Skip to content

Dorado® DR8098Recalls & reported problems

Bard Peripheral Vascular, Inc. · Dorado® PTA Balloon Dilatation Catheter 9 mm x 80 mm, 80 cm Catheter · Class 2

All models of Dorado (49)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741059476), received between Apr 14, 2017 and Dec 9, 2023 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Detachment of Device or Device Component3
  • Retraction Problem3
  • Difficult to Remove2
  • Inflation Problem1
  • Leak/Splash1
  • Material Rupture1

Patient problems

  • No Consequences Or Impact To Patient5
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 9, 2023Break; Difficult to Remove; Break; Difficult to Remove

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY WOULD NOT COME OUT OF SHEATH. IT WAS FURTHER REPORTED THAT THE BALLOON AND THE SHEATH WAS REMOVED TOGETHER. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18298144 ↗
  • MalfunctionNov 29, 2023Break; Difficult to Remove; Material Rupture; Break; Difficult to Remove; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON HAD ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THERE WAS DIFFICULTY IN RETRACTING THE BALLOON THROUGH THE INTRODUCER SHEATH. REPORTEDLY, THE SHEATH AND THE BALLOON WERE REMOVED AS ONE UNIT AND THE TIP OF THE SHEATH WAS APPEARED TO BE DAMAGED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18225466 ↗
  • MalfunctionJan 16, 2020Break; Inflation Problem; Leak/Splash; Break; Inflation Problem; Leak/Splash

    THIS REPORT SUMMARIZES ONE MALFUNCTION. THE INFORMATION REVIEWED INDICATED THAT MODEL DR8098 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED INFLATION ISSUE, BREAK, AND A LEAK. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED CONSEQUENCES. AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9597967 ↗
  • MalfunctionDec 18, 2017Retraction Problem; Detachment of Device or Device Component; Retraction Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUBCLAVIAN VEIN, RESISTANCE WAS ALLEGEDLY FELT WHEN THE BALLOON WAS RETRACTED TO THE SHEATH AND SUBSEQUENTLY DETACHED FROM THE CATHETER SHAFT. THE PROCEDURE WAS CONCLUDED, AND NO FURTHER TREATMENT WAS PROVIDED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7126345 ↗
  • MalfunctionSep 13, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THE PTA BALLOON ALLEGEDLY DETACHED FROM THE BALLOON CATHETER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6862542 ↗
  • MalfunctionMay 2, 2017Break; Retraction Problem; Break; Retraction Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A BRACHIOCEPHALIC FISTULA, THE HEALTH CARE PROVIDER (HCP) ALLEGEDLY HAD DIFFICULTY REMOVING THE PTA BALLOON THROUGH THE 6FR SHEATH AFTER THE SECOND INFLATION. REPORTEDLY, THE SHEATH AND BALLOON WERE REMOVED AS A SINGLE UNIT. WHILE MAINTAINING THE WIRE ACCESS, THE HCP USED ANOTHER SHEATH AND BALLOON TO COMPLETE THE PROCEDURE. THE PATIENT WAS REPORTED TO BE ASYMPTOMATIC AT THE CONCLUSION OF THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6538277 ↗
  • MalfunctionApr 14, 2017Break; Retraction Problem; Detachment of Device or Device Component; Break; Retraction Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A ARTERIAL VENOUS GRAFT, THE PTA BALLOON ALLEGEDLY HAD DIFFICULTY RETRACTING THROUGH THE 6FR SHEATH. DURING RETRACTION THE BALLOON BROKE INSIDE THE SHEATH AND THE HEALTH CARE PROVIDER REMOVED BOTH BALLOON AND SHEATH TOGETHER AS ONE UNIT. THE PROCEDURE WAS CONCLUDED, AND NO FURTHER TREATMENT WAS PROVIDED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6495635 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates