Denali® Vena Cava Filter DL900FRecalls & reported problems
Bard Peripheral Vascular, Inc. · DENALI® Vena Cava Filter Femoral · Class 2 · Implant
All models of Denali Vena Cava Filter (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Bard Denali vena cava filter has 529 adverse event reports, with 22 deaths and 238 injuries. The reports often describe the filter sitting in the wrong position, parts such as the flush port detaching during placement, and difficult removal, and some mention pulmonary embolism or blood clots. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 530
- Deaths reported
- 22
- Injuries reported
- 239
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 530 adverse event reports that list this exact device (DI 00801741040801), received between Oct 21, 2015 and Sep 28, 2026 — classified as 22 deaths, 239 injuries, 269 malfunctions.
The most frequently reported specific device problem is “Malposition of Device” (171 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Malposition of Device171
- Patient Device Interaction Problem142
- Difficult to Remove123
- Patient-Device Incompatibility96
- Detachment of Device or Device Component81
- Adverse Event Without Identified Device or Use Problem72
- Migration or Expulsion of Device65
- Positioning Problem34
Patient problems
- No Consequences Or Impact To Patient149
- No Clinical Signs, Symptoms or Conditions144
- Pulmonary Embolism93
- Abdominal Pain70
- Thrombosis/Thrombus28
- Thrombus23
- Death21
- Pain19
Recent adverse event reports
The 25 most recent of 530 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Migration
A patient underwent a percutaneous inferior vena cava (IVC) filter implantation procedure using a Denali Filter via femoral vein access. During the procedure, it was reported that the filter had allegedly migrated from its intended position. It was further reported that two attempts were made to retrieve the Denali Filter; however, the retrieval attempts were unsuccessful. Subsequently, another physician ordered an abdominal MRI for further evaluation. The patient's current status is unknown.
Patient problems: Foreign Body In Patient
MAUDE report 50910808 ↗MalfunctionAug 10, 2026Detachment of Device or Device Component
On (b)(6) 2026, a patient was prepped for a percutaneous inferior vena cava filter implantation via the femoral vein using a Denali Filter. Prior to use, the user pulled the Denali Y adapter from the protective tubing and the Denali broke apart from the Y adapter and the pusher. The procedure was completed using another device. There was no patient contact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50191944 ↗MalfunctionAug 10, 2026Detachment of Device or Device Component
On (b)(6) 2026, a patient underwent a percutaneous inferior vena cava filter implantation via the femoral vein using a Denali filter. During the procedure, the flushing port on the delivery sheath was allegedly detached. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50191009 ↗MalfunctionJun 4, 2026Detachment of Device or Device Component
ON (B)(6) 2026, A PATIENT WAS PREPPED FOR A PERCUTANEOUS INFERIOR VENA CAVA FILTER IMPLANTATION USING A DENALI FILTER. PRIOR TO THE PROCEDURE, IT WAS NOTICED THAT THE SIDE ARM FLUSH FELL OFF. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25389765 ↗MalfunctionMar 25, 2026Failure to Advance; Detachment of Device or Device Component
ON (B)(6) 2025, PRIOR TO A PERCUTANEOUS INFERIOR VENA CAVA FILTER IMPLANTATION USING A DENALI FILTER. PRIOR TO THE PROCEDURE, IT WAS REPORTED THAT THE FILTER ALLEGEDLY UNABLE TO ADVANCE THROUGH THE DEPLOYMENT SYSTEM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24686795 ↗MalfunctionMar 18, 2026Improper or Incorrect Procedure or Method; Detachment of Device or Device Component
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS INFERIOR VENA CAVA FILTER IMPLANTATION IN FEMORAL VEIN USING A DENALI FILTER. DURING THE PROCEDURE, IT WAS FOUND THAT THE SHEATH WAS ALLEGEDLY DETACHED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24631360 ↗MalfunctionFeb 3, 2026Malposition of Device
ON (B)(6) 2026, A PATIENT UNDERWENT A PERCUTANEOUS INFERIOR VENA CAVA FILTER IMPLANTATION USING A DENALI FILTER. DURING THE PROCEDURE, THE SHEATH GRABBED THE FILTER AND ALLEGEDLY TURNS THE FILTER INTO SIDEWAY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24244424 ↗MalfunctionJan 29, 2026Detachment of Device or Device Component
