Navarre NNU8LPTRecalls & reported problems
Bard Access Systems, Inc. · Navarre Universal Drainage Catheter with Nitinol, Locking Pigtail, 8 Fr x 30 cm · Class 1
All models of Navarre (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Bard Navarre Universal drainage catheter has 754 reports, with no deaths and 18 injuries. Most reports describe the catheter fracturing or breaking, often at its connector, and some describe material separating or fluid leaking out. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 738
- Deaths reported
- 0
- Injuries reported
- 18
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 738 adverse event reports that list this exact device (DI 00801741037009), received between Mar 22, 2017 and Sep 30, 2026 — classified as 18 injuries, 720 malfunctions.
The most frequently reported specific device problem is “Fracture” (389 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture389
- Material Separation169
- Fluid/Blood Leak130
- Break101
- Defective Component99
- Crack37
- Material Integrity Problem28
- Nonstandard Device25
Patient problems
- No Clinical Signs, Symptoms or Conditions673
- No Consequences Or Impact To Patient25
- No Known Impact Or Consequence To Patient24
- Extravasation11
- No Patient Involvement4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Unspecified Infection2
- Abdominal Pain1
Recent adverse event reports
The 25 most recent of 738 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Material Integrity Problem
It was reported that on (b)(6) 2026, during a gallbladder surgery procedure, a single use drainage catheter was selected for use. Upon opening the sterile package, the healthcare professional observed that the suture attached to the catheter was damaged. As a result, the device did not meet the requirements for intraoperative use and could not be used as intended. The affected device was immediately removed from use, and a replacement drainage catheter was obtained to continue the procedure. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944182 ↗MalfunctionSep 28, 2026Fracture; Material Separation
A patient underwent an ultrasound guided percutaneous transhepatic cholangiography drainage (PTCD) catheter placement procedure using the Navarre Universal Drainage Catheter. Following the procedure, on (b)(6) 2026, it was reported that the drainage catheter connector had fractured. The catheter was subsequently replaced, and the procedure was completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910914 ↗MalfunctionSep 24, 2026Break; Material Separation
On (b)(4) 2026, a patient underwent an ultrasound guided percutaneous drainage catheter placement procedure using Navarre Universal Drainage Catheter. Prior to the procedure, the wire is disconnected at the from the pigtail. The procedure was completed using another device. There was no patient contact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50881315 ↗MalfunctionSep 24, 2026Nonstandard Device
On (b)(6) 2026, a patient underwent an ultrasound guided cyst drainage catheter placement procedure using Navarre Universal Drainage Catheter. Prior to the procedure, it was noted that the trocar needle was shorter in length than the catheter. The procedure was completed using another device. There was no patient contact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50877870 ↗MalfunctionSep 21, 2026Break; Material Separation
On (b)(6) 2026, a patient underwent a computed tomography guided percutaneous transhepatic cholangiography drainage (PTCD) catheter placement procedure using a Navarre Universal Drainage Catheter. During the procedure, it was reported that the Sure-Lok thread completely disengaged from the catheter. The procedure was completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50837140 ↗MalfunctionSep 10, 2026Fluid/Blood Leak
On (b)(6) 2026, a patient underwent an ultrasound guided ascites percutaneous drainage catheter placement procedure using a Navarre Universal Drainage Catheter. During the procedure, it was reported that leakage was observed from the drainage catheter. The procedure was subsequently completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713087 ↗MalfunctionSep 10, 2026Fluid/Blood Leak
On (b)(6), 2026, a patient underwent an ultrasound guided ascites percutaneous drainage catheter placement procedure using a Navarre Universal Drainage Catheter. During the procedure, it was reported that leakage was observed from the drainage catheter. The procedure was subsequently completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713080 ↗MalfunctionSep 10, 2026Fluid/Blood Leak
On (b)(6) 2026, a patient underwent an ultrasound guided ascites percutaneous drainage catheter placement procedure using a Navarre Universal Drainage Catheter. During the procedure, it was reported that leakage was observed from the drainage catheter. The procedure was subsequently completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713079 ↗MalfunctionSep 9, 2026Fluid/Blood Leak
On (b)(6) 2026, a patient underwent an ultrasound guided ascites percutaneous drainage procedure using a Navarre Universal Drainage Catheter. It was reported that during the procedure, leakage was noted. The procedure was completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50689902 ↗MalfunctionSep 9, 2026Fluid/Blood Leak
