Bard® Touchless® Plus Unisex Pre-Lubricated Rubber Urethral Catheter Kit 4A7044Recalls & reported problems
C. R. Bard, Inc. · Bard® Touchless® Plus Unisex Pre-Lubricated Urethral Catheter Kit, Rubber Urethral Catheter · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is a Bard Touchless Plus pre-lubricated rubber urethral catheter kit, a tray used to drain the bladder. Most of the 54 reports describe catheters arriving with no lubricant or not enough lubricant, tips folded or bent over so they could not be inserted, missing parts such as the eyelet, and units that were damaged or misassembled in the box. Four reports mention injuries, including urinary tract infections, and there are no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 00801741030536), received between Sep 24, 2017 and Apr 25, 2024 — classified as 4 injuries, 50 malfunctions.
The most frequently reported specific device problem is “Product Quality Problem” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Product Quality Problem14
- Nonstandard Device10
- Incorrect Measurement7
- Material Integrity Problem7
- Difficult to Insert6
- Misassembled5
- Component Missing3
- Improper Chemical Reaction3
Patient problems
- No Consequences Or Impact To Patient32
- No Known Impact Or Consequence To Patient29
- No Clinical Signs, Symptoms or Conditions13
- Urinary Tract Infection3
- No Information2
- Pain2
- Unspecified Infection2
- Blood Loss1
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2024Improper Chemical Reaction; Improper Chemical Reaction
IT WAS REPORTED THAT SOME OF THE PRE- LUBRICATED RUBBER URETHRAL CATHETERS WERE LUBRICATED, AND SOME WERE NOT. PER CUSTOMER ON (B)(6) 2024, IT WAS REPORTED THAT SOME OF THE CATHETERS WERE NOT LUBRICATED ENOUGH TO BE USED PROPERLY. CUSTOMER HAS SWITCHED TO A DIFFERENT COMPANY WITH CATHETERS THAT WORK BETTER FOR THEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19180524 ↗MalfunctionFeb 26, 2024Component Misassembled; Component Misassembled
IT WAS REPORTED THAT SOME OF THE PRE- LUBRICATED RUBBER URETHRAL CATHETERS WERE LUBRICATED, AND SOME WERE NOT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18776650 ↗MalfunctionAug 18, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT PRE LUBRICATED RUBBER URETHRAL CATHETER SUPPLIED WERE DEFECTIVE AND UNABLE TO USE. CUSTOMER STATED THAT THE TIPS WERE DEFECTIVE AND UNABLE TO INSERT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15250672 ↗MalfunctionOct 28, 2021Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT ONE TOUCHLESS PLUS COUDE TIP INTERMITTENT CATHETER FROM A BOX OF 50 HAD NO LUBE IN IT, IT WAS NOT LIKE THE LUBE WAS DRIED OUT AND IT WAS THAT NONE WAS ADDED (LOT# NGFS1953). IT WAS STATED THAT 7 EACH FROM A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFT2379). IT WAS ALSO STATED THAT 8 EACH FROM A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFQ1655). PER ADDITIONAL INFORMATION ON (B)(6)2021, IT WAS STATED THAT 2 EACH FROM OF A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFR2863).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12712108 ↗MalfunctionOct 11, 2021Improper Chemical Reaction; Improper Chemical Reaction
IT WAS REPORTED THAT ONE TOUCHLESS PLUS COUDE TIP INTERMITTENT CATHETER FROM A BOX OF 50 HAD NO LUBE IN IT, IT WAS NOT LIKE THE LUBE WAS DRIED OUT AND IT WAS THAT NONE WAS ADDED (LOT# NGFS1953). IT WAS STATED THAT 7 EACH FROM A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFT2379). IT WAS ALSO STATED THAT 8 EACH FROM A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFQ1655). PER ADDITIONAL INFORMATION ON (B)(6) 2021, IT WAS STATED THAT 2 EACH FROM OF A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFR2863).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605572 ↗MalfunctionOct 11, 2021Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT ONE TOUCHLESS PLUS COUDE TIP INTERMITTENT CATHETER FROM A BOX OF 50 HAD NO LUBE IN IT. IT WAS NOT LIKE THE LUBE WAS DRIED OUT AND IT WAS THAT NONE WAS ADDED ( (LOT: NGFS1953). IT WAS STATED THAT 7 EACH FROM A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT: NGFT2379). IT WAS ALSO STATED THAT 8 EACH FROM OF A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT: NGFQ1655). PER ADDITIONAL INFORMATION ON 07OCT2021, IT WAS STATED THAT 2 EACH FROM OF A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT: NGFR2863).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605485 ↗MalfunctionOct 11, 2021Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT ONE TOUCHLESS PLUS COUDE TIP INTERMITTENT CATHETER FROM A BOX OF 50 HAD NO LUBE IN IT, ALSO IT WAS NOT LIKE THE LUBE WAS DRIED OUT AND IT WAS THAT NONE WAS ADDED (LOT# NGFS1953). IT WAS STATED THAT 7 EACH FROM A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFT2379). IT WAS ALSO STATED THAT 8 EACH FROM A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFQ1655). PER ADDITIONAL INFORMATION ON (B)(6)2021, IT WAS STATED THAT 2 EACH FROM OF A BOX OF 50 HAD THE WHOLE TIP CURVED OVER OR FOLDED IN HALF AND IT WAS UNUSABLE (LOT# NGFR2863).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605478 ↗MalfunctionJul 1, 2021Component Misassembled; Component Misassembled
