Conquest® 40 CQF7574Recalls & reported problems

Bard Peripheral Vascular, Inc. · Conquest® 40 PTA Dilatation Catheter 7 mm x 40 mm, 75 cm catheter · Class 2

Quick summary

This Conquest40 balloon dilatation catheter is a Bard Peripheral Vascular device used to open narrowed blood vessels during angioplasty. The 56 reports mostly describe balloon problems: failing to deflate, bursting during inflation, and being hard to withdraw or remove, plus a few reports of wet or damaged packaging. One report describes a vessel rupture with bleeding; no deaths were reported. There are no recalls naming this device, and reports have dropped sharply, from 3 in the prior12 months to 1.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
55
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 55 adverse event reports that list this exact device (DI 00801741060496), received between Oct 12, 2016 and Sep 22, 2025 — classified as 1 injury, 54 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem38
  • Material Rupture17
  • Retraction Problem8
  • Difficult to Remove4
  • Inflation Problem3
  • Device Contamination with Chemical or Other Material2
  • Material Frayed2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Consequences Or Impact To Patient40
  • No Clinical Signs, Symptoms or Conditions13
  • Hemorrhage/Blood Loss/Bleeding1
  • No Patient Involvement1
  • Perforation of Vessels1

Recent adverse event reports

The 25 most recent of 55 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2025Device Contamination with Chemical or Other Material; Packaging Problem; Device Contamination with Chemical or Other Material; Packaging Problem

    ON (B)(6) 2025, A PATIENT UNDERWENT FORA ANGIOPLASTY PROCEDURE USING CONQUEST 40. PIOR TO AN ANGIOPLASTY PROCEDURE USING THE CONQUEST 40 PTA BALLOON, INNER AND OUTER PACKAGES ALLEGEDLY ARRIVED WET. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23107239 ↗
  • MalfunctionSep 12, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    ON (B)(6) 2025, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE CONQUEST 40 PTA BALLOON. IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING ALLEGEDLY HAVE WATER DAMAGE. IT WAS FURTHER REPORTED THAT THE WATER MARK IS BELOW THE STERILE PACKAGE LINE. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23037208 ↗
  • InjurySep 2, 2025Material Rupture; Material Rupture

    ON (B)(6) 2025, THE PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE UTILIZING THE CONQUEST 40 PTA BALLOON. DURING THIS PROCEDURE, THE BALLOON REPORTEDLY RUPTURED. THE PATIENT IS SAID TO HAVE EXPERIENCED A VESSEL RUPTURE AND SUBSEQUENT BLEEDING. IT HAS BEEN REPORTED THAT ANOTHER BALLOON WAS EMPLOYED TO FINALIZE THE PROCEDURE. THE CURRENT STATUS OF THE PATIENT REMAINS UNKNOWN.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Perforation of Vessels

    MAUDE report 22931792 ↗
  • MalfunctionAug 21, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT AROUND 29 OR 30 ATM. IT WAS FURTHER REPORTED THAT THE BALLOON WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20018562 ↗
  • MalfunctionMay 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE TIP OF THE PTA BALLOON ALLEGEDLY BROKE OFF WHILE THE BALLOON WAS BEING FLUSHED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19368879 ↗
  • MalfunctionOct 17, 2023Material Rupture; No Apparent Adverse Event; Material Rupture; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE FISTULA PUNCTURE AREA, WHEN SUBJECT TO THE FIRST PRESSURIZATION TO 30 ATM, THE PTA BALLOON WAS INFLATED AND ALLEGEDLY GOT FRACTURED. IT WAS FURTHER REPORTED THAT THE BALLOON WAS DEFLATED AND REMOVED DIRECTLY THROUGH SHEATH. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17951961 ↗
  • MalfunctionOct 16, 2023Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, DURING THE FIRST PRESSURIZATION TO 20 ATM, THE PTA BALLOON WAS ALLEGEDLY NOT INFLATED COMPLETELY. IT WAS FURTHER REPORTED THAT WITH CONTINUOUS PRESSURIZATION, THE PTA BALLOON ALLEGEDLY LEAKED LIQUID. FURTHERMORE, THE BALLOON WAS DEFLATED AND REMOVED DIRECTLY THROUGH THE SHEATH. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17937272 ↗
  • MalfunctionOct 13, 2023Deflation Problem; Inflation Problem; Leak/Splash; Deflation Problem; Inflation Problem; Leak/Splash

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, DURING THE SECOND INFLATION, THE PTA BALLOON WAS ALLEGEDLY DEFLATED ON ITS OWN WHEN THE PRESSURE WAS INCREASED TO AROUND 35 ATM. IT WAS FURTHER REPORTED THAT THE BALLOON WAS REMOVED ROTATABLY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17928683 ↗
  • MalfunctionMar 2, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16465513 ↗
  • MalfunctionSep 13, 2021Difficult to Insert; Difficult to Remove; Material Rupture; Difficult to Insert; Difficult to Remove; Material Rupture