On (b)(6) 2026, a patient underwent an ultrasound guided ascites percutaneous drainage procedure using a Navarre Universal Drainage Catheter. It was reported that during the procedure, leakage was noted. The procedure was completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50689900 ↗MalfunctionSep 9, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50688754 ↗MalfunctionSep 7, 2026Fracture
A patient underwent an ultrasound guided ascites percutaneous drainage procedure using a Navarre Universal Drainage Catheter. On (b)(6) 2026, the drainage catheter connector was found to be fractured. The procedure was subsequently completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50650591 ↗MalfunctionSep 7, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50650588 ↗MalfunctionSep 7, 2026Fracture; Material Separation
A patient underwent an ultrasound guided ascites percutaneous drainage procedure using a Navarre Universal Drainage Catheter. On (b)(6) 2026, the drainage catheter connector was found to be fractured. The procedure was subsequently completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50649815 ↗MalfunctionSep 7, 2026Fluid/Blood Leak; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50649814 ↗MalfunctionSep 7, 2026Break; Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50649100 ↗MalfunctionSep 7, 2026Device Contamination with Chemical or Other Material
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50642634 ↗MalfunctionSep 7, 2026Device Contamination with Chemical or Other Material
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50642631 ↗MalfunctionJun 26, 2026Fluid/Blood Leak; Fracture; Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING A NAVARRE UNIVERSAL DRAINAGE CATHETER. POST PROCEDURE, THE DRAINAGE CATHETER CONNECTOR WAS REPORTEDLY FRACTURED AND LEAKING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25633692 ↗MalfunctionJun 23, 2026Fracture; Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE DRAINAGE CATHETER. POST THE PROCEDURE, THE DRAINAGE CATHETER CONNECTOR WAS ALLEGEDLY FRACTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25590266 ↗MalfunctionJun 23, 2026Fracture
ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE DRAINAGE CATHETER. POST THE PROCEDURE, THE DRAINAGE CATHETER CONNECTOR WAS ALLEGEDLY FRACTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25590128 ↗MalfunctionJun 9, 2026Break; Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE UNIVERSAL DRAINAGE CATHETER. PRIOR TO THE PROCEDURE, IT WAS NOTED THAT SURE LOK THREAD WAS ALLEGEDLY DIGRESSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25440052 ↗MalfunctionJun 3, 2026Break; Material Separation
ON (B)(6) 2026, THE PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE BY USING NAVARRE UNIVERSAL DRAINAGE CATHETER. IT WAS REPORTED THAT DURING THE PROCEDURE, IT WAS NOTED THAT THE SURE LOK THREAD WAS ALLEGEDLY BROKEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25372192 ↗MalfunctionJun 3, 2026Break; Material Separation
ON (B)(6) 2026, THE PATIENT UNDERWENT AN ULTRASOUND GUIDED PERCUTANEOUS DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE UNIVERSAL DRAINAGE CATHETER. IT WAS REPORTED THAT DURING THE PROCEDURE, IT WAS NOTED THAT THE SURE LOK THREAD WAS ALLEGEDLY BROKEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25372189 ↗MalfunctionMay 25, 2026Fracture; Material Separation
ON (B)(6) 2026, THE PATIENT UNDERWENT AN ULTRASOUND GUIDED CYST DRAINAGE CATHETER PLACEMENT PROCEDURE USING NAVARRE UNIVERSAL DRAINAGE CATHETER. POST PROCEDURE, IT WAS NOTED THAT THE DRAINAGE CATHETERS CONNECTOR WAS FRACTURED AND SHATTERED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267807 ↗
FDA clearance and classification
The FDA regulates product code GBX (catheter, irrigation) under 21 CFR 878.4200 as a Class I device. 21 CFR 878.4200 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Navarre NNU8LPT
Has the Navarre NNU8LPT been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741037009), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Navarre NNU8LPT?
The FDA's MAUDE database lists 738 adverse event reports for this device, received between Mar 22, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Fracture” (389), “Material Separation” (169) and “Fluid/Blood Leak” (130); the most common patient problems are “Extravasation” (11), “Unspecified Infection” (2) and “Abdominal Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Navarre NNU8LPT?
None of the reports are classified as deaths, 18 as injuries and 720 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Navarre NNU8LPT FDA approved?
Devices of this type (product code GBX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Bard Access Systems, Inc.
Who makes the Navarre NNU8LPT, and is it still sold?
The FDA's device database lists Bard Access Systems, Inc. as the company responsible for it (brand Navarre, model NNU8LPT). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, IRRIGATION devices
- Navarre Opti-Drain NOD8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION900 reports
- BLAKE 2227Ethicon Inc. · CATHETER, IRRIGATION331 reports
- Navarre NNU6LPTBard Access Systems, Inc. · CATHETER, IRRIGATION213 reports
- Navarre Opti-Drain NOD10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION180 reports
- Navarre NNU10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION127 reports
- BLAKE 2233Ethicon Inc. · CATHETER, IRRIGATION114 reports
- Vasoshield VH-5001Maquet Cardiovascular LLC · Catheter, Irrigation1 recall27 reports
- BLAKE 2229Ethicon Inc. · CATHETER, IRRIGATION76 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page