PowerPort ClearVUE 1606052Recalls & reported problems
Bard Access Systems, Inc. · PowerPort ClearVUE Implantable Port with Attachable 6F Polyurethane Open-Ended Single-Lumen Venous Catheter Microintroducer Kit · Class 2 · Implant
All models of PowerPort ClearVUE (27)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This PowerPort ClearVUE implantable port is a small vessel implanted under the skin for chemotherapy and other infusions. The 56 reports mostly describe the port or its catheter breaking apart, separating from the port body, or leaking fluid and blood;17 reports involve injuries and no deaths. Reports dropped from 14 in the prior 12 months to 8. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 15
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 00801741026409), received between Jan 15, 2019 and Jul 27, 2026 — classified as 15 injuries, 39 malfunctions.
The most frequently reported specific device problem is “Fracture” (19 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture19
- Material Separation13
- Fluid/Blood Leak9
- Deformation Due to Compressive Stress8
- Adverse Event Without Identified Device or Use Problem5
- Difficult or Delayed Separation5
- Difficult to Flush5
- Detachment of Device or Device Component4
Patient problems
- No Clinical Signs, Symptoms or Conditions36
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
- Unspecified Infection4
- Swelling/ Edema2
- Bacterial Infection1
- Chest Pain1
- Discomfort1
- Fever1
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2026Difficult to Flush
Patient problems: Chest Pain; Discomfort
MAUDE report 50032761 ↗MalfunctionApr 6, 2026Fracture; Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE POWERPORT CLEARVUE ISP. DURING THE PROCEDURE, IT WAS NOTED THAT THE ATTACHED HUBER NEEDLE BROKE OFF. THE PROCEDURE WAS COMPLETED BY ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24787524 ↗MalfunctionDec 19, 2025Disconnection; Fracture; Material Separation; Detachment of Device or Device Component; Disconnection; Fracture; Material Separation; Detachment of Device or Device Component
A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE POWERPORT CLEARVUE ISP. POST THE PROCEDURE ON (B)(6) 2025, IT WAS NOTED THAT THE CATHETER DISCONNECTED FROM THE PORT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23847555 ↗MalfunctionDec 10, 2025Fracture; Material Separation; Fracture; Material Separation
ON (B)(6) 2025, IT WAS REPORTED THAT DURING THE CV PORT PLACEMENT PROCEDURE USING THE POWERPORT CLEARVUE ISP, UPON OPENING THE KIT, THE SYRINGE PLUNGER BOTTOM WAS FOUND BROKEN. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23764954 ↗MalfunctionDec 8, 2025Disconnection; Detachment of Device or Device Component; Disconnection; Detachment of Device or Device Component
ON (B)(6) A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F INTO THE RIGHT SUBCLAVIAN VEIN. ON (B)(6) 2025, IT WAS REPORTED THAT THE PORT BODY AND CATHETER CONNECTION WAS ALLEGEDLY FOUND DETACHED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23739455 ↗InjuryNov 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME LATER PORT PLACEMENT PROCEDURE, THE PATIENT DEVELOPED WITH UNSPECIFIED INFECTION. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 23655491 ↗MalfunctionOct 8, 2025Fluid/Blood Leak; Fluid/Blood Leak
ON (B)(6) 2025, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE VIA LEFT SUBCLAVIAN VEIN APPROACH USING THE POWERPORT CLEARVUE ISP. ON (B)(6) 2025, SUBCUTANEOUS LEAKAGE WAS FOUND. THE PORT WAS REMOVED ON(B)(6) 2025 AND WAS REPLACED WITH A NEW PORT VIA RIGHT SUBCLAVIAN APPROACH. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23242644 ↗MalfunctionSep 25, 2025Component Missing; Device Markings/Labelling Problem; Missing Information; Component Missing; Device Markings/Labelling Problem; Missing Information
