Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814Recalls & reported problems

C. R. Bard, Inc. · Bardex® Lubri-Sil® All Silicone Lubricious Coated 5 cc Foley Catheter · Class 2

All models of Bardex Lubri-Sil All-Silicone Foley Catheter (21)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Bard Lubri-Sil all-silicone Foley catheter is a retention-type balloon catheter. Its 151 adverse event reports include 5 injuries and no deaths, and many describe holes or tears in the material, tubing splitting, and balloons that burst or would not deflate. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
153
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 153 adverse event reports that list this exact device (DI 00801741034152), received between May 24, 2017 and Sep 28, 2026 — classified as 5 injuries, 148 malfunctions.

The most frequently reported specific device problem is “Material Puncture/Hole” (42 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 8 reportsJun 2024: 2 reportsJul 2024: 5 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 4 reportsJan 2025: 3 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 3 reportsMay 2025: 5 reportsJun 2025: 4 reportsJul 2025: 4 reportsAug 2025: 8 reportsSep 2025: 3 reportsOct 2025: 3 reportsNov 2025: 15 reportsDec 2025: 6 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Puncture/Hole42
  • Difficult to Remove15
  • Material Split, Cut or Torn15
  • Material Fragmentation11
  • Material Rupture11
  • Deflation Problem10
  • Inflation Problem8
  • Fluid/Blood Leak6

Patient problems

  • No Clinical Signs, Symptoms or Conditions119
  • No Consequences Or Impact To Patient19
  • No Known Impact Or Consequence To Patient17
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
  • Pain5
  • Foreign Body In Patient4
  • Discomfort2
  • Hematuria2

Recent adverse event reports

The 25 most recent of 153 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Nonstandard Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923637 ↗
  • MalfunctionSep 24, 2026Material Fragmentation

    It was reported that the customer had a Bard Medical 14Fr Foley tray system catheter (Reference: (b)(4) was placed on the patient for intravesical chemotherapy. When the nurse went to instill the chemotherapy, the needle-free sampling port, where the Equashield connector was to be attached, was noted to be missing or broken, preventing attachment of the Equashield device for chemotherapy administration. The drainage bag and tubing were removed, and a new bag (Medline Add-A-Cath) was attached to the catheter, which had the correct port. The customer stated that the same issue was observed in multiple kits from the same lot number on the same day.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889830 ↗
  • MalfunctionSep 2, 2026Deflation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50563775 ↗
  • MalfunctionMay 6, 2026Material Rupture

    IT WAS REPORTED THAT PATIENT FOLEY CATHETER CAME OUT WHEN THEY WERE ON THE TOILET. WAS NOT PULLED BUT FELL OUT. THERE WAS ABOUT A 5MM U-SHAPED TEAR NOTED IN THE BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096384 ↗
  • MalfunctionMar 18, 2026Material Puncture/Hole

    IT WAS REPORTED THAT BEDSIDE STAFF REPOSITIONED THE PATIENT AFTER A COMPLAINT OF URINARY URGENCY WHILE A FOLEY CATHETER WAS IN PLACE. REPOSITIONING WAS PERFORMED TO ENSURE THE FOLEY CATHETER WAS NOT OCCLUDED, AT WHICH TIME THE CATHETER WAS FOUND DEFLATED AND LYING ON THE BED. FURTHER INSPECTION IDENTIFIED A LEAK AT THE FOLEY CATHETER PORT SITE NEAR THE SALINE PORT. A NEW FOLEY CATHETER WAS PLACED, AND NO PATIENT HARM OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24630402 ↗
  • MalfunctionMar 13, 2026Gel Leak

    IT WAS REPORTED THAT DURING USE WHILE INSERTING A URINARY FOLEY CATHETER INTO THE PATIENT, THE PRE-FILLED STERILE WATER SYRINGE USED FOR INFLATION BURST AND WAS LEAKING WATER OUT OF THE SIDES. THEN REMOVED SYRINGE FROM CATHETER, DREW UP 10ML STERILE WATER TO USE TO INFLATE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24596485 ↗
  • MalfunctionJan 23, 2026Material Rupture

