Skip to content

SlimPort 0605550Recalls & reported problems

Bard Access Systems, Inc. · SlimPort Titanium Implantable Port with Attachable 6F Open-Ended Single-Lumen Venous Catheter Intermediate Kit · Class 2 · Implant

Adverse event reports
11
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741025976), received between Feb 7, 2019 and May 19, 2026 — classified as 4 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Break1
  • Component Missing1
  • Defective Component1
  • Deformation Due to Compressive Stress1
  • Difficult to Flush1
  • Fluid/Blood Leak1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient3
  • Bacteremia1
  • Foreign Body In Patient1
  • Injury1
  • Thrombosis/Thrombus1
  • Unspecified Infection1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Migration

    IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT, SOMETIME AFTER THE PORT PLACEMENT, THE PORT HAD ALLEGEDLY MIGRATED. HOWEVER, THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25215122 ↗
  • InjuryNov 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT SOMETIME AFTER THE PORT PLACEMENT, THE PATIENT WAS DIAGNOSED WITH THROMBOSIS. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 23448389 ↗
  • InjuryMay 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME LATER PORT PLACEMENT, THE PATIENT DEVELOPED WITH UNSPECIFIED INFECTION. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 22024279 ↗
  • InjuryMar 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH LITIGATION PROCESS THAT APPROXIMATELY SIX MONTHS AND TWENTY-FIVE DAYS POST FILTER PLACEMENT THE PATIENT DEVELOPED WITH BACTEREMIA. IT WAS FURTHER REPORTED THAT THE PORT WAS REMOVED, AND NEW PORT HAS BEEN IMPLANTED. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Bacteremia

    MAUDE report 18892476 ↗
  • MalfunctionMar 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT SOMETIME PORT PLACEMENT, THE PATIENT ALLEGEDLY EXPERIENCED PAIN IN SHOULDER. HOWEVER, THERE IS NO INFORMATION AVAILABLE REGARDING ADDITIONAL INTERVENTION OR MEDICATION TO TREAT SHOULDER PAIN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16461137 ↗
  • MalfunctionApr 5, 2022Difficult to Flush; Deformation Due to Compressive Stress; Material Twisted/Bent; Difficult to Flush; Deformation Due to Compressive Stress; Material Twisted/Bent

    IT WAS REPORTED THAT SOMETIMES POST PORT PLACEMENT PROCEDURE, THE DEVICE ALLEGEDLY HAD A TWIST. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14011986 ↗
  • MalfunctionDec 16, 2020Component Missing; Packaging Problem; Component Missing; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO A PORT PLACEMENT PROCEDURE FOR CHEMOTHERAPY, THE PACKAGING WAS ALLEGEDLY INCOMPLETE. IT WAS FURTHER REPORTED THAT FEW COMPONENT WERE MISSING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11017090 ↗
  • MalfunctionJan 16, 2020Material Frayed; Material Frayed

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0605550. IMPLANTABLE PORT ALLEGEDLY EXPERIENCED MATERIAL FRAYED. THIS REPORT WAS RECEIVED FROM A SINGLE SOURCE. THIS EVENT DID INVOLVE PATIENT WITH NO PATIENT CONSEQUENCES. THE PATIENT IS (B)(6) YEARS OF AGE, THE GENDER IS FEMALE, AND THE WEIGHT IS (B)(6) KGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9598349 ↗
  • MalfunctionJul 23, 2019Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES ONE (1) MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0605550 IMPLANTED PORT ALLEGEDLY EXPERIENCED A FLUID LEAK. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE PATIENT¿S AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8819799 ↗
  • MalfunctionJul 17, 2019Defective Component; Defective Component

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 605550 PORT & CATHETER ALLEGEDLY HAD DEFECTIVE COMPONENTS. THIS INFORMATION WAS RECEIVED FROM A SINGLE SOURCE. THE ALLEGED MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE PATIENT WAS REPORTED TO BE A (B)(6) MALE WHOSE AGE WAS NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8800160 ↗
  • InjuryFeb 7, 2019Break; Break

    IT WAS REPORTED THAT APPROXIMATELY TWO YEARS POST PORT IMPLANT, THE PORT WAS ALLEGEDLY DIFFICULT TO INFUSE. IT WAS FURTHER REPORTED THAT THE PORT CATHETER ALLEGEDLY RUPTURED AND BROKE INTO TWO SEGMENTS. REPORTEDLY, THE DISTAL END OF THE PORT CATHETER MIGRATED INTO THE CARDIAC CAVITY AND WAS REMOVED BY INTERVENTIONAL RADIOLOGY WITHOUT COMPLICATION. IT WAS FURTHER REPORTED THAT THE REMAINING PORT CATHETER AND PORT BODY WERE REMOVED AT A LATER DATE. THE PATIENT STATUS POST REMOVAL PROCEDURE IS UNKNOWN.

    Patient problems: Foreign Body In Patient; Injury

    MAUDE report 8316704 ↗

Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices

All PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.