Conquest® 40 CQF5064Recalls & reported problems
Bard Peripheral Vascular, Inc. · Conquest® 40 PTA Dilatation Catheter 6 mm x 40 mm, 50 cm catheter · Class 2
All models of Conquest 40 (25)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741060151), received between Jul 16, 2020 and Jul 15, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture2
- Break1
- Burst Container or Vessel1
- Device-Device Incompatibility1
- Difficult to Remove1
- Failure to Advance1
- Leak/Splash1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2025Difficult to Remove; Failure to Advance; Device-Device Incompatibility; Difficult to Remove; Failure to Advance; Device-Device Incompatibility
ON (B)(6) 2025, A 60-YEAR-OLD MALE PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE CONQUEST 40 PTA BALLOON CATHETER FOR CEPHALIC VEIN FISTULA. DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY CANNOT REWIRE. IT WAS FURTHER REPORTED THAT THERE WAS A DIFFICULTY IN RETRACTING THE BALLOON THRU THE SHEATH. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22505572 ↗MalfunctionApr 11, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21823856 ↗MalfunctionJan 31, 2025Burst Container or Vessel; Leak/Splash; Burst Container or Vessel; Leak/Splash
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THERE WAS LEAKAGE AT THE Y-CONNECTION DURING BALLOON PRESSURIZATION. REPORTEDLY, THE Y BASE OF THE CATHETER WAS ALLEGEDLY RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21280747 ↗MalfunctionApr 27, 2023Break; Break
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PROXIMAL TIP OF THE PTA BALLOON ALLEGEDLY HAD A CRACK. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16822795 ↗MalfunctionJul 16, 2020Peeled/Delaminated; Material Rupture; Peeled/Delaminated; Material Rupture
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQF5064 PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED PINHOLE RUPTURE AND PEELED PEBAX. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. OF THE ONE MALFUNCTION, ONE PATIENT WAS INVOLVED WITH NO PATIENT CONSEQUENCE OR IMPACT. THE AGE, WEIGHT AND GENDER OF THE PATIENT WAS NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10286882 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page