IT WAS REPORTED THAT THE RECEIVED SHIPMENT BOX OF THE CATHETER WAS DAMAGED AND THE TOP OF THE BOX FOLDED SHUT AND THE TAPE WAS MISSING. THE CATHETER PRESENT IN THE SHIPMENT WAS ALSO DAMAGED BUT THE CUSTOMER FELT LIKE THEY WERE INTACT TO USE BUT FOUND THAT THE LOT OF CATHETERS WERE DEFECTIVE AND THE EYELET WAS ALSO MISSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12099555 ↗MalfunctionJun 30, 2021Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT THE INTERMITTENT COUDE CATHETERS WERE NOT CURVED ENOUGH. ALSO STATED THAT THE COUDE WAS EITHER NOT DEFINED ENOUGH OR NOT THERE AT ALL. PER FOLLOW UP ON 16JUN2021 IT WAS STATED THAT THE COUDE CATHETERS WORKS GREAT BUT 90 PERCENTAGE OF THEM DID NOT HAVE NEARLY ENOUGH COUDE TO BE EFFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12093132 ↗MalfunctionJun 30, 2021Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT THE INTERMITTENT COUDE CATHETERS WERE NOT CURVED ENOUGH. ALSO STATED THAT THE COUDE WAS EITHER NOT DEFINED ENOUGH OR NOT THERE AT ALL. PER FOLLOW UP ON (B)(6) 2021 IT WAS STATED THAT THE COUDE CATHETERS WORK THEY WERE GREAT BUT 90 PERCENTAGE OF THEM DID NOT HAVE NEARLY ENOUGH COUDE TO BE EFFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12092950 ↗MalfunctionMar 17, 2021Nonstandard Device; Material Twisted/Bent; Nonstandard Device; Material Twisted/Bent
IT WAS REPORTED THAT THE COUDE WAS CURVED AND SOME JUST HAD A SLIGHT CURVE. FOLLOW UP VIA PHONE ON 26FEB2021, WHEN THE CATHETERS WEREN¿T CURVED ENOUGH, THE CUSTOMER HAD A HARD TIME INSERTING AND USING THEM. PER FOLLOW UP ON 02JUN2021, CUSTOMER STATED THAT THE COUDE TOUCHLESS CATHETER KITS HAVE VIRTUALLY NO COUDE. AS A PARAPLEGIC, THESE KITS COULD BE INVALUABLE IF THEY WERE MORE USER FRIENDLY. WHEN THE CATHETER HAD A WELL DEFINED COUDE AT LEAST 45 DEGREES, THEY WORK ALMOST 100 PERCENT OF THE TIME. PER FOLLOW UP ON 08JUN2021, CUSTOMER STATED THAT THE ISSUE WAS NOT WITH ONE PARTICULAR LOT AND IT HAD BEEN MANY LOTS OVER THE YEARS AND THE ISSUE WAS CONSTANT. ALSO ADDED THAT COUDE WAS EITHER NOT DEFINED ENOUGH OR NOT THERE AT ALL. PER FOLLOW UP ON 16JUN2021, IT WAS STATED THAT THE COUDE CATHETERS WORK GREAT, BUT 90 PERCENTAGE OF THEM DO NOT HAVE NEARLY ENOUGH COUDE TO BE EFFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11510120 ↗MalfunctionDec 22, 2020Nonstandard Device; Nonstandard Device
IT WAS REPORTED THAT THERE WAS HARDLY ANY COUDE AT THE TIP OF THE CATHETER. PER FOLLOW UP VIA PHONE ON (B)(6) 2020 THE CUSTOMER NOTED THAT THE CATHETERS WERE NOT CURVED ENOUGH THEY SEEM STRAIGHT. PER FOLLOW UP ON (B)(6) 2021 THE CUSTOMER STATED THAT THE COUDE TOUCHLESS CATHETER KITS HAVE VIRTUALLY NO COUDE. AS A PARAPLEGIC THESE KITS COULD BE INVALUABLE IF THEY WERE MORE USER FRIENDLY. WHEN THE CATHETER HAD A WELL DEFINED COUDE AT LEAST 45 DEGREES. PER FOLLOW UP ON (B)(6) 2021 THE CUSTOMER STATED THAT THE ISSUE WAS NOT WITH ONE PARTICULAR LOT AND WITH MANY LOTS OVER THE YEARS AND THE ISSUE WAS CONSTANT. ALSO STATED THAT THE COUDE WAS EITHER NOT DEFINED ENOUGH OR NOT THERE AT ALL. PER FOLLOW UP ON (B)(6) 2021 IT WAS STATED THAT THE COUDE CATHETERS WORK THEY ARE GREAT BUT 90 PERCENTAGE OF THEM DID NOT HAVE NEARLY ENOUGH COUDE TO BE EFFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11060206 ↗MalfunctionDec 15, 2020Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE PATIENT HAD NOTICED FOR SOME TIME NOW THAT THE MANUFACTURE WAS CUTTING THE CATHETERS TOO SHORT, THEY WERE SHORTER THAN THAT THEY USED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11013352 ↗MalfunctionJul 30, 2020Inaccurate Flow Rate; Obstruction of Flow; Inaccurate Flow Rate
IT WAS REPORTED THAT THE PATIENT WAS HALF WAY THROUGH A BOX OF 50 TOUCHLESS PLUS CATHETER AND HAD 2 THAT DID NOT ALLOW FLUID TO PASS.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 10345984 ↗InjuryJul 16, 2020Biocompatibility; Adverse Event Without Identified Device or Use Problem; Biocompatibility
IT WAS REPORTED THAT THE PATIENT HAD A URINARY TRACT INFECTION AND THE CAUSE WAS UNKNOWN. IT IS UNKNOWN WHAT TREATMENT WAS PROVIDED FOR THE UTI.
