Groshong Catheter 7715405Recalls & reported problems
Bard Access Systems, Inc. · Groshong Catheter 4F Single-Lumen Basic Kit · Class 2 · Implant
All models of Groshong Catheter (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Groshong PICC catheter is a small tube placed through the arm into a large vein for long-term IV access. Most of the 72 reports describe fluid or blood leaking from the catheter, cracks or splits in the tubing, and connectors or tubing that come apart, sometimes leaving a piece of the catheter inside the patient or requiring a replacement line. Twelve reports involve injuries and no deaths are described. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 72
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 72 adverse event reports that list this exact device (DI 00801741035296), received between Feb 4, 2016 and Jun 3, 2026 — classified as 12 injuries, 60 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (40 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak40
- Break12
- Material Deformation4
- Leak/Splash3
- Detachment of Device or Device Component2
- Device Dislodged or Dislocated2
- Failure to Infuse2
- Material Frayed2
Patient problems
- No Clinical Signs, Symptoms or Conditions42
- No Known Impact Or Consequence To Patient8
- Foreign Body In Patient6
- No Consequences Or Impact To Patient5
- Swelling/ Edema5
- No Patient Involvement3
- Extravasation2
- Aneurysm1
Recent adverse event reports
The 25 most recent of 72 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 3, 2026Fluid/Blood Leak
IT WAS REPORTED THAT ON (B)(6) 2025, A PICC CATHETER WAS INSERTED INTO THE PATIENT¿S RIGHT UPPER ARM WITH AN INTERNAL LENGTH OF 40 CM. THE CIRCUMFERENCES OF BOTH THE LEFT AND RIGHT ARMS WERE 23 CM AT THE TIME OF INSERTION. DURING ROUTINE MAINTENANCE ON THE PICC CATHETER, A PERFORATION WAS IDENTIFIED ON THE EXTERNAL PORTION OF THE CATHETER APPROXIMATELY 5 CM FROM THE INSERTION SITE DURING FLUSHING. IN RESPONSE, THE EXTERNAL DAMAGED SEGMENT WAS TRIMMED. FOLLOWING ADJUSTMENT, THE CATHETER REMAINED WITH 40 CM INTERNAL LENGTH AND APPROXIMATELY 3 CM EXTERNAL SEGMENT. THE LINE WAS SUBSEQUENTLY FLUSHED WITHOUT OBSTRUCTION. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25375059 ↗MalfunctionMar 18, 2026Fluid/Blood Leak
IT WAS REPORTED ON (B)(6) 2026, A GYNECOLOGY PATIENT UNDERWENT PICC CATHETER PLACEMENT VIA THE LEFT BRACHIAL VEIN UNDER LOCAL ANESTHESIA. THE CATHETER WAS INSERTED TO A DEPTH OF 38 CM. 2.23 DURING INFUSION, CLEAR SEROUS FLUID WAS OBSERVED AT THE PUNCTURE SITE. APPROXIMATELY 5 ML WAS VISIBLE WITHIN THE TRANSPARENT DRESSING. THE DRESSING WAS REPLACED FOR MAINTENANCE. FOLLOWING A VENOUS SPECIALIST CONSULTATION THE NEXT DAY, A TEAR WAS IDENTIFIED 1.5 CM WITHIN THE CATHETER, PRESENTING AS A SMALL JET OF FLUID UNDER POSITIVE PRESSURE. THE CATHETER WAS TRIMMED AND THE CONNECTOR REPLACED. CURRENT CATHETER LENGTH WITHIN THE BODY IS 33 CM, WITH 3 CM EXPOSED EXTERNALLY. THE CATHETER TIP IS POSITIONED AT T5 AND SECURELY FIXED. IV INFUSION IS UNOBSTRUCTED. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24629967 ↗MalfunctionFeb 23, 2026Fluid/Blood Leak
IT WAS REPORTED THAT DURING INTRAVENOUS INFUSION FOLLOWING A PICC LINE INSERTION, LEAKAGE WAS DETECTED NEAR THE CONNECTOR. AFTER TRIMMING THE CATHETER AND REPLACING THE CONNECTOR, BLOOD RETURN WAS RESTORED, AND THE CATHETER REMAINED IN USE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24416046 ↗MalfunctionJan 20, 2026Fluid/Blood Leak
IT WAS REPORTED DURING THE IMPLANTATION OF A PERIPHERALLY INSERTED CENTRAL CATHETER (PICC) FOR THE PATIENT, IT WAS DISCOVERED THAT THE DISTAL END OF THE CATHETER HAD SPLIT AND WAS UNUSABLE. CONSEQUENTLY, THE CENTRAL VENOUS CATHETER WAS REPLACED AND REINSERTED FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24122992 ↗InjurySep 4, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, THE PATIENT UNDERWENT PICC PUNCTURE ON THE MORNING OF (B)(6), AND THE BLOOD RETURN WAS SMOOTH AND FIXED. AT ABOUT 20:00, THE NURSE FOUND THE INFUSION PUMP ALARM DURING THE INFUSION PROCESS, CHECKED THE PATIENT AND FOUND THAT THE CATHETER WAS RETURNING TO BLOOD, AND THE CATHETER NEAR THE LINKER WAS DAMAGED AND LEAKING, TRIMMED THE EXTRACORPOREAL CATHETER AND REPLACED THE NEW CATHETER CONNECTOR, AND THERE WAS NO BLOOD RETURN AFTER CONNECTION. AFTER THE LEAKAGE OCCURRED TWICE, THE PRUNED EXPOSED CATHETER WAS USED TO RECONNECT THE EXTENSION TUBE AND CONTINUE TO USE IT. THERE WERE TWO REPORTED LEAKS ON THE SAME DEVICE. THIS REPORT ADDRESSES BOTH INCIDENCES.
Patient problems: Insufficient Information
MAUDE report 22967565 ↗MalfunctionMay 15, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED FROM THE HOSPITAL, ON (B)(6) 2024, A CATHETER KINK OCCURRED IN THE PICC LINE OF A PATIENT IN THE REHABILITATION DEPARTMENT. AFTER SEVERAL ATTEMPTS, THE CATHETER COULD NOT BE DELIVERED TO THE SUPERIOR VENA CAVA. X-RAY EXAMINATION REVEALED VASCULAR MALFORMATION, WHICH PREVENTED THE CATHETER FROM REACHING THE SUPERIOR VENA CAVA. FINALLY, THE CATHETER WAS LEFT IN THE ARMPIT. A FEW DAYS AFTER CATHETERISATION, IT WAS FOUND THAT THERE WAS NO BLOOD RETURN WHEN THE CATHETER WAS WITHDRAWN. AS THE DRUG INFUSION WAS SMOOTH, THE CATHETER WAS LEFT IN PLACE AND THE PATIENT WAS KEPT UNDER CLOSE OBSERVATION. ON (B)(6) 2025, IT WAS FOUND THAT THE DRUG COULD NOT BE INFUSED, SO THE CATHETER WAS REMOVED. AFTER REMOVAL, A TEAR WAS FOUND AT THE TIP OF THE CATHETER. REPEATED ATTEMPTS TO INSERT THE CATHETER DURING THE CATHETERISATION PROCESS MAY HAVE DAMAGED THE BLOOD VESSELS, INCREASING THE LIKELIHOOD OF SUBSEQUENT PHLEBITIS IN THE PATIENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22031923 ↗MalfunctionApr 9, 2025Misassembled; Device Dislodged or Dislocated; Misassembled; Device Dislodged or Dislocated
IT WAS REPORTED BY THE CUSTOMER, ON (B)(6) 2025 THE PATIENT WITH PICC IN PLACE WAS ADMITTED TO THE HOSPITAL AND WAS FOUND TO BE OOZING MEDICATION DURING INFUSION. UPON INSPECTION BY THE CHARGE NURSE IT WAS FOUND THAT THERE WAS NO CATHETER PRESENT AT THE PUNCTURE POINT AND ONLY THE EXTERNAL EXTENSION TUBING AND CONNECTOR WAS FIXED OUTSIDE THE PATIENT'S BODY. THE HOSPITAL THEN INSPECTED THE EXTENSION TUBING DECOMPRESSION SLEEVE ON ITS OWN, WHICH MAY HAVE DAMAGED THE DECOMPRESSION SLEEVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21810059 ↗MalfunctionMar 28, 2025Break; Break
IT WAS REPORTED BY THE CUSTOMER THAT AFTER RADICAL MASTECTOMY FOR BREAST CANCER, THE NURSE MANAGER CARRIED OUT PICC INTUBATION OPERATION FOR THE PATIENT AT 10:00 ON MARCH 6, THE INTUBATION PROCESS WENT SMOOTHLY, AND THE GUIDEWIRE WAS WITHDRAWN TO INSTALL THE LOU GEHRIG'S CONNECTOR, AND IN THE PROCESS OF INSTALLING THE CONNECTOR, THE LOU GEHRIG'S CONNECTOR FLANK BROKE, AND THE PATIENT WAS IMMEDIATELY REPLACED WITH A SPARE LOU GEHRIG'S CONNECTOR, AND THE PATIENT'S PICC INTUBATION OPERATION WENT SMOOTHLY WITHOUT ANY ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION PROVIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21723951 ↗MalfunctionMar 28, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED ON (B)(6) 2025, A NURSE ON UNIT 1-5B LEFT A CENTRAL VENOUS CATHETER IN PLACE, PUMPED BACK INTO THE CATHETER TO OBSERVE NO BLOOD RETURN, AND UPON GIVING THE CATHETER FOR REMOVAL, FOUND THAT THE FRONT END OF THE CATHETER HAD RUPTURED. THE CATHETER WAS IMMEDIATELY REPLACED WITH A NEW ONE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21719485 ↗MalfunctionMar 12, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT THE PICC CATHETER WAS BROKEN AND LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21585860 ↗MalfunctionFeb 11, 2025Suction Problem; Failure to Infuse; Suction Problem; Failure to Infuse
IT WAS REPORTED, AFTER PLACING OUR GROSHONG PICC (MODEL 7717405) CATHETER, THE NEXT DAY, THE LIQUID DID NOT DRIP, THROUGH A VARIETY OF TROUBLESHOOTING, THIS SITUATION STILL OCCURS, AND THEN DECIDED TO PULL OUT THE CATHETER IN VITRO INSPECTION, DID NOT FIND ANY RESIDUES OF DRUGS IN THE CATHETER AND BLOOD STAGNATION AND OTHER FOREIGN OBJECTS, CONNECTED TO THE SYRINGE INJECTION SMOOTH, BUT CONNECTED TO THE INFUSER, THE INFUSER BOTTLE ELEVATED AND THEN DID NOT DRIP, SALES WAS AWARE OF THIS AND WENT TO CHECK IT BY BRIEFLY FOLDING BACK TO LOOSEN THE INFUSION TUBE, THE LIQUID DOES NOT DRIP, AND THEN FORCE TO THE INFUSION BOTTLE AFTER THE PRESSURE OF THE LIQUID CAN DRIP, AND THEN TURN OFF THE INFUSION DEVICE AND THEN OPEN THE SITUATION DOES NOT DRIP, SUSPECTED THAT THE PRESSURE AT THE TIP OF THE CATHETER THREE-WAY VALVE THAN THE PRESSURE OF OTHER NORMAL USE OF THE CATHETER TO BE GREATER, RESULTING IN THE LIQUID CAN NOT BE DRIPPED WITHIN THE NORMAL RANGE OF PRESSURE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21360732 ↗MalfunctionFeb 11, 2025Component Incompatible; Detachment of Device or Device Component; Component Incompatible; Detachment of Device or Device Component
IT WAS REPORTED THAT THREE DAYS AFTER THE CATHETER WAS PLACED, THE CATHETER WAS EXPOSED AND THE CONNECTOR DISCONNECTED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21358767 ↗MalfunctionSep 26, 2024Break; Fluid/Blood Leak; Break; Fluid/Blood Leak
IT WAS REPORTED, THE PATIENT WAS PLACED IN A PICC CATHETER FOR ELECTIVE CHEMOTHERAPY ON MAY 22, THE CATHETER BROKE OCCURRED OUTSIDE THE BODY ON MAY 23. THE CATHETER WAS NOT REMOVED FROM THE PATIENT, BUT THE EXPOSED PART WAS TRIMMED DIRECTLY. REPLACE WITH A NEW COMPRESSION SLEEVE. NO HARM TO PATIENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20312177 ↗MalfunctionAug 28, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING THE VISIT TO REPORTING FACILITY THAT ON (B)(6) 2024, MORE THAN THREE WEEKS AFTER THE SUDDEN SWELLING OF THE ARM, SWELLING AREA OF 10 X 10CM, IN ACCORDANCE WITH THE MEDICAL ADVICE OF PICC EXTRACTION, FOUND THAT THE PICC CATHETER INTO THE BODY OF THE PLACE OF 1 ~ 2 CENTIMETERS OF THE PLACE OF THE BREAKAGE, AND FOAM DRESSING WAS IMMEDIATELY GIVEN TO REDUCE SWELLING EXTERNALLY, AND THE SWELLING SUBSIDED WITHOUT OTHER ADVERSE REACTIONS AFTER 3 DAYS. AS THERE WAS BLOOD RESIDUE ON THE CATHETER, IT WAS DISCARDED AS MEDICAL WASTE. NO OTHER INFORMATION PROVIDED, AT THIS TIME.
Patient problems: Swelling/ Edema
MAUDE report 20085896 ↗MalfunctionAug 26, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, A PICC TUBE WAS INSERTED IN FACILITY ICU FOR MORE THAN THREE WEEKS. ON THE AFTERNOON OF THE 24TH, INFUSION TREATMENT SUDDENLY CAUSED A 10*10CM SWELLING AT THE INSERTION POINT. AFTER EXTUBATION ACCORDING TO THE DOCTOR'S INSTRUCTIONS, IT WAS FOUND THAT THE PICC TUBE WAS DAMAGED AT THE POINT WHERE IT WAS EMBEDDED IN THE BODY. IT'S AN EXCEPTION. REPLACE THE INTRAVENOUS CATHETER WITH A NEW ONE, NOTIFY THE PATIENT, OBTAIN UNDERSTANDING, AND NOTIFY RELEVANT HOSPITAL PERSONNEL.
Patient problems: Swelling/ Edema
MAUDE report 20065091 ↗MalfunctionAug 23, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED ON (B)(6) 2024, THE ICU DEPARTMENT PLACED A GROSHONGPICC CATHETER IN THE PATIENT AND WAS TREATED IN THE ICU. ON (B)(6) 2024, THE PATIENT'S FAMILY FOUND SWELLING IN THE PATIENT'S ARM ON THE SIDE WHERE THE CATHETER WAS PLACED, AND REPORTED IT TO THE HEAD NURSE. AFTERWARDS, THEY CHECKED THE PICC CATHETER AND FOUND THAT IT WAS IN THE ICU. LEAKAGE WAS FOUND ABOUT 2CM FROM THE PUNCTURE POINT TOWARD THE PROXIMAL END, AND THE CATHETER WAS RUPTURED. PATIENT WAS EXTUBATED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Swelling/ Edema
MAUDE report 20048473 ↗MalfunctionApr 25, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, "ON (B)(6) 2024, THE PATIENT HAD SWELLING AT THE PICC PUNCTURE SITE DURING TARGETED THERAPY, AND IMMEDIATELY SUSPENDED THE INFUSION AND ASKED THE PICC OUTPATIENT CLINIC TO ASSIST IN THE TREATMENT, AND FOUND THAT THE PICC CATHETER WAS DAMAGED, AND THE INTERCEPTION CATHETER WAS GIVEN 5 CM, AND THE CHEST X-RAY SHOWED THAT THE END OF THE CATHETER WAS LOCATED AT THE LEVEL OF THE FIFTH POSTERIOR RIB ON THE RIGHT SIDE, AND INTRAVENOUS INFUSION COULD BE CONTINUED, BUT THE POSSIBILITY OF VENOUS THROMBOSIS AND THE RISK OF CATHETER-RELATED VENOUS BLOODSTREAM INFECTION WOULD BE INCREASED DURING USE, AND THE PATIENT WAS GIVEN REASSURANCE, AND THE PATIENT HAD NO OBJECTION." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Swelling/ Edema
MAUDE report 19183012 ↗MalfunctionApr 12, 2024Break; Break
IT WAS REPORTED THAT WHEN NURSE CHANGES DRESSING THE TUBE WAS BROKEN. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19096354 ↗MalfunctionDec 14, 2023Defective Component; Defective Component
IT WAS REPORTED ON (B)(6) 2023, THE PATIENT WAS ADMITTED TO OUR DEPARTMENT WITH CEREBRAL ARTERY OCCLUSION AND CEREBRAL INFARCTION FOR 4.5 HOURS DUE TO LEFT LIMB WEAKNESS AND SPEECH IMPAIRMENT, AND ON NOVEMBER 17, WHEN THE PICC CATHETER WAS INDWELLED, IT WAS FOUND THAT THE CENTRAL VENOUS CATHETER THROUGH THE PERIPHERAL CANNULA WAS NOT SMOOTH AND COULD NOT BE USED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18332229 ↗MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WHEN THE CATHETER WAS PLACED BY THE TEACHER, IT WAS FOUND THAT THE CATHETER WAS LEAKING, WHICH AFFECTED THE NORMAL USE. REOPEN THE NEW CATHETER FOR PATIENT PLACEMENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18238288 ↗MalfunctionNov 17, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THE CATHETERIZATION TEACHER FOUND THAT THERE WAS A LEAKAGE DURING THE OPERATION, WHICH AFFECTED THE NORMAL USE. REOPEN THE NEW CATHETER FOR PATIENT PLACEMENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18161333 ↗InjuryNov 8, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THE PATIENT REQUESTED CHEMOTHERAPY MAINLY DUE TO BREAST CANCER SURGERY FOR MORE THAN 3 MONTHS, CAME TO OUR DEPARTMENT ON (B)(6) 2023, AND WAS ADMITTED TO THE HOSPITAL WITH "MILK ROCK" IN THE OUTPATIENT CLINIC, AND UNDERWENT THE FIRST INTRAVENOUS CHEMOTHERAPY ON (B)(6) 2023, AND UNDERWENT PERCUTANEOUS TRANSPERIPHERAL CATHETERIZATION IN OUR DEPARTMENT. AT THE TIME OF ADMISSION, THE PATIENT COMPLAINED OF EARTHWORM-LIKE VARICOSE VEINS IN THE SUPERFICIAL BLOOD VESSELS AROUND THE CATHETER, WITH SUPERFICIAL VEIN BULGING AND DILATION, AND PERIPHERAL INDURATION CHEST X-RAY RE-EXAMINATION ON (B)(6) 2023: ECTOPIC CATHETER. PHLEBITIS IS CONSIDERED TO BE CAUSED BY ECTOPIC CENTRAL VENOUS CATHETERS THROUGH PERIPHERAL TEAHOUSES.