ON (B)(6) 2025, A PATIENT WAS PREPPED FOR A PORT PLACEMENT PROCEDURE USING THE POWER PORT CLEAR VUE ISP. PRIOR TO THE PROCEDURE, THE LABEL WAS ALLEGEDLY FOUND TO BE REMOVED WHEN THE ITEM WAS DELIVERED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23141401 ↗MalfunctionSep 11, 2025Entrapment of Device; Fracture; Material Separation; Deformation Due to Compressive Stress; Entrapment of Device; Fracture; Material Separation; Deformation Due to Compressive Stress
ON (B)(6) 2025, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE INVOLVING THE INSERTION OF A DILATOR OVER A 0.035-INCH GUIDEWIRE USING THE POWERPORT CLEARVUE ISP. DURING THE PROCEDURE, IT WAS REPORTED THAT THE DILATOR BECAME CAUGHT ON THE SKIN, RESULTING IN A PARTIAL LATERAL TEAR OF THE OUTER SHEATH. ADDITIONALLY, THE TIP OF THE SHEATH BENT OVER DURING USE, THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23023421 ↗InjurySep 10, 2025Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
ON (B)(6) 2024, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE POWERPORT CLEARVUE ISP. 11 MONTHS AND 6 DAYS POST PROCEDURE, THE CATHETER WAS ALLEGEDLY FOUND RAPTURED. REPORTEDLY, THE CATHETER ALLEGEDLY FOUND LEAKING. THE DEVICE WAS REMOVED BY INTERVENTIONAL RADIOLOGY. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 23013188 ↗InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME LATER PORT PLACEMENT PROCEDURE, THE PATIENT DEVELOPED WITH UNSPECIFIED INFECTION. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 22661993 ↗MalfunctionJul 11, 2025Fluid/Blood Leak; Fluid/Blood Leak
ON (B)(6) 2024, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE LOW ARTIFACT POWER PORT. ON (B)(6) 2025 THE PORT ALLEGEDLY LEAKED IN THE MIDDLE OF THE CATHETER. PORT WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22478043 ↗MalfunctionJun 24, 2025Fracture; Material Separation; Stretched; Unraveled Material; Failure to Advance; Deformation Due to Compressive Stress; Fracture; Material Separation; Stretched; Unraveled Material; Failure to Advance; Deformation Due to Compressive Stress
ON (B)(6) 2025, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE LOW ARTIFACT POWER PORT. DURING THE PROCEDURE WHEN GUIDE WIRE WAS INSERTED INTO A PUNCTURE NEEDLE, THE TIP ALLEGEDLY BROKE OFF, CREATING A LOOP IN THE NEEDLE'S LUMEN AND PREVENTING IT FROM PASSING THROUGH THE NEEDLE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22322147 ↗MalfunctionJun 24, 2025Fluid/Blood Leak; Deformation Due to Compressive Stress; Fluid/Blood Leak; Deformation Due to Compressive Stress
ON (B)(6) 2025, A PATIENT UNDERWENT FOR A PORT PLACEMENT USING POWERPORT CLEARVUE ISP IMPLANTABLE PORT,CHRONOFLEX SINGLE-LUMEN, 6F. SOMETIMES POST A PORT PROCEDURE THE DEVICE ALLEGEDLY FOUND LEAKAGE. IT WAS FURTHER REPORTED THAT THE CATHETER ALLEGEDLY FOUND TO BE KINKED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22313507 ↗InjuryJun 20, 2025Disconnection; Fracture; Material Separation; Migration; Disconnection; Fracture; Material Separation; Migration
ON (B)(6) 2025, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE LOW ARTIFACT POWER PORT TO SECURE THE ROUTE FOR BLOOD COLLECTION. ON MAY 15TH, THE CATHETER WAS FOUND TO HAVE COME OFF AND FLOWN INTO THE RIGHT VENTRICLE BY CT. HOWEVER, THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22294950 ↗InjuryJun 20, 2025Improper or Incorrect Procedure or Method; Suction Problem; Improper or Incorrect Procedure or Method; Suction Problem
ON (B)(6), 2025, A PATIENT UNDERWENT A PORT PLACEMENT PROCEDURE USING THE LOW ARTIFACT POWER PORT TO SECURE THE ROUTE FOR BLOOD COLLECTION. IT WAS LEFT FOR TWO MONTHS WITHOUT FLUSHING. ON (B)(6), WHEN BLOOD WAS COLLECTED, THE BLOOD DID NOT DRAIN AND THERE WAS SWELLING AROUND THE AREA. HOWEVER, THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Swelling/ Edema
MAUDE report 22294830 ↗InjuryMar 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT TEN DAYS POST PORT PLACEMENT, THE PATIENT ALLEGEDLY HAD FEVER AND REDNESS. IT WAS FURTHER REPORTED THAT THE METHICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS (MRSA) WAS DETECTED AT THE CATHETER TIP. REPORTEDLY, THE PORT WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: Bacterial Infection; Fever
