Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741060038), received between Oct 6, 2016 and Oct 6, 2016 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Material Rupture1
  • Retraction Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 6, 2016Retraction Problem; Material Rupture; Detachment of Device or Device Component; Retraction Problem; Material Rupture; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A LEFT UPPER ARM AV FISTULA, THE PTA BALLOON ALLEGEDLY RUPTURED (ATM UNKNOWN). REPORTEDLY, AS THE HEALTH CARE PROVIDER(HCP) WAS RETRACTING THE BALLOON THROUGH THE SHEATH, A SMALL SEGMENT OF THE BALLOON DETACHED AND TRAVELED PROXIMAL TO THE CEPHALIC VEIN. THE HCP USED A SNARE DEVICE IN AN ATTEMPT TO REMOVE THE BALLOON SEGMENT, BUT WAS UNSUCCESSFUL. THE HCP THEN USED A STENT TO PIN THE BALLOON SEGMENT AGAINST THE VESSEL WALL. REPORTEDLY, THE HCP USED ANOTHER BALLOON TO COMPLETE THE ANGIOPLASTY PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6003097 ↗