Patient problems: Aneurysm; Vasculitis
MAUDE report 18097075 ↗MalfunctionNov 1, 2023Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED BY THE CUSTOMER "WHEN THE CATHETER WAS OPENED AND THE CATHETER WAS PRE-FLUSHED, A WHITE FOREIGN BODY WAS FOUND FLUSHED OUT OF THE CATHETER VALVE, AND THE CATHETER WAS AFRAID TO USE THE CATHETER IN THE HUMAN BODY, AND RE-OPENED THE NEW CATHETER FOR INSERTION." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18049721 ↗MalfunctionOct 26, 2023Material Frayed; Material Frayed
IT WAS REPORTED THE PATIENT WAS ADMITTED TO THE ONCOLOGY DEPARTMENT OF OUR HOSPITAL AS "COUGHING UP SPUTUM AND SHORTNESS OF BREATH FOR MORE THAN 5 MONTHS, AND 20 DAYS AFTER THE FIRST CYCLE OF CHEMOTHERAPY AFTER RADIOTHERAPY FOR LUNG CANCER", AND NOW IN ORDER TO SEEK FURTHER ANTI-TUMOR TREATMENT, THE OUTPATIENT CLINIC IS ADMITTED TO THE ONCOLOGY DEPARTMENT OF OUR HOSPITAL WITH "LUNG CANCER". PRELIMINARY DIAGNOSIS: 1. UPPER LOBE SQUAMOUS CELL CARCINOMA OF THE LEFT LUNG WITH MEDIASTINAL AND BILATERAL HILAR LYMPH NODE METASTASIS (CT4N3M0, STAGE III.B) AFTER RADIOTHERAPY; 2. MAINTENANCE CHEMOTHERAPY FOR MALIGNANT TUMORS; 3. REFLUX ESOPHAGITIS; 4. FUNCTIONAL DYSPHAGIA; 5. LUNG ABSCESS; 6. OBSTRUCTIVE PNEUMONIA; 7. HYPERLIPIDEMIA; 8. CHRONIC GASTRITIS; 9. PULMONARY EMBOLISM?; 10. KIDNEY STONES; 11. KIDNEY CYSTS. TC CHEMOTHERAPY WAS GIVEN, PACLITAXEL (ALBUMIN-BOUND) 350MG IVGTT D1 + CARBOPLATIN 330MG IVGTT, Q3W. HE WAS GIVEN ACID SUPPRESSION, ANTIEMETIC AND HEPATOPROTECTIVE THERAPY DURING CHEMOTHERAPY, AND BEGAN TO USE PICC CATHETER INFUSION CHEMOTHERAPY DRUGS THROUGH PERIPHERAL INTUBATION ON (B)(6) 2023. WHEN THE PATIENT IS PUNCTURED, THE GUIDEWIRE OF THE CENTRAL VENOUS CATHETER PICC THRO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18012798 ↗MalfunctionOct 10, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WHEN PRE-FLUSHING THE CATHETER, THE OPERATOR FOUND THAT LIQUID LEAKS OCCURRED AT THE SITE 2 CM FROM THE VALVE AT THE PROXIMAL END OF THE CATHETER. NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION: IT WAS REPORTED THE CATHETER WAS NOT USED ON THE PATIENT AND THERE WAS NO HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17908769 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page