Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741032417), received between Jan 16, 2017 and Sep 23, 2026 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image3
  • Adverse Event Without Identified Device or Use Problem2
  • Therapeutic or Diagnostic Output Failure2
  • Break1
  • Patient-Device Incompatibility1
  • Positioning Failure1
  • Premature Activation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Hypersensitivity/Allergic reaction1
  • Reaction1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjurySep 23, 2026Adverse Event Without Identified Device or Use Problem

    A patient underwent an ultrasound guided marker placement procedure using the UltraClip. Post procedure, the patient reported being recently diagnosed with a nickel allergy and expressed concern that the implanted clips may have contributed to the rashes and other symptoms experienced over the past few years. The current status of the patient was unknown.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50866574 ↗
  • MalfunctionMay 20, 2026Break; Positioning Failure

    ON (B)(6) 2026, A PATIENT UNDERWENT A BREAST TISSUE MARKER PLACEMENT PROCEDURE USING THE ULTRACLIP DUAL TRIGGER BREAST TISSUE MARKER. DURING THE PROCEDURE, A FRAGMENT CAME OFF THE CLIP. REPORTEDLY, THERE WAS AN ISSUE WITH MARKER DEPLOYMENT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25226208 ↗
  • MalfunctionMar 4, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT DURING AN ULTRASOUND GUIDED BREAST MARKER PLACEMENT PROCEDURE, THE CLIP WAS ALLEGEDLY NOT DEPLOYED. IT WAS FURTHER REPORTED THAT THE CLIP WAS ALLEGEDLY NOT NOTED WHEN THE TECH TOOK THE POST MAMMO IMAGES. NO COAXIAL WAS USED AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18828191 ↗
  • MalfunctionSep 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY FOUR YEARS POST ULTRASOUND GUIDED BREAST TISSUE MARKER PLACEMENT, THE PATIENT ALLEGEDLY EXPERIENCED ALLERGY FLARED UP WITH BURNING SENSATIONS ON HANDS AND ITCHY RASHES DEVELOPED ON SEVERAL PARTS OF BODY. REPORTEDLY, DAILY STEROID CREAMS WERE USED BY THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: Hypersensitivity/Allergic reaction; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17787193 ↗
  • MalfunctionMar 20, 2023Premature Activation; Premature Activation

    IT WAS REPORTED THAT DURING A BREAST TISSUE MARKER PLACEMENT PROCEDURE THROUGH, THE MARKER ALLEGEDLY DEPLOYED PREMATURELY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16579134 ↗
  • MalfunctionDec 1, 2021Poor Quality Image; Therapeutic or Diagnostic Output Failure; Poor Quality Image; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT POST BREAST MARKER PLACEMENT PROCEDURE, THE MARKER WAS NOT VISIBLE UNDER MAMMOGRAPHY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12908320 ↗
  • MalfunctionNov 30, 2021Poor Quality Image; Therapeutic or Diagnostic Output Failure; Poor Quality Image; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT POST BREAST TISSUE MARKER PLACEMENT THROUGH HETEROGENEOUSLY DENSITY TISSUE UNDER MAMMOGRAPHY, THE MARKER WAS ALLEGEDLY NOT VISIBLE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12891065 ↗
  • InjuryJan 16, 2017Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT FIVE DAYS POST BREAST TISSUE MARKER PLACEMENT, THE PATIENT ALLEGED A REACTION TO THE MARKER PLACED. IT WAS FURTHER REPORTED THAT THE PATIENT HAD FORMED A RASH ALL OVER THE BODY AND PRESENTED THE REACTION TO THE DOCTOR ELEVEN DAYS AFTER THE PROCEDURE. REPORTEDLY, THIRTEEN DAYS AFTER THE PROCEDURE THE PATIENT WENT TO THE EMERGENCY ROOM AS THE RASH HAS NOT DISSIPATED. THE PATIENT HAD STOPPED TAKING ALL MEDICATIONS AND WAS TAKING ALLERGY MEDICATION TO HELP STOP THE RASH, HOWEVER IT IS STILL PRESENT. IT WAS REPORTED THAT THE PATIENT HAS SCHEDULED AN APPOINTMENT WITH AN ALLERGIST.