BardPort, Groshong 0602810Recalls & reported problems

Bard Access Systems, Inc. · BardPort Titanium Implantable Port with Attachable 8F Groshong Single-Lumen Venous Catheter Intermediate Kit · Class 2 · Implant

All models of BardPort, Groshong (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00801741025693), received between Aug 9, 2018 and Jan 12, 2026 — classified as 10 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture8
  • Material Separation6
  • Migration5
  • Fluid/Blood Leak3
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Deformation Due to Compressive Stress1
  • Difficult to Remove1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions2
  • Extravasation1
  • Injury1
  • No Consequences Or Impact To Patient1
  • Pain1
  • Swelling/ Edema1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 12, 2026Fracture; Material Separation; Migration

    ON (B)(6) 2021, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE TITANIUM IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F. POST THE PROCEDURE ON (B)(6) 2025, DURING REMOVAL OF THE CIP, IT WAS NOTED THAT THE CATHETER WAS RUPTURED AND SEPARATED. THE CATHETER WAS REMOVED. REPORTEDLY, IN A POST-REMOVAL CHEST X-RAY, IT WAS FURTHER REPORTED THAT A PIECE OF CATHETER HAD MOVED INTO THE RIGHT HEART CHAMBER. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24033327 ↗
  • InjuryJan 12, 2026Fracture; Material Separation; Migration

    ON (B)(6) 2021, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE TITANIUM IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F. POST THE PROCEDURE ON (B)(6) 2025, DURING REMOVAL OF THE CIP, IT WAS NOTED THAT THE CATHETER WAS RUPTURED AND SEPARATED. THE CATHETER WAS REMOVED. REPORTEDLY, IN A POST-REMOVAL CHEST X-RAY, IT WAS FURTHER REPORTED THAT A PIECE OF CATHETER HAD MOVED INTO THE RIGHT HEART CHAMBER. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24033281 ↗
  • InjuryDec 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME LATER PORT PLACEMENT PROCEDURE, THE PATIENT DEVELOPED WITH UNSPECIFIED INFECTION. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 23825592 ↗
  • MalfunctionApr 10, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT ONE DAY POST PORT PLACEMENT PROCEDURE, MICRODROPLETS FORMATION WAS IDENTIFIED IN THE CATHETER WALL, AND CAUSED THE LEAKAGE. PORT WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21812894 ↗
  • InjuryJan 30, 2025Fracture; Difficult to Remove; Material Separation; Fracture; Difficult to Remove; Material Separation

    IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME POST PORT PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY FRACTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: Extravasation; Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21270789 ↗
  • InjuryNov 27, 2023Fracture; Material Separation; Migration; Fracture; Material Separation; Migration

    IT WAS REPORTED THAT APPROXIMATELY TWO YEARS POST A PORT PLACEMENT, THE CATHETER WAS ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT A PART OF THE CATHETER HAD MIGRATED TO THE RIGHT PULMONARY ARTERY. REPORTEDLY, PROGRAMMING OF SURGICAL PROCEDURES WAS DONE TO REMOVE THE CATHETER AND THE IMPLANTABLE CHAMBER. THE PATIENT IS REPORTED TO BE STABLE NOW.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18208113 ↗
  • InjuryNov 27, 2023Fracture; Material Separation; Migration; Fracture; Material Separation; Migration

    IT WAS REPORTED THAT APPROXIMATELY TWO YEARS POST A PORT PLACEMENT, THE CATHETER WAS ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT A PART OF THE CATHETER HAD MIGRATED TO THE RIGHT PULMONARY ARTERY. REPORTEDLY, PROGRAMMING OF SURGICAL PROCEDURES WAS DONE TO REMOVE THE CATHETER AND THE IMPLANTABLE CHAMBER. THE PATIENT IS REPORTED TO BE STABLE NOW.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18208112 ↗
  • InjuryNov 24, 2023Fracture; Material Separation; Deformation Due to Compressive Stress; Naturally Worn; Migration; Fracture; Material Separation; Deformation Due to Compressive Stress; Naturally Worn; Migration

    IT WAS REPORTED THAT APPROXIMATELY TWO YEARS POST A PORT PLACEMENT, THE CATHETER WAS ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT A PART OF THE CATHETER HAD MIGRATED TO THE RIGHT PULMONARY ARTERY. REPORTEDLY, PROGRAMMING OF SURGICAL PROCEDURES WAS DONE TO REMOVE THE CATHETER AND THE IMPLANTABLE CHAMBER. THE PATIENT IS REPORTED TO BE STABLE NOW.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18203911 ↗
  • InjuryMay 13, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT APPROXIMATELY THIRTY ONE DAYS POST PORT PLACEMENT PROCEDURE, THE PORT CATHETER ALLEGEDLY LEAKED AT THE SUTURE LINE. IT WAS FURTHER REPORTED THAT THE SUTURE SITE SHOWED SIGNS OF INFECTION AND THE INFECTION WAS TREATED. REPORTEDLY THE PORT CATHETER WAS REMOVED. THE PATIENT REPORTED TO BE STABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Swelling/ Edema; Unspecified Infection

    MAUDE report 14394990 ↗
  • InjuryMar 25, 2020Fracture; Fracture

    IT WAS REPORTED THAT APPROXIMATELY TWO YEARS SEVEN MONTHS POST PORT DEVICE IMPLANT, THE CATHETER ALLEGEDLY BROKE AT THE CONNECTION POINT AND MIGRATED TO THE HEART, WHICH WAS CONFIRMED BY X-RAY IMAGING. REPORTEDLY, THE CATHETER WAS REMOVED SURGICALLY BY INTERVENTIONAL RADIOLOGY. THE PATIENT IS REPORTED AS STABLE.

    Patient problems: Injury

    MAUDE report 9878121 ↗
  • MalfunctionJul 22, 2019Failure to Infuse; Failure to Infuse

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0602810 IMPLANTED PORT ALLEGEDLY WAS UNABLE TO BE INFUSED. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. PATIENT AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8814186 ↗
  • InjuryJan 24, 2019Break; Break

    IT WAS REPORTED THAT TWO YEARS AND TWO MONTHS POST PORT PLACEMENT, A SCAN ALLEGEDLY DEMONSTRATED THE PORT CATHETER RUPTURED AND MIGRATED TO THE ARTERY LEVEL OF THE LEFT LUNG. IT WAS FURTHER REPORTED THAT THE ADDITIONAL INTERVENTION WAS USED TO COMPLETE REMOVAL OF THE MIGRATED SEGMENT. REPORTEDLY, THE DEVICE WAS REMOVED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 8276398 ↗
  • MalfunctionAug 9, 2018Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    IT WAS REPORTED THAT THE PORT WAS PLACED IN THE RIGHT INTERNAL JUGULAR VEIN. PATIENT COMPLAINED OF PAIN AND SWELLING DURING INFUSION. ULTRASOUND REVEALED A CRACK IN THE CATHETER WITH A CONTRAST LEAK NEAR THE LIMB AND INSERTION SITE. DEVICE WAS REMOVED IN INTERVENTIONAL RADIOLOGY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: Pain

    MAUDE report 7770364 ↗

Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices

All PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices →