Bard® Clean-Cath® Intermittent Catheter 420712Recalls & reported problems
C. R. Bard, Inc. · Bard® Clean-Cath® Intermittent Catheter · Class 2
All models of Bard Clean-Cath Intermittent Catheter (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741038730), received between Aug 17, 2020 and Sep 21, 2021 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Biocompatibility” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jan 11, 2021Class IITerminated
The device catheter tip was cut off (no tip) and the product packaging was not fully sealed.
Z-1431-2021 · Root cause: Packaging process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Biocompatibility1
- Break1
- Component Missing1
- Detachment of Device or Device Component1
- Material Protrusion/Extrusion1
- Material Split, Cut or Torn1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- Pain2
- Patient Problem/Medical Problem2
- Blood Loss1
- Hemorrhage/Blood Loss/Bleeding1
- Laceration(s)1
- No Clinical Signs, Symptoms or Conditions1
- Urinary Tract Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 21, 2021Biocompatibility; Biocompatibility
IT WAS REPORTED THAT THE INTERMITTENT CATHETER CAUSED A URINARY TRACT INFECTION TO THE PATIENT. IT WAS UNKNOWN WHAT MEDICAL INTERVENTION WAS PROVIDED FOR URINARY TRACT INFECTION.
Patient problems: Urinary Tract Infection
MAUDE report 12505916 ↗MalfunctionMar 9, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WERE CONFIRMED REPORTS OF THE CATHETER TIP COULD POTENTIALLY BE CUT OFF RESULTING IN THE CATHETER POUCH SEAL HAVING AN OPENING NEAR THE TIP OF THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11446844 ↗MalfunctionAug 21, 2020Component Missing; Detachment of Device or Device Component; Material Protrusion/Extrusion; Material Protrusion/Extrusion
IT WAS REPORTED THAT THE TIP OF A CATHETERS WAS CLEANLY CUT OFF AND TIP WAS MISSING. ALSO STATED PATIENT EXPERIENCED BLEEDING AND PAIN. NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain; Patient Problem/Medical Problem
MAUDE report 10439606 ↗MalfunctionAug 17, 2020Break; Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE TIP OF A CATHETERS WAS PERFECTLY CUT OFF. ALSO STATED PATIENT EXPERIENCED BLEEDING AND PAIN, FELT A SCRAPING AGAINST THE PATIENT'S URETHRA. NO MEDICAL INTERVENTION WAS REPORTED. PER SAMPLE EVALUATION DETERMINED THROUGH EVALUATION THAT THE TIP OF THE CATHETER WAS NOT FORMED PROPERLY. DURING SAMPLE EVALUATION, IT WAS DETERMINED THAT THE CUT CATHETER WAS CAUSED BECAUSE THE CATHETER WAS STICKING OUT OF THE PACKAGING DURING SEALING CAUSING A STERILE BARRIER BREACH.
Patient problems: Blood Loss; Laceration(s); Pain; Patient Problem/Medical Problem
MAUDE report 10414481 ↗
Other CATHETER, UROLOGICAL devices
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50614C. R. Bard, Inc. · CATHETER, UROLOGICAL883 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53614GC. R. Bard, Inc. · CATHETER, UROLOGICAL692 reports
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50616C. R. Bard, Inc. · CATHETER, UROLOGICAL2 deaths443 reports
- GENTLECATH GLIDE 421567Convatec, Purchasing Department · Catheter, Urological342 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53616GC. R. Bard, Inc. · CATHETER, UROLOGICAL301 reports
- Bardex® Foley Catheter Silicone Coated 0167V22SC. R. Bard, Inc. · CATHETER, UROLOGICAL187 reports
- Bardex® Foley Catheter Silicone Coated 0167V20SC. R. Bard, Inc. · CATHETER, UROLOGICAL175 reports
- Bardia® Coude Olive Tip Urethral Foley Catheter 802516C. R. Bard, Inc. · CATHETER, UROLOGICAL170 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page