    IT WAS REPORTED DURING AN ANGIOPLASTY PROCEDURE AT LEFT BRACHYCEPHALIC FISTULA, THE BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THE BALLOON WAS ALLEGEDLY DIFFICULT TO BE REMOVED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12458101 ↗
  • MalfunctionJun 9, 2021Material Rupture; Material Rupture

    IT WAS REPORTED DURING THE ANGIOPLASTY PROCEDURE AT LEFT BRACHY CEPHALIC FISTULA, THE BALLOON WAS ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11963879 ↗
  • MalfunctionJun 8, 2021Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture

    IT WAS REPORTED DURING AN ANGIOPLASTY PROCEDURE AT LEFT BRACHYCEPHALIC FISTULA, THE BALLOON ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THE BALLOON WAS ALLEGEDLY DIFFICULT TO BE REMOVED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11953753 ↗
  • MalfunctionJan 11, 2021Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY HAD A DEFLATION ISSUE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11145269 ↗
  • MalfunctionOct 28, 2020Inflation Problem; Material Rupture; Inflation Problem; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL CQF7574 PTA BALLOON DILATATION CATHETER ALLEGEDLY INFLATION PROBLEM AND MATERIAL RUPTURE. THE INFORMATION WAS RECEIVED FROM A SINGLE SOURCE. ONE PATIENT WAS INVOLVED WITH NO REPORTED PATIENT INJURY. THE 57 YEAR OLD FEMALE PATIENT WAS 50 KGS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10749764 ↗
  • MalfunctionOct 26, 2020Deflation Problem; Deflation Problem

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQF7574 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED DEFLATION PROBLEM. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE AGE, WEIGHT AND GENDER OF THE PATIENT WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10738191 ↗
  • MalfunctionOct 23, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AT 16 ATM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10726060 ↗
  • MalfunctionJul 16, 2020Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL CQF7574 PTA DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE AND RETRACTION ISSUE. THIS REPORT WAS RECEIVED FROM ONE SOURCE. ONE PATIENT WAS INVOLVED WITH NO REPORTED PATIENT INJURY. THE PATIENTS¿ AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10286719 ↗
  • MalfunctionJan 16, 2020Deflation Problem; Deflation Problem

    THIS REPORT SUMMARIZES FOUR MALFUNCTIONS. A REVIEW OF THE MALFUNCTIONS INDICATED THAT MODEL CQF7574 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED DEFLATION ISSUES. THIS REPORT WAS RECEIVED FROM VARIOUS SOURCES. THE MALFUNCTIONS INVOLVED 4 PATIENTS WITH NO PATIENT CONSEQUENCES. THE FOUR PATIENTS RANGED FROM 58-68 YEARS OF AGE AND WEIGHT RANGE 165-194 LB. OF THE REPORTED PATIENTS 2 WERE MALES AND 2 WERE FEMALE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9596757 ↗
  • MalfunctionJan 16, 2020Material Rupture; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL CQF7574 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THE INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO PATIENT INJURY. THE MALE PATIENT'S AGE AND WEIGHT WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9596601 ↗
  • MalfunctionOct 16, 2019Burst Container or Vessel; Material Rupture; Burst Container or Vessel; Material Rupture

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQF7574 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED BURST AND MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. ONE PATIENT WAS INVOLVED AND THERE WAS NO REPORTED PATIENT INJURY. THE PATIENT IS MALE, (B)(6) YEARS OLD, AND (B)(6) POUNDS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9198424 ↗
  • MalfunctionJul 30, 2019Material Frayed; Material Rupture; Material Deformation; Material Frayed; Material Rupture; Material Deformation

    THIS REPORT SUMMARIZES 2 MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQF7574 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED A MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. OF THE 2 RUPTURES, BOTH DID NOT INVOLVE PATIENTS. PATIENT AGE, WEIGHT, AND GENDER WAS NOT PROVIDED FOR EITHER PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 8843031 ↗
  • MalfunctionJul 22, 2019Deflation Problem; Deflation Problem

    THIS REPORT SUMMARIZES ONE (1) MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL CQF7574 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED A DEFLATION ISSUE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE (B)(6) MALE PATIENT WAS (B)(6).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8812358 ↗
  • MalfunctionJan 7, 2019Deflation Problem; Material Rupture; Deflation Problem; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A DECLOT THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8224600 ↗
  • MalfunctionJan 7, 2019Deflation Problem; Material Rupture; Deflation Problem; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A VENOUS ANASTOMOSIS THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8224565 ↗
  • MalfunctionJan 7, 2019Deflation Problem; Material Frayed; Material Rupture; Deflation Problem; Material Frayed; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A DECLOT THE PTA BALLOON WAS ALLEGEDLY DIFFICULT TO DEFLATE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8224442 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →