Skip to content

Bardex® Lubricath® Foley Catheter Emmett Model 0138L24Recalls & reported problems

C. R. Bard, Inc. · Bardex® Lubricath® Emmett Model Foley Catheter Three Eyes Long Length Round Tip 3 Way Continuous Irrigation 30cc Balloon Lubricious Coated · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741018862), received between Nov 22, 2024 and Nov 22, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Inaccurate Flow Rate” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inaccurate Flow Rate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2024Inaccurate Flow Rate; Inaccurate Flow Rate

    IT WAS REPORTED THAT THE URINE FLOWS FROM THE 3RD CHANNEL PORT TIP AND THE DRAINAGE CHANEL WAS CONNECTED TO THE URINARY BAG FOR DRAINING, BUT THE URINE GOES TO BOTH THESE 2 CHANNELS WHICH WAS NOT SUPPOSED TO. PER ADDITIONAL INFORMATION RECEIVED VIA MAIL ON 23DEC2024, STATED THAT THE URINE WAS LEAKING FROM THE 3RD LUMEN WHICH WAS SUPPOSED TO USE FOR IRRIGATION ONLY AND THEY PUT A LABOR BAND ON THE CATHETER TIP TO KINK AND PREVENT THE LEAKAGE. PER CLARIFICATION MAIL RECEIVED FROM UCC ON 30DEC2024, STATED THAT THE PATIENT HAD ISSUES WITH 2 MATERIAL# (MATERIAL#0138L24 AND MATERIAL#0167L24).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20759496 ↗

Other CATHETER, UROLOGICAL devices

All CATHETER, UROLOGICAL devices →