BardPort 0604970Recalls & reported problems

Bard Access Systems, Inc. · BardPort Low-Profile Dual-Lumen Implantable Port with Attachable 7F Open-Ended Dual-Lumen Venous Catheter. Intermediate Kit · Class 2 · Implant

All models of BardPort (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00801741025884), received between Sep 14, 2018 and Mar 30, 2026 — classified as 1 death, 2 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Fracture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture6
  • Adverse Event Without Identified Device or Use Problem3
  • Fluid/Blood Leak2
  • Crack1
  • Difficult to Flush1
  • Failure to Infuse1
  • Loose or Intermittent Connection1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Consequences Or Impact To Patient3
  • Unspecified Infection3
  • Thrombosis/Thrombus1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • DeathMar 30, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT, SOMETIME AFTER THE PORT PLACEMENT, THE PATIENT WAS DIAGNOSED WITH AN UNSPECIFIED INFECTION. IT WAS FURTHER REPORTED THAT THE PATIENT EVENTUALLY EXPIRED. HOWEVER, THE CAUSE OF DEATH WAS NOT REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 24726864 ↗
  • MalfunctionAug 12, 2025Fracture; Fracture

    IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT SOMETIME POST A PORT PLACEMENT, THE CATHETER HAD ALLEGEDLY FRACTURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22781525 ↗
  • InjuryJun 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME POST PORT PLACEMENT, THE PATIENT DEVELOPED INFECTION AND THROMBOSIS. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Thrombosis/Thrombus; Unspecified Infection

    MAUDE report 22296953 ↗
  • MalfunctionApr 28, 2025Fracture; Fracture

    ON (B)(6) 2003, PATIENT UNDERWENT BARD SLIMPORT DUAL-LUMEN ROSENBLATT IMPLANTABLE PORT PLACEMENT PROCEDURE FOR UNKNOWN MEDICAL CONDITION. ABOUT SOMETIMES POST A PORT PLACEMENT, THE CATHETER ALLEGEDLY HAD FRACTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21916480 ↗
  • InjuryFeb 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME POST A PORT PLACEMENT, THE PATIENT ALLEGEDLY DEVELOPED INFECTION. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 21353796 ↗
  • MalfunctionFeb 10, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A PREPARATION OF PORT PLACEMENT PROCEDURE, THE TIP WAS ALLEGEDLY BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21341800 ↗
  • MalfunctionFeb 10, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A PREPARATION OF PORT PLACEMENT PROCEDURE, THE TIP WAS ALLEGEDLY BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21341625 ↗
  • MalfunctionJan 26, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, THE PORT ALLEGEDLY LEAKED WHILE FLUSHING. IT WAS FURTHER REPORTED THAT THE PORT WAS REMOVED AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16245180 ↗
  • MalfunctionApr 12, 2022Material Deformation; Material Deformation

    IT WAS REPORTED THAT PRIOR TO A PORT PLACEMENT PROCEDURE, THE TIP OF THE DILATOR WAS ALLEGEDLY NOTED TO BE SPLIT/FRAYED UPON OPENING THE DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14081967 ↗
  • MalfunctionOct 27, 2021Difficult to Flush; Difficult to Flush

    IT WAS FURTHER REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, THE PORT WAS ALLEGEDLY UNABLE TO BE FLUSHED. THE DEVICE WAS REMOVED FROM THE PATIENT AND REPLACED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12704636 ↗
  • MalfunctionMar 8, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, THE DEVICE COULDN'T ATTACH THE CATHETER TO THE CHAMBER HUB. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11441255 ↗
  • MalfunctionFeb 25, 2021Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    IT WAS REPORTED THAT SOME TIME POST PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY HAD DROPLETS WHEN CONNECTED TO THE DUAL PORT HUB. IT WAS FURTHER REPORTED THAT ANOTHER CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11377003 ↗
  • MalfunctionOct 17, 2019Fracture; Fracture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0604970. IMPLANTABLE PORT ALLEGEDLY EXPERIENCED A FRACTURE. THIS REPORT WAS RECEIVED FROM A SINGLE SOURCE. THIS EVENT DID INVOLVE PATIENT WITH NO REPORTED PATIENT INJURY. THE PATIENTS AGE, WEIGHT AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9205802 ↗
  • MalfunctionOct 17, 2019Crack; Crack

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0604970 PORT AND CATHETER ALLEGEDLY EXPERIENCED A CRACK. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. ONE PATIENT WAS INVOLVED AND THERE WAS NO REPORTED PATIENT INJURY. THE MALE PATIENT IS (B)(6) OLD AND WEIGHS (B)(6) KGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9205798 ↗
  • MalfunctionSep 14, 2018Failure to Infuse; Failure to Infuse

    IT WAS REPORTED THAT THE PATIENT¿S DUAL PORT WOULD NOT ASPIRATE OR INFUSE. DURING THE PATIENT¿S CHEMOTHERAPY TREATMENT ONLY ONE PORT WORKED AND HE HAD TO HAVE AN IV PLACED FOR FLUIDS. APPROXIMATELY ONE WEEK AGO BOTH PORTS STOPPED WORKING AND THE PATIENT HAD TO HAVE CHEMOTHERAPY THROUGH HIS IV. THE PORT WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7879531 ↗

Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices

All PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices →