOT TO HAVE THEM REMOVED. THE CATHETER WAS PLACED BY A PHYSICIAN DURING A CYSTOGRAM ON (B)(6) 2017.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6709449 ↗
Other CATHETER, COUDE devices
- Bardex® Lubricath® Foley Catheter 0165L14C. R. Bard, Inc. · CATHETER, COUDE562 reports
- Bardex® Lubricath® Foley Catheter 0165L16C. R. Bard, Inc. · CATHETER, COUDE539 reports
- Bardia® Foley Catheter Silicone Coated 123516AC. R. Bard, Inc. · CATHETER, COUDE381 reports
- Bardia® Foley Catheter Silicone Coated 123514AC. R. Bard, Inc. · CATHETER, COUDE195 reports
- SureStep™ Foley Tray System Bardex® Lubricath® Foley Catheter A899916C. R. Bard, Inc. · CATHETER, COUDE3 recalls19 reports
- Bardia® Foley Catheter Silicone Coated 123518AC. R. Bard, Inc. · CATHETER, COUDE153 reports
- SureStep™ Foley Tray System Bardex® Lubricath® Foley Catheter A902916C. R. Bard, Inc. · CATHETER, COUDE2 recalls50 reports
- Bard® Lubricath® Foley Catheter 365716C. R. Bard, Inc. · CATHETER, COUDE117 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.