Patient problems: Urinary Tract Infection
MAUDE report 10283219 ↗MalfunctionJul 16, 2020Improper Chemical Reaction; Material Integrity Problem; Improper Chemical Reaction
IT WAS REPORTED THAT THE CATHETERS DID NOT HAD ENOUGH LUBRICANT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 10282579 ↗MalfunctionMay 20, 2020Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THE COUDE TIP WERE NOT CURVED ENOUGH. IT WAS NOTED THAT ALL 4 CATHETERS WERE FOUND TO NOT HAVING COUDE INDICATOR BUTTON.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10074229 ↗MalfunctionApr 2, 2020Product Quality Problem; Material Integrity Problem; Product Quality Problem
IT WAS REPORTED THAT THE CATHETER LACKED LUBRICATION INSIDE OF THE GEL PACKAGE. THE PATIENT STATED SOME PACKAGES ARRIVED EMPTY, AND OTHERS DID NOT HAVE ENOUGH LUBRICATION WHICH MADE IT UNCOMFORTABLE TO INSERT THE CATHETER. NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9918135 ↗MalfunctionMar 18, 2020Nonstandard Device; Material Deformation; Nonstandard Device
IT WAS REPORTED THAT THE COUDE WASN'T CURVED ENOUGH. NO MEDICAL INTERVENTION.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9848168 ↗MalfunctionDec 27, 2019Misassembled; Misassembled; Component Missing
IT WAS REPORTED THAT THE COUDE INDICATORS WERE MISSING FROM THE CATHETERS.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 9525309 ↗MalfunctionDec 18, 2019Product Quality Problem; Product Quality Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE CATHETER HAD DRIED LUBRICANT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9491664 ↗MalfunctionNov 26, 2019Incorrect Measurement; Incorrect Measurement; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THE COUDE TOUCHLESS PLUS CATHETER WAS 4 INCHES SHORTER THAN ANOTHER TOUCHLESS CATHETER.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9379177 ↗MalfunctionNov 13, 2019Misassembled; Misassembled; Material Integrity Problem
IT WAS REPORTED THAT THERE WAS NO BUMP AT THE FUNNEL END OF THE COUDE CATHETER TO INDICATE THAT IT IS COUDE BUT THE CATHETER DID HAVE A CURVE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9314619 ↗MalfunctionNov 7, 2019Misassembled; Misassembled; Material Deformation
IT WAS REPORTED THAT THERE WAS NO BUMP AT THE FUNNEL END OF THE COUDE CATHETER TO INDICATE THAT IT IS COUDE BUT THE CATHETER DID HAVE A CURVE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9293101 ↗MalfunctionOct 24, 2019Difficult to Insert; Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE CATHETER APPEARED SHORTER THAN NORMAL CAUSING A DIFFICULT INSERTION PROCESS. THE PATIENT WAS UNABLE TO USE THE CATHETER. NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9234135 ↗
Other Tray, catheterization, sterile urethral, with or without catheter (kit) devices
- Bard® Touchless® Plus Unisex Pre-Lubricated Vinyl Urethral Catheter Kit 4A5144C. R. Bard, Inc. · Tray, catheterization, sterile urethral, with or without catheter (kit)170 reports
- Dover 3450LFCardinal Health, Inc. · Tray, catheterization, sterile urethral, with or without catheter (kit)1 recall3 reports
- Dover 3450Cardinal Health, Inc. · Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)1 recall1 report
- Dover 3410Cardinal Health, Inc. · Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)1 recall0 reports
- Dover 8887600784Cardinal Health, Inc. · Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)1 recall0 reports
- Dover 8887600776Cardinal Health, Inc. · Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)1 recall0 reports
- Dover 3408Cardinal Health 200, LLC · Tray, catheterization, sterile urethral, with or without catheter (kit)1 recall0 reports
- Dover 2480Cardinal Health, Inc. · Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page