Bardex® Lubricath® Foley Catheter Tiemann Model Coude 0103L24Recalls & reported problems
C. R. Bard, Inc. · Bardex® Lubricath® Tiemann Model Foley Catheter Coude Tip 30cc Balloon Red Latex Radiopaque Lubricious Coated · Class 2
All models of Bardex Lubricath Foley Catheter Tiemann Model Coude (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741018220), received between Oct 18, 2022 and Feb 23, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Deflation Problem1
- Deformation Due to Compressive Stress1
- Difficult to Insert1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 23, 2023Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED THAT THE SURVEY RESPONDENT STATED NO, PATIENT DID NOT AGREE THAT THE BENEFITS OF USING THE BARD/BECTON DICKINSON BIOCATH PRODUCT OUTWEIGH THE POTENTIAL RISKS BECAUSE IT MIGHT GET IN A KNOT IN RELATION TO TIEMANN MODEL, LUBRICATH, 2-WAY, 30CC BALLOON, FRENCH SIZE 24.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16428393 ↗InjuryOct 18, 2022Deflation Problem; Deflation Problem
IT WAS REPORTED THAT THE PATIENT HAD A FOLEY CATHETER WITH 30 CC BALLOON INSERTED DUE TO STRAIGHT CATHETER UNSUCCESSFUL. ON (B)(6) 2022, THE PATIENT WAS PULLING AT FOLEY AND CAUSING TRAUMA AND DEVELOPED SANG DISCHARGE IN FOLEY. WHEN ATTEMPTED TO REMOVE FOLEY, IT WAS UNABLE TO REMOVE DUE TO BULB NOT DEFLATING. CONFIRMED ON ULTRASOUND THAT BULB WAS FULL WITH FLUID. ATTEMPTED CLIENT REPOSITIONING AND MULTIPLE ATTEMPTS TO WITHDRAW FLUID. SURGICAL CONSULT WAS RECOMMENDED TO HAVE AN ULTRASOUND GUIDED PROCEDURE COMPLETED. ON (B)(6) 2022, THE PATIENT WAS LEFT TO FACILITY 2 FOR SURGICAL CONSULT. BULB WAS DEFLATED AND NEW FOLEY INSERTED. REMOVED WHEN BACK IN FACILITY. INCIDENT WAS THOUGHT TO HAVE OCCURRED DUE TO FAULTY EQUIPMENT AND WAS ABLE TO REMOVE ABOUT 2.5 ML OR LESS OF FLUID, UNABLE TO REMOVE MORE. BULB WAS NOT TESTED/INFLATED PRIOR TO INSERTION. THIS WAS NOT DONE DUE TO RECENT CHANGES TO BEST PRACTICE FOR FOLEY CATHETER INSERTION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 15626905 ↗MalfunctionOct 18, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE PATIENT HAD A FOLEY CATHETER WITH 30 CC BALLOON INSERTED DUE TO STRAIGHT CATHETER UNSUCCESSFUL. ON (B)(6) 2022, THE PATIENT WAS PULLING AT FOLEY AND CAUSING TRAUMA AND DEVELOPED SANG DISCHARGE IN FOLEY. WHEN ATTEMPTED TO REMOVE FOLEY, IT WAS UNABLE TO REMOVE DUE TO BULB NOT DEFLATING. CONFIRMED ON ULTRASOUND THAT BULB WAS FULL WITH FLUID. ATTEMPTED CLIENT REPOSITIONING AND MULTIPLE ATTEMPTS TO WITHDRAW FLUID. SURGICAL CONSULT WAS RECOMMENDED TO HAVE AN ULTRASOUND GUIDED PROCEDURE COMPLETED. ON (B)(6) 2022, THE PATIENT WAS LEFT TO FACILITY 2 FOR SURGICAL CONSULT. BULB WAS DEFLATED AND NEW FOLEY INSERTED. REMOVED WHEN BACK IN FACILITY. INCIDENT WAS THOUGHT TO HAVE OCCURRED DUE TO FAULTY EQUIPMENT AND WAS ABLE TO REMOVE ABOUT 2.5 ML OR LESS OF FLUID, UNABLE TO REMOVE MORE. BULB WAS NOT TESTED/INFLATED PRIOR TO INSERTION. THIS WAS NOT DONE DUE TO RECENT CHANGES TO BEST PRACTICE FOR FOLEY CATHETER INSERTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15624446 ↗
Other CATHETER, UROLOGICAL devices
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50614C. R. Bard, Inc. · CATHETER, UROLOGICAL883 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53614GC. R. Bard, Inc. · CATHETER, UROLOGICAL692 reports
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50616C. R. Bard, Inc. · CATHETER, UROLOGICAL2 deaths443 reports
- GENTLECATH GLIDE 421567Convatec, Purchasing Department · Catheter, Urological342 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53616GC. R. Bard, Inc. · CATHETER, UROLOGICAL301 reports
- Bardex® Foley Catheter Silicone Coated 0167V22SC. R. Bard, Inc. · CATHETER, UROLOGICAL187 reports
- Bardex® Foley Catheter Silicone Coated 0167V20SC. R. Bard, Inc. · CATHETER, UROLOGICAL175 reports
- Bardia® Coude Olive Tip Urethral Foley Catheter 802516C. R. Bard, Inc. · CATHETER, UROLOGICAL170 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page