ON (B)(6) 2026, A PATIENT UNDERWENT A PERCUTANEOUS INFERIOR VENA CAVA FILTER IMPLANTATION BY USING A DENALI FILTER. IT WAS REPORTED THAT DURING THE PROCEDURE, THE FILTER TRANSDUCER PORT WAS ALLEGEDLY DETACHED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24207151 ↗InjuryDec 1, 2025Difficult to Remove; Difficult to Remove
ON (B)(6) 2024, A PATIENT UNDERWENT A PERCUTANEOUS INFERIOR VENA CAVA FILTER IMPLANTATION USING A DENALI FILTER. POST THE PROCEDURE ON (B)(6) 2025, DURING REMOVAL, THE FILTER PUNCTURED PATIENT'S HEART WHICH REQUIRED EMERGENCY CARDIO SURGERY/REPAIR AND TRANSPORT TO ANOTHER FACILITY. FURTHERMORE, THE HCP ATTEMPTED TO REMOVE THE DENALI IVC FILTER USING A BENSON WIRE. WHILE INTRODUCING AND ADVANCING THE SHEATH THROUGH THE IVC FILTER, THE SHEATH BUCKLED. THE HCP SUBSEQUENTLY REALIGNED THE SHEATH AND WAS ABLE TO SUCCESSFULLY RETRIEVE THE IVC FILTER. THE REPORTER RECANTED THE ALLEGATION OF DEVICE FAILURE, CONFIRMING THAT THE REPORTED ISSUE WAS ASSOCIATED WITH A NON-BD DEVICE.
Patient problems: Perforation of Vessels
MAUDE report 23678446 ↗MalfunctionNov 20, 2025Positioning Failure; Deformation Due to Compressive Stress; Detachment of Device or Device Component; Positioning Failure; Deformation Due to Compressive Stress; Detachment of Device or Device Component
ON (B)(6) 2025, A 87-YEAR-OLD MALE PATIENT UNDERWENT AN INFERIOR VENA CAVA FILTER PLACEMENT PROCEDURE IN RIGHT FEMORAL GROIN VEIN FOR DEEP VEIN THROMBOSIS. DURING THE PROCEDURE THE FILTER ALLEGEDLY DID NOT DEPLOY 2X SAME PT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23602912 ↗MalfunctionAug 4, 2025Premature Activation; Detachment of Device or Device Component; Premature Activation; Detachment of Device or Device Component
ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS INFERIOR VENA CAVA FILTER IMPLANTATION USING A DENALI FILTER. DURING THE PROCEDURE, IT WAS REPORTED THAT THE Y-CONNECTION CAME APART AND WHEN DOING SO THE FILTER DEPLOYED PRE-MATURELY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22697533 ↗MalfunctionApr 28, 2025Activation Failure; Activation Failure
ON (B)(6) 2025, A PATIENT UNDERWENT AN INFERIOR VENA CAVA FILTER IMPLANTATION PROCEDURE USING DENALI FILTER. IT WAS REPORTED THAT THE DENALI FILTER ALLEGEDLY DID NOT OPEN WHEN DEPLOYED, FILTER MANUALLY OPEN WITH A BALLOON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21916860 ↗InjuryJan 9, 2025Improper or Incorrect Procedure or Method; Migration; Improper or Incorrect Procedure or Method; Migration
IT WAS REPORTED THAT DURING AN INFERIOR VENA CAVA FILTER PLACEMENT PROCEDURE, THE FILTER HAD ALLEGEDLY MIGRATED THROUGH THE HEART TO THE PULMONARY ARTERY. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 21119614 ↗InjuryNov 1, 2024Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT A VENA CAVA FILTER WAS PLACED IN A PATIENT WITH LOWER EXTREMITY DEEP VEIN THROMBOSIS AND PULMONARY EMBOLUS. IT WAS REPORTED THAT THE TWO YEARS, ONE MONTH AND TEN DAYS POST A FILTER DEPLOYMENT, IT WAS ALLEGED THAT FILTER STRUT DETACHED AND ALLEGEDLY DIFFICULT TO REMOVE THE FILTER. IT WAS FURTHER REPORTED THAT THE FILTER REMOVED AND NEW FILTER HAS BEEN PLACED. THE FRACTURED STRUT COMPLETELY BURIED WITHIN THE BRANCH OF THE HEPATIC VEIN. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Foreign Body In Patient
MAUDE report 20586331 ↗MalfunctionJul 10, 2024Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A VENA CAVA FILTER PLACEMENT PROCEDURE VIA THE FEMORAL ARTERY, THE FILTER ALLEGEDLY WOULD NOT BE OPENED UP. IT WAS FURTHER REPORTED THE FILTER WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19708252 ↗InjuryMar 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT THREE YEARS, SEVEN MONTHS, AND TWENTY-ONE DAYS POST A VENA CAVA FILTER PLACEMENT VIA THE RIGHT COMMON FEMORAL VEIN, IT WAS ALLEGED THAT THE FILTER STRUT HAD DETACHED AND MIGRATED INTO THE PATIENT'S RIGHT VENTRICLE. IT WAS FURTHER REPORTED THAT THE PATIENT ALLEGEDLY EXPERIENCED CHEST PAIN AND LATER DIED DUE TO COMPLICATIONS OF METASTATIC LUNG CANCER. HOWEVER, THE RELATIONSHIP OF THE DEATH WAS NOT RELATED TO THE DEVICE. REPORTEDLY, THE FILTER WAS REMOVED DURING AUTOPSY.
Patient problems: Chest Pain; Insufficient Information
MAUDE report 18957253 ↗MalfunctionFeb 9, 2024Difficult to Remove; Positioning Problem; Activation Failure; Difficult to Remove; Positioning Problem; Activation Failure
IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT A VENA CAVA FILTER WAS DEPLOYED SUCCESSFULLY IN A PATIENT WITH DEEP VEIN THROMBOSIS. DURING DEPLOYMENT, IT WAS FOUND THAT THE FILTER WAS ALLEGEDLY DEPLOYED IN AN UNSATISFACTORY POSITION AND THE FILTER DID NOT OPEN WELL. IT WAS FURTHER REPORTED THAT THE FILTER WAS ALLEGEDLY NOT ABLE TO REMOVE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18675098 ↗InjuryFeb 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT A VENA CAVA FILTER WAS DEPLOYED SUCCESSFULLY, IN A PATIENT WITH RECURRENT PULMONARY EMBOLI. IT WAS FURTHER REPORTED THAT IN TEN MONTHS FIVE DAYS POST A FILTER DEPLOYMENT, PATIENT WAS ALLEGEDLY DIAGNOSED WITH THROMBOSIS AND THROMBUS ABOVE THE FILTER. REPORTEDLY, THE DEVICE HAS BEEN REMOVED. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Thrombosis/Thrombus
MAUDE report 18638310 ↗InjuryJan 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT THROUGH THE LEFT COMMON FEMORAL VEIN APPROACH, AN INFERIOR VENA CAVA FILTER WAS DEPLOYED IN A PATIENT AFTER BEING DIAGNOSED WITH CLOT EXTENDING FROM THE RIGHT FEMORAL VEIN TO THE LEVEL OF THE RIGHT EXTERNAL ILIAC VEIN. SIX DAYS POST THE FILTER DEPLOYMENT, IT WAS ALLEGED THAT THE PATIENT ALLEGEDLY EXPERIENCED PULMONARY EMBOLISM. THE DEVICE HAS NOT BEEN REMOVED, AND THERE HAVE BEEN NO REPORTED ATTEMPTS MADE TO RETRIEVE THE FILTER. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Pulmonary Embolism
MAUDE report 18481434 ↗InjuryDec 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT SOMETIMES POST A VENA CAVA FILTER PLACEMENT, THE PATIENT ALLEGEDLY EXPERIENCED INFECTION AND ARTHRITIC SORE IN THE LEG. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 18361651 ↗InjuryDec 11, 2023Malposition of Device; Patient-Device Incompatibility; Malposition of Device; Patient-Device Incompatibility
IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT A VENA CAVA FILTER WAS DEPLOYED VIA RIGHT COMMON FEMORAL VEIN APPROACH FOR A PATIENT WITH HIGH RISK FOR DEEP VEIN THROMBOSIS AND UNABLE TO ANTI-COAGULATE SECONDARY TO INTRACRANIAL BLEEDING. IT WAS FURTHER REPORTED THAT THREE YEARS, SIX MONTHS AND TWENTY-SIX DAYS POST FILTER DEPLOYMENT, THE FILTER WAS ALLEGEDLY TILTED ANTERIORLY WITH ITS CONE LYING ON THE POSTERIOR WALL OF THE INFERIOR VENA CAVA. FURTHERMORE, THE FILTER STRUTS HAD ALLEGEDLY PERFORATED THE INFERIOR VENA CAVA INTO THE PERICAVAL/MESENTERIC FAT AND ONE OF THE LEGS WAS IN CONTACT WITH THE ANTERIOR PERIOSTEUM OF A LUMBAR VERTEBRA. REPORTEDLY, THE PATIENT ALLEGEDLY EXPERIENCED HIP PAIN. HOWEVER, THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Pain
MAUDE report 18299239 ↗InjuryNov 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT A VENA CAVA FILTER WAS DEPLOYED SUCCESSFULLY IN A PATIENT AFTER BEING DIAGNOSED WITH PULMONARY EMBOLISM AND HYPERCOAGULABLE STATE. SIX MONTHS AND SIX DAYS POST FILTER DEPLOYMENT, IT WAS ALLEGED THAT THE PATIENT WAS DIAGNOSED WITH THROMBUS ABOVE THE INFERIOR VENA CAVA FILTER. REPORTEDLY, THE FILTER HAS BEEN REMOVED WITH A SNARE DEVICE UNDER FLUOROSCOPY. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Thrombosis/Thrombus
