Bardex® Lubricath® Foley Catheter 0165GL16Recalls & reported problems
C. R. Bard, Inc. · Bardex® Lubricath® Foley Catheter 5cc Balloon Lubricious Coated · Class 2
All models of Bardex Lubricath Foley Catheter (38)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741046216), received between Apr 12, 2019 and Oct 7, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Deflation Problem2
- Break1
- Component Missing1
- Detachment of Device or Device Component1
- Incorrect Measurement1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 7, 2024Deflation Problem; Deflation Problem
IT WAS REPORTED THAT NURSE TESTED THE BALLOON OF FOLEY CATHETER, AND IT WOULD NOT DEFLATE AFTERWARDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20384920 ↗MalfunctionMay 1, 2019Break; Incorrect Measurement; Component Missing; Detachment of Device or Device Component; Incorrect Measurement
IT WAS REPORTED THAT UPON DEFLATION OF THE FOLEY CATHETER, THE TIP OF THE CATHETER WAS NOT PRESENT AND BELIEVED TO HAVE BEEN BROKEN OFF INSIDE OF THE PATIENT. X-RAY AND CYSTOGRAM WERE PERFORMED BUT THE TIP WAS NOT LOCATED. UPON RECEIPT OF SAMPLE, INVESTIGATORS FOUND CATHETER TO BE INCORRECT MEASUREMENT MEASURING AS AN 18FR.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8567958 ↗InjuryApr 12, 2019Deflation Problem
IT WAS REPORTED THAT THE CATHETER WOULD NOT DEFLATE. THE PATIENT WAS TAKEN TO SURGERY FOR FOREIGN BODY REMOVAL.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 8511183 ↗
Other CATHETER, COUDE devices
- Bardex® Lubricath® Foley Catheter 0165L14C. R. Bard, Inc. · CATHETER, COUDE562 reports
- Bardex® Lubricath® Foley Catheter 0165L16C. R. Bard, Inc. · CATHETER, COUDE539 reports
- Bardia® Foley Catheter Silicone Coated 123516AC. R. Bard, Inc. · CATHETER, COUDE381 reports
- Bardia® Foley Catheter Silicone Coated 123514AC. R. Bard, Inc. · CATHETER, COUDE195 reports
- SureStep™ Foley Tray System Bardex® Lubricath® Foley Catheter A899916C. R. Bard, Inc. · CATHETER, COUDE3 recalls19 reports
- Bardia® Foley Catheter Silicone Coated 123518AC. R. Bard, Inc. · CATHETER, COUDE153 reports
- SureStep™ Foley Tray System Bardex® Lubricath® Foley Catheter A902916C. R. Bard, Inc. · CATHETER, COUDE2 recalls50 reports
- Bard® Lubricath® Foley Catheter 365716C. R. Bard, Inc. · CATHETER, COUDE117 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page