BardPort M.R.I. 0603840Recalls & reported problems
Bard Access Systems, Inc. · BardPort M.R.I. Low-Profile Implantable Port with Attachable 6.6F Open-Ended Single-Lumen Venous Catheter Basic Kit · Class 2 · Implant
All models of BardPort M.R.I. (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741025822), received between May 8, 2019 and Sep 22, 2021 — classified as 4 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture4
- Break3
- Fluid/Blood Leak3
- Disconnection2
- Material Separation2
- Dent in Material1
- Material Deformation1
- Migration or Expulsion of Device1
Patient problems
- Injury2
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Extravasation1
- No Known Impact Or Consequence To Patient1
- Swelling/ Edema1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 22, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT SOMETIME POST PORT DEVICE PLACEMENT, LIQUID WAS ALLEGEDLY NOTED TO BE LEAKING SUBCUTANEOUSLY FROM THE CONNECTION OF THE PIPE. IT WAS FURTHER REPORTED THAT SWELLING UNDER THE SKIN AND EXTRAVASATION WAS NOTED ON THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Extravasation; No Clinical Signs, Symptoms or Conditions
MAUDE report 12511421 ↗InjuryJan 18, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT POST DEVICE IMPLANT, THE DEVICE ALLEGEDLY LEAKED DRUG FROM THE DIAPHRAGM. IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED REDNESS AND SWELLING DUE TO THE LEAK. THE PORT WAS REMOVED AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Swelling/ Edema
MAUDE report 11184397 ↗MalfunctionNov 23, 2020Disconnection; Fluid/Blood Leak; Fracture; Stretched; Material Deformation; Disconnection; Fluid/Blood Leak; Fracture; Stretched; Material Deformation
IT WAS REPORTED THAT POST PORT DEVICE IMPLANT, THE CATHETER ALLEGEDLY DISCONNECTED FROM THE PORT STEM AND HAD A SUBCUTANEOUS LEAKAGE. IT WAS FURTHER REPORTED THAT THE CATHETER WAS ALLEGEDLY DAMAGED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10883721 ↗InjuryFeb 14, 2020Disconnection; Fracture; Migration or Expulsion of Device; Disconnection; Fracture; Migration or Expulsion of Device
IT WAS REPORTED THAT APPROXIMATELY TWO YEARS THREE MONTHS POST PORT DEVICE IMPLANT, THE CATHETER ALLEGEDLY CRACKED AND DISCONNECTED FROM THE PORT STEM. IT WAS FURTHER REPORTED THAT THE CATHETER ALLEGEDLY MIGRATED TO THE ATRIUM. REPORTEDLY, THE DEVICE WAS REMOVED. THE PATIENT WAS REPORTED AS STABLE.
Patient problems: Injury; No Known Impact Or Consequence To Patient
MAUDE report 9714886 ↗MalfunctionJan 15, 2020Break; Fracture; Material Separation; Dent in Material; Break; Fracture; Material Separation; Dent in Material
THIS REPORT SUMMARIZES ONE MALFUNCTION. THE INFORMATION REVIEWED INDICATED THAT MODEL 0603840 IMPLANTABLE PORT ALLEGEDLY EXPERIENCED A BREAK. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED CONSEQUENCES. THE MALE PATIENT'S AGE AND WEIGHT WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9592520 ↗MalfunctionOct 18, 2019Break; Fracture; Material Separation; Break; Fracture; Material Separation
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0603840, IMPLANTABLE PORT, ALLEGEDLY EXPERIENCED A BREAK IN THE CATHETER. THIS INFORMATION WAS OBTAINED FROM A SINGLE SOURCE. THIS MALFUNCTION INVOLVED A FIVE YEAR OLD FEMALE PATIENT WITH NO CONSEQUENCES, THE PATIENT'S WEIGHT WAS NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9208942 ↗InjuryMay 8, 2019Break; Break
IT WAS REPORTED THAT APPROXIMATELY THREE MONTHS POST PORT PLACEMENT, AN ALLEGED BREAK IN THE PORT WAS IDENTIFIED. REPORTEDLY, A SEGMENT OF THE PORT WAS REMOVED AND A SECOND SEGMENT WAS REMOVED THE FOLLOWING DAY. THE PATIENT WAS STABLE AFTER PROCEDURE.
Patient problems: Injury
MAUDE report 8591978 ↗
Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices
- PowerPort M.R.I., Groshong 9808560Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR3 deaths834 reports
- PowerPort ClearVUE 1608062Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR18 deaths571 reports
- PowerPort ClearVUE 5608062Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR31 deaths542 reports
- PowerPort M.R.I., ChronoFlex 1808000Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR19 deaths513 reports
- PowerPort M.R.I., ChronoFlex 1808060Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall15 deaths383 reports
- PowerPort M.R.I., Groshong 1808560Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR10 deaths304 reports
- PowerPort M.R.I., ChronoFlex 1829500Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR6 deaths197 reports
- BardPort M.R.I. 0602680Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR2 deaths179 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page