MAUDE report 18147184 ↗MalfunctionOct 20, 2023Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT PRIOR TO A VENA CAVA FILTER PLACEMENT PROCEDURE, UPON OPENING THE PACKAGE, THE FILTER WAS ALLEGEDLY PREDEPLOYED PARTIALLY INTO THE SHEATH. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17972279 ↗MalfunctionOct 4, 2023Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING PREPARATION OF A VENA CAVA FILTER PLACEMENT PROCEDURE, THE FILTER WAS ALLEGEDLY DETACHED IN THE INTRODUCER SHEATH AND REMAINED IN THE SHEATH. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17866681 ↗InjurySep 8, 2023Difficult to Remove; Obstruction of Flow; Migration; Difficult to Remove; Obstruction of Flow; Migration
IT WAS REPORTED THROUGH THE LITIGATION PROCESS THAT A VENA CAVA FILTER WAS PLACED IN A PATIENT AFTER BEING DIAGNOSED WITH PULMONARY EMBOLISM AND A HYPERCOAGULABLE STATE OF MALIGNANCY. APPROXIMATELY TWO YEARS AND EIGHT MONTHS POST FILTER DEPLOYMENT, SUBSEQUENT IMAGING DEMONSTRATED OCCLUSION OF THE FILTER, AND THE PATIENT ALLEGEDLY EXPERIENCED THROMBOSIS, WHICH REQUIRED THROMBOLYSIS. IT WAS FURTHER REPORTED THAT THE FILTER HAD ALLEGEDLY MIGRATED. REPORTEDLY, THE DEVICE HAS BEEN REMOVED PERCUTANEOUSLY AFTER UNSUCCESSFUL ATTEMPTS. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Thrombosis/Thrombus
MAUDE report 17708890 ↗
FDA clearance and classification
The FDA regulates product code DTK (filter, intravascular, cardiovascular) under 21 CFR 870.3375 as a Class II device. 21 CFR 870.3375 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Denali® Vena Cava Filter DL900F
Has the Denali® Vena Cava Filter DL900F been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801741040801), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Denali® Vena Cava Filter DL900F?
The FDA's MAUDE database lists 530 adverse event reports for this device, received between Oct 21, 2015 and Sep 28, 2026. In those reports, the most common specific device problems are “Malposition of Device” (171), “Patient Device Interaction Problem” (142) and “Difficult to Remove” (123); the most common patient problems are “Pulmonary Embolism” (93), “Abdominal Pain” (70) and “Thrombosis/Thrombus” (28). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Denali® Vena Cava Filter DL900F?
22 reports are classified as deaths, 239 as injuries and 269 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Denali® Vena Cava Filter DL900F MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Denali® Vena Cava Filter DL900F, and is it still sold?
The FDA's device database lists Bard Peripheral Vascular, Inc. as the company responsible for it (brand Denali® Vena Cava Filter, model DL900F). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other FILTER, INTRAVASCULAR, CARDIOVASCULAR devices
- DENALI® Vena Cava Filter DL900JBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR15 deaths491 reports
- Option Elite Retrievable Vena Cava Filter System 352506070EArgon Medical Devices, Inc. · Filter, Intravascular, Cardiovascular2 recalls3 deaths340 reports
- Gunther Tulip G33016Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths348 reports
- Gunther Tulip G33017Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths286 reports
- Cook Celect G34309William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall1 death230 reports
- Gunther Tulip G52916William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths189 reports
- Cook Celect G34502William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths165 reports
- Cook Celect G34505William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall1 death131 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page