    IT WAS REPORTED THAT, FOLEYS CATHETER INSERTED PRIOR TO TOTAL HYSTERECTOMY AND IT GOT BURST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24172976 ↗
  • MalfunctionJan 9, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THE PATIENT¿S URINARY FOLEY CATHETER WAS FOUND BROKEN, WITH URINE LEAKING FROM THE TUBING. THE PATIENT WAS UNDER A DEPRIVATION OF LIBERTY SAFEGUARDS (DOLS) AND WAS AGITATED AND RESTLESS IN BED. POSEY MITTENS WERE APPLIED TO THE RIGHT ARM AS THE LEFT ARM WAS PLEGIC. THE PATIENT HAD NOT BEEN OBSERVED ACTIVELY PULLING AT THE CATHETER. ON INSPECTION, THERE WAS A CLEAR SPLIT ALONG THE CATHETER TUBING, THOUGH THE STAT LOCK REMAINED IN PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24031311 ↗
  • MalfunctionJan 9, 2026Material Fragmentation

    IT WAS REPORTED THAT THE PATIENT¿S URINARY FOLEY CATHETER WAS FOUND BROKEN, WITH URINE LEAKING FROM THE TUBING. THE PATIENT WAS UNDER A DEPRIVATION OF LIBERTY SAFEGUARDS (DOLS) AND WAS AGITATED AND RESTLESS IN BED. POSEY MITTENS WERE APPLIED TO THE RIGHT ARM AS THE LEFT ARM WAS PLEGIC. THE PATIENT HAD NOT BEEN OBSERVED ACTIVELY PULLING AT THE CATHETER. ON INSPECTION, THERE WAS A CLEAR SPLIT ALONG THE CATHETER TUBING, THOUGH THE STAT LOCK REMAINED IN PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24031075 ↗
  • MalfunctionDec 30, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE BALLOON ON X2 FOLEY CATHETERS FROM THIS LOT NUMBER HAD BURST WHILE INSIDE THE PATIENT CAUSING THE CATHETER TO FALL OUT AND BE NO LONGER USEABLE. THE PATIENT HAS NOT BEEN INJURED BUT OTHER CATHETERS FROM THIS LOT MAY BE DEFECTIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23931065 ↗
  • MalfunctionDec 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT ON INSERTION OF URINARY CATHETER, NOTICED LEAKAGE OF URINE COMING OUT THE ACTUAL CATHETER TUBE. A TEAR WAS FOUND IN THE TUBE (UNABLE TO SEE BEFORE INSERTION AS COVERED IN THE BLUE PACKAGING UPON INSERTION). IMMEDIATELY REMOVED THE CATHETER AND INSERTED A NEW CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23915492 ↗
  • MalfunctionDec 24, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT ON CHECKING PATIENT, BED FOUND SOAKING WET AND THE FOLEY CATHETER FOUND BROKEN INTO 2 PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23889136 ↗
  • MalfunctionDec 19, 2025Unclear Information; Wrong Label; Unclear Information; Wrong Label

    IT WAS REPORTED THAT THEY DISCOVERED THAT ALL OF OUR SUPPLY OF BARD LUBRI-SIL ALL-SILICONE FOLEY CATHETERS SIZES 8FR TO 20FR IN THE STOLLERY OR LEVEL 1 CORE HAVE LABELLING DISCREPANCIES. THE LOT NUMBER ON OUTSIDE OF PACKAGE DID NOT MATCH WITH LOT NUMBER ON PRODUCT INSIDE PACKAGE. THE BALLOON FILL AMOUNTS DO NOT MATCH ON THE 20FR AND 12FR PRODUCTS/PACKAGING. (THE REFERENCE NUMBERS DO APPEAR TO MATCH WITH PRODUCT AND PACKAGING). FOLEY CATHETER SIZE 8FR . FOLEY CATHETER SIZE 10FR . FOLEY CATHETER SIZE 12FR. FOLEY CATHETER SIZE 14FR . FOLEY CATHETER SIZE 16FR . FOLEY CATHETER SIZE 18FR . FOLEY CATHETER SIZE 20FR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23850893 ↗
  • MalfunctionDec 12, 2025Wrong Label; Wrong Label

