Bard® Tiemann Model Urological Catheter Coude Tip 010112Recalls & reported problems
C. R. Bard, Inc. · Bard® Tiemann Model Urological Catheter Coude Tip · Class 2
All models of Bard Tiemann Model Urological Catheter Coude Tip (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00801741029851), received between Nov 14, 2017 and Jul 18, 2026 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Bent” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Bent5
- Difficult to Insert4
- Material Split, Cut or Torn2
- Break1
- Device Operates Differently Than Expected1
- Failure to Advance1
- Material Deformation1
- Material Protrusion/Extrusion1
Patient problems
- No Known Impact Or Consequence To Patient8
- No Clinical Signs, Symptoms or Conditions3
- Sepsis1
- Skin Inflammation/ Irritation1
- Unspecified Infection1
- Urinary Tract Infection1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2026Material Protrusion/Extrusion
Patient problems: Skin Inflammation/ Irritation
MAUDE report 25867180 ↗InjuryJul 24, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE USER USED EXPIRED COUDE CATHETER AND SHORTLY AFTER THE USE GOT A URINARY TRACT INFECTION. COUDE CATHETER WAS EXPIRED IN 2022 AND WAS USED BY (B)(6)2025. REPRESENTATIVE INFORMED PATIENT THAT THEY DID NOT HAVE AN EXPIRATION EXTENSION LETTER AND THEY DID NOT RECOMMEND THE USE OF EXPIRED PRODUCTS.. THEY SUGGESTED TO REACH OUT TO HEALTH CARE PROFESSIONAL AS BD COULDN'T IDENTIFY THE CAUSE OF UTI.
Patient problems: Sepsis; Unspecified Infection; Urinary Tract Infection
MAUDE report 22589515 ↗MalfunctionJul 23, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE PATIENT ARRIVED FOR BACILLE CALMETTE-GUÉRIN TREATMENT AND WAS PREPPED AND USED A BARD #12 FRENCH UROLOGICAL CATHETER TIEMANN MODEL, COUDE TIP INSERTED BLADDER WAS DRAINED, BCG PLACED TO THE HUB OF THE CATHETER AND BEGAN TO INFUSE. A PIN HOLE NOTED AT THE HUB OF THE CATHETER. IMMEDIATELY STOPPED TREATMENT. SECOND CATHETER OPENED, NOTED PIN HOLE, NOT USED. OPENED BARD #14 COUDE TIP, NO PIN HOLE NOTED. UNDER STERILE TECHNIQUE, A CATHETER INSERTED AND TREATMENT GIVEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22576225 ↗MalfunctionJun 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PATIENT ARRIVED FOR THEIR BCG TREATMENT WAS PREPPED AND USED BARD #12 FRENCH UROLOGICAL CATHETER TIEMANN MODEL COUDE TIP WAS INSERTED INTO BLADDER WAS DRAINED, BCG PLACE THE HUB OF THE CATHETER AND BEGAN TO INFUSE, PIN HOLE NOTED AT THE HUB OF THE INTERMITTENT CATHETER. IMMEDIATELY STOPPED TREATMENT. SECOND CATHETER OPENED NOTED PIN HOLE, NOT USED. OPENED BARD #14 COUDE TIP NO PIN HOLE NOTED. UNDER STERILE TECHNIQUE CATHETER INSERTED AND TREATMENT GIVEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22293187 ↗MalfunctionJun 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PATIENT ARRIVED FOR THEIR BCG TREATMENT WAS PREPPED AND USED BARD #12 FRENCH UROLOGICAL CATHETER TIEMANN MODEL COUDE TIP WAS INSERTED INTO BLADDER WAS DRAINED, BCG PLACE THE HUB OF THE CATHETER AND BEGAN TO INFUSE, PIN HOLE NOTED AT THE HUB OF THE INTERMITTENT CATHETER. IMMEDIATELY STOPPED TREATMENT. SECOND CATHETER OPENED NOTED PIN HOLE, NOT USED. OPENED BARD #14 COUDE TIP NO PIN HOLE NOTED. UNDER STERILE TECHNIQUE CATHETER INSERTED AND TREATMENT GIVEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22293034 ↗MalfunctionNov 16, 2017Bent; Difficult to Insert
IT WAS REPORTED THAT THE COUDE TIP WAS BENT MORE THAN USUAL; THEREFORE, THE PATIENT COULD NOT INSERT THE CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7038647 ↗MalfunctionNov 16, 2017Bent; Failure to Advance
IT WAS REPORTED THAT THE COUDE TIP WAS BENT MORE THAN USUAL; THEREFORE, THE PATIENT COULD NOT INSERT THE CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7038339 ↗MalfunctionNov 16, 2017Bent; Difficult to Insert
IT WAS REPORTED THAT THE COUDE TIP WAS BENT MORE THAN USUAL; THEREFORE, THE PATIENT COULD NOT INSERT THE CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7038234 ↗MalfunctionNov 16, 2017Break; Difficult to Insert
IT WAS REPORTED THAT THE COUDE TIP WAS BENT MORE THAN USUAL; THEREFORE, THE PATIENT COULD NOT INSERT THE CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7038216 ↗MalfunctionNov 16, 2017Material Deformation
IT WAS REPORTED THAT THE COUDE TIP WAS BENT MORE THAN USUAL, THEREFORE, THE PATIENT COULD NOT INSERT THE CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7036767 ↗MalfunctionNov 16, 2017Bent; Difficult to Insert
IT WAS REPORTED THAT THE COUDE TIP WAS BENT MORE THAN USUAL; THEREFORE, THE PATIENT COULD NOT INSERT THE CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7036626 ↗MalfunctionNov 16, 2017Bent
IT WAS REPORTED THAT THE COUDE TIP WAS BENT MORE THAN USUAL; THEREFORE, THE PATIENT COULD NOT INSERT THE CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7036446 ↗MalfunctionNov 14, 2017Device Operates Differently Than Expected
IT WAS REPORTED THAT THE CATHETER WAS CURLED MAKING IT DIFFICULT TO INSERT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7028470 ↗
Other CATHETER, UROLOGICAL devices
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50614C. R. Bard, Inc. · CATHETER, UROLOGICAL883 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53614GC. R. Bard, Inc. · CATHETER, UROLOGICAL692 reports
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50616C. R. Bard, Inc. · CATHETER, UROLOGICAL2 deaths443 reports
- GENTLECATH GLIDE 421567Convatec, Purchasing Department · Catheter, Urological342 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53616GC. R. Bard, Inc. · CATHETER, UROLOGICAL301 reports
- Bardex® Foley Catheter Silicone Coated 0167V22SC. R. Bard, Inc. · CATHETER, UROLOGICAL187 reports
- Bardex® Foley Catheter Silicone Coated 0167V20SC. R. Bard, Inc. · CATHETER, UROLOGICAL175 reports
- Bardia® Coude Olive Tip Urethral Foley Catheter 802516C. R. Bard, Inc. · CATHETER, UROLOGICAL170 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page