MAUDE report 21706038 ↗MalfunctionDec 6, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, THE DEVICE WAS ALLEGEDLY BROKEN WHEN IT WAS ABOUT TO SUCK THE SALINE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20861849 ↗MalfunctionNov 27, 2024Detachment of Device or Device Component; Packaging Problem; Physical Resistance/Sticking; Detachment of Device or Device Component; Packaging Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, THE FLUSHING CONNECTOR WAS ATTACHED TO THE CUSHIONING MATERIAL, SO WHEN PICKED UP THE CUSHIONING MATERIAL, THE FLUSHING CONNECTOR ALSO LIFTED UP AND FELL OFF. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20793906 ↗InjuryNov 14, 2024Device Tipped Over; Device Tipped Over
IT WAS REPORTED THAT SOMETIME POST A PORT PLACEMENT, THE REGULAR CT SCAN REVEALED THAT PORT WAS ALLEGEDLY ROTATED ONCE IN THE SUPERIOR VENA CAVA. IT WAS FURTHER REPORTED THAT THERE WAS A CONSIDERABLE AMOUNT OF BLOOD CLOT ON THE TIP END POINT OF THE CATHETER AND ANTICOAGULATION WAS PERFORMED AND IT WAS REMOVED. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Thrombosis/Thrombus
MAUDE report 20681434 ↗MalfunctionOct 14, 2024Fracture; Material Separation; Difficult or Delayed Separation; Fracture; Material Separation; Difficult or Delayed Separation
IT WAS REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, AFTER INSERTING THE CATHETER INTO THE SHEATH TO ATTEMPT TEAR, THE SHEATH ALLEGEDLY COULD NOT BE TORN SUCCESSFULLY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20439250 ↗InjurySep 30, 2024Expulsion; Expulsion
IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME LATER POST A PORT PLACEMENT, THE PATIENT ALLEGEDLY DEVELOPED WITH INFECTION AND SKIN EROSION AT THE PORT SITE. HOWEVER, THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Skin Erosion; Unspecified Infection
MAUDE report 20331006 ↗MalfunctionAug 28, 2024Fracture; Fitting Problem; Defective Component; Fracture; Fitting Problem; Defective Component
IT WAS REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, THE STEM ALLEGEDLY BROKE WHILE ATTEMPTING TO CONNECT IT TO THE CATHETER. IT WAS FURTHER REPORTED THAT THE CATHETER ALLEGEDLY DID NOT CONNECT TO THE STEM PROPERLY, AND IT ALLEGEDLY BROKE WHEN TOUCHED IT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20092946 ↗MalfunctionAug 12, 2024Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT DURING PREPARATION FOR A PORT PLACEMENT PROCEDURE, THE PORT WAS ALLEGEDLY NOT ABLE TO PASS WATER AS IT WAS FOUND TO BE CLOGGED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19961493 ↗MalfunctionAug 12, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT TWENTY-FIVE DAYS POST A PORT PLACEMENT, THE FLUID ALLEGEDLY LEAKED SUBCUTANEOUSLY EVEN AFTER BEING PUNCTURED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19960261 ↗
Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices
- PowerPort M.R.I., Groshong 9808560Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR3 deaths834 reports
- PowerPort ClearVUE 1608062Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR18 deaths571 reports
- PowerPort ClearVUE 5608062Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR31 deaths542 reports
- PowerPort M.R.I., ChronoFlex 1808000Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR19 deaths513 reports
- PowerPort M.R.I., ChronoFlex 1808060Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall15 deaths383 reports
- PowerPort M.R.I., Groshong 1808560Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR10 deaths304 reports
- PowerPort M.R.I., ChronoFlex 1829500Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR6 deaths197 reports
- BardPort M.R.I. 0602680Bard Access Systems, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR2 deaths179 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page