    IT WAS REPORTED THAT FOLEY CATHETER HAD A DIFFERENT LOT NUMBER THAN THE PACKAGING IT CAME OUT OF. THAT WAS THE FIRST TIME THEY HAD HEARD OF THAT COMPLAINT AND WOULD DO EVERYTHING THEY CAN TO GET TO THE BOTTOM OF THAT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23794672 ↗
  • MalfunctionDec 5, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THERE WAS A HAIR INSIDE THE UN-OPENED STERILE PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23735178 ↗
  • MalfunctionNov 17, 2025Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE REGISTERED NURSE TRIED TO DEFLATE THE BALLOON OF THE FOLEY CATHETER BUT THE FOLEY BALLOON DID NOT DEFLATE ¿ NO SALINE ASPIRATED. NEXT, THIS RN ASPIRATED TWICE, AND AIR FILLED THE SYRINGE. LASTLY, THIS RN TRIED TO DEFLATE THE BALLOON AGAIN AND FINALLY THE BALLOON DEFLATED WITH SALINE ASPIRATED. THE FOLEY CATHETER WAS THEN REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23570316 ↗
  • MalfunctionNov 13, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THE FOLEY CATHETER WAS INSERTED AND WHEN BALLOON WAS INFLATED THERE WAS A PINHOLE LEAK NEAR INSERTION PORT. FOLEY WAS REMOVED AND NEW FOLEY INSERTED WITHOUT ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23542965 ↗
  • MalfunctionNov 12, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT FOLEY CATHETER KEPT DEFLATING DUE TO SMALL HOLE IN THE BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533670 ↗
  • MalfunctionNov 10, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE OPERATING STAFF WAS STATED THERE WAS A HOLE IN THE CATHETER ALONG THE SEAM. THE FOLEY CATHETER BALLOON DID NOT STAY INFLATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23512474 ↗
  • MalfunctionNov 7, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THE FOLEY CATHETER WAS PLACED FOR SCHEDULED CESAREAN SECTION. ATTEMPTED TO INFLATE FOLEY BALLOON, FLUID FROM SYRINGE WENT DIRECTLY TO FOLEY BAG, THUS THE BALLOON COULD NOT BE INFLATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23497745 ↗
  • MalfunctionNov 7, 2025Inflation Problem; Material Puncture/Hole; Inflation Problem; Material Puncture/Hole

    IT WAS REPORTED THAT WHEN INFLATING BALLOON IN FOLEY THE SALINE FLUSH SQUIRTED OUT OF SITE AND DID NOT INFLATE. SECOND CATHETER HAD TO BE PLACED FOR PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23494943 ↗
  • MalfunctionNov 7, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THE CATHETER PLACED AND WHEN CATHETER BALLOON WAS INFLATED, THE FLUID SQUIRTED OUT OF THE Y JUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23494841 ↗
  • MalfunctionNov 7, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THE FOLEY CATHETER CAME OUT DURING A POSITION CHANGE. UPON EVALUATION THE BALLON DID NOT APPEAR TO BE INFLATED. WHEN REINFLATED THE BALLON QUICKLY DEFLATED DUE TO A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23494840 ↗
  • MalfunctionNov 6, 2025Inaccurate Flow Rate; Inaccurate Flow Rate

    IT WAS REPORTED THAT EMERGENCY CESAREAN SECTION. FOLEY CATHETER PLACED IN OPERATING ROOM AND NOT DRAINING URINE. CATHETER DISCONTINUED. DIFFERENT LOT NUMBER FOLEY CATHETER PLACED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23488457 ↗
  • MalfunctionNov 6, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THE RN PLACED FOLEY, WENT TO INFLATE BALLOON AND FLUID SHOT OUT OF THE Y SITE. IT FELT LIKE THERE WAS A HOLE. PRODUCT IS IN MANAGER'S OFFICE. NEW FOLEY HAD TO BE PLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23488454 ↗

Other CATHETER, RETENTION TYPE, BALLOON devices

All CATHETER, RETENTION TYPE, BALLOON devices →