Adverse event reports
9
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741025198), received between May 15, 2018 and Nov 7, 2025 — classified as 4 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem1
  • Difficult to Remove1
  • Entrapment of Device1
  • Fluid/Blood Leak1
  • Fracture1
  • Hole In Material1
  • Inaccurate Flow Rate1
  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient2
  • No Consequences Or Impact To Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Known Impact Or Consequence To Patient1
  • Patient Problem/Medical Problem1
  • Urinary Tract Infection1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2025Difficult to Remove; Difficult to Remove

    IT WAS REPORTED FROM THE UROLOGY SERVICE COORDINATOR THERE THAT (B)(6). THEY WERE HOPING THEY COULD HELP THEM OUT. THEY WERE HAVING SOME ISSUES WITH THE LUBRICATH HEMATURIA FOLEY CATHETERS. THE 22F AND 24F (REF 2557H22 AND 2557H24) WERE HAVING ISSUES WITH THE INFLOW PORT NOT WORKING PROPERLY AND THE BALLOON HAVING ISSUES WITH DEFLATING. THOSE ISSUES THAT LEAD TO RISKY CATHETER CHANGES AFTER A DELICATE PROCEDURE CAUSING EITHER ADDITIONAL BLEEDING AND/OR ADDITIONAL PROCEDURES.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 23497562 ↗
  • InjuryApr 1, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE CUSTOMER MENTIONED ABOUT CATHETER BLOCKAGE, LEAKAGE, PAIN AND RETAINED FRAGMENTS FROM CATHETER WHEN THEY WERE ASKED OF COMPLICATIONS WHILE USING THE FOLEY CATHETER. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION IS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 13977393 ↗
  • MalfunctionMar 31, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE CUSTOMER MENTIONED ABOUT CATHETER BLOCKAGE, LEAKAGE, PAIN AND RETAINED FRAGMENTS FROM CATHETER WHEN THEY WERE ASKED OF COMPLICATIONS WHILE USING THE FOLEY CATHETER. MEDICAL INTERVENTION WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13971478 ↗
  • MalfunctionMar 31, 2022Partial Blockage; Partial Blockage

    IT WAS REPORTED THAT THE CUSTOMER MENTIONED ABOUT CATHETER BLOCKAGE, LEAKAGE, PAIN AND RETAINED FRAGMENTS FROM CATHETER WHEN THEY WERE ASKED ABOUT THE COMPLICATIONS WHILE USING THE FOLEY CATHETER. NO MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13959924 ↗
  • MalfunctionJun 23, 2021Fracture; Material Split, Cut or Torn; Fracture; Material Split, Cut or Torn

    IT WAS REPORTED THAT A HOLE WAS FOUND IN THE FOLEY CATHETER WHICH LEAKED IRRIGATION AND TWO CATHETERS WERE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12054180 ↗
  • MalfunctionJan 3, 2021Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THERE WAS A PUNCTURE IN THE BALLOON SURFACE AFTER INSERTING THE HEMATURIA CATHETER AND FILLING THE BALLOON WITH LIQUID. HENCE THE CATHETER COULD NOT REMAIN IN PLACE AND WAS REPLACED BY ANOTHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11105626 ↗
  • InjuryFeb 12, 2020Deflation Problem; Entrapment of Device; Deflation Problem

    IT WAS REPORTED THAT THE CATHETER GOT STUCK IN THE PATIENT'S URETHRA. THE CATHETER WAS REMOVED BY SURGERY.

    Patient problems: Foreign Body In Patient; Patient Problem/Medical Problem

    MAUDE report 9701924 ↗
  • InjuryFeb 21, 2019Inaccurate Flow Rate; Infusion or Flow Problem; Inaccurate Flow Rate

    IT WAS REPORTED THAT THE CATHETER DID NOT VOID AT ALL. ALLEGEDLY, AS A RESULT THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR AN URINARY TRACT INFECTION. THE INFECTION WAS TREATED WITH ANTIBIOTICS.

    Patient problems: Urinary Tract Infection

    MAUDE report 8359209 ↗
  • MalfunctionMay 15, 2018Material Rupture; Hole In Material; Material Rupture

    IT WAS REPORTED THAT A HOLE WAS OBSERVED IN THE CATHETER BALLOON WHEN IT WAS REMOVED FROM THE PATIENT. NO PIECES OF THE BALLOON WERE REPORTED MISSING AND NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7516436 ↗

Other CATHETER, UROLOGICAL devices