Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741038136), received between Dec 6, 2017 and Dec 6, 2017 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 6, 2017Break
IT WAS REPORTED BY THE FACILITY THAT THE CLAMP BROKE ON THE END OF THE SLIDE. IT WAS STATED THE FACILITY USED CHG TO CLEAN THE CATHETER BUT NOT THE CLAMP. FACILITY CONTACT REPORTED THAT THE DRESSING IS CHANGED EVERY SIX DAYS WITH USE OF CHG AND THE CATHETER IS ACCESSED FOR INFUSION AND ASPIRATION EVERY EIGHT HOURS OR WITH INTERMITTENT USE. THE CLAMP BROKE AFTER A FEW DAYS OF USE. THIS REPORT ADDRESS THE SECOND CLAMP.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7089378 ↗MalfunctionDec 6, 2017Break
IT WAS REPORTED BY THE FACILITY THAT THE CLAMP BROKE ON THE END OF THE SLIDE. IT WAS STATED THE FACILITY USED CHG TO CLEAN THE CATHETER BUT NOT THE CLAMP. FACILITY CONTACT REPORTED THAT THE DRESSING IS CHANGED EVERY SIX DAYS WITH USE OF CHG AND THE CATHETER IS ACCESSED FOR INFUSION AND ASPIRATION EVERY EIGHT HOURS OR WITH INTERMITTENT USE. THE CLAMP BROKE AFTER A FEW DAYS OF USE. THIS REPORT ADDRESS THE THIRD CLAMP.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7089333 ↗MalfunctionDec 6, 2017Break
IT WAS REPORTED BY THE FACILITY THAT THE CLAMP BROKE ON THE END OF THE SLIDE. IT WAS STATED THE FACILITY USED CHG TO CLEAN THE CATHETER BUT NOT THE CLAMP. FACILITY CONTACT REPORTED THAT THE DRESSING IS CHANGED EVERY SIX DAYS WITH USE OF CHG AND THE CATHETER IS ACCESSED FOR INFUSION AND ASPIRATION EVERY EIGHT HOURS OR WITH INTERMITTENT USE. THE CLAMP BROKE AFTER A FEW DAYS OF USE. THIS REPORT ADDRESS THE FIRST CLAMP.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7089284 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page