Skip to content

Bard® All Purpose Urethral Catheter 0094120Recalls & reported problems

C. R. Bard, Inc. · Bard® All Purpose Urethral Catheter · Class 2

All models of Bard All Purpose Urethral Catheter (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741034015), received between Sep 16, 2020 and Feb 19, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Contamination /Decontamination Problem1
  • Material Split, Cut or Torn1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THEIR CLINICAL STAFF STATED THAT THEY NOTICED CRACKING IN GENERAL PURPOSE LATEX INTERMITTENT CATHETERS [PCN #277710(BATCH #UNK), (BATCH #NGFS3051), (BATCH #NGHV0932), (BATCH #NGGR2390, (BATCH #NGGQ3730], [PCN #277708 (BATCH #NGHS3453), (BATCH #NGGZ0495), (BATCH #NGGT3524), (BATCH #NGGY3130), (BATCH #NGGS2682], [PCN #94100 (BATCH #NGEY3152], ]PCN #94120(BATCH #NGEV4510], [PCN #56110(BATCH #NGFW2958], [PCN #56114(BATCH #NGGV2774]. THEY WERE CONCERNED THAT THEY WERE BRITTLE AND COULD AFFECT PATIENT SAFETY. BASED OFF OF THEIR CONCERNS, THEY SWEPT THEIR SUPPLY AREAS AND FOUND THE CATHETERS BELOW TO SHOW THE SAME CHARACTERISTICS. THEY WOULD LIKE TO HAVE THOSE EVALUATED. THEY HAVE ALSO INCLUDED A PHOTO BELOW, WHICH SHOWS THE CRACKING. PER CUSTOMER FOLLOW UP INFORMATION RECEIVED VIA EMAIL ON 05MAR2025, STATED THAT THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21409856 ↗
  • MalfunctionSep 16, 2020Break; Contamination /Decontamination Problem; Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE CATHETERS HAD AN ALLEGED WHITE POWDER, FILM OR RESIDUE ON THEM AND WERE NOT USED. AS PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 25AUG2020 FROM THE INVESTIGATOR, THE CATHETER WAS FOUND TO HAVE OZONE DAMAGE DURING THE EVALUATION.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 10541630 ↗

Other CATHETER, UROLOGICAL devices

All CATHETER, UROLOGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page