Atlas® AT75126Recalls & reported problems

Bard Peripheral Vascular, Inc. · Atlas® PTA Balloon Dilatation Catheter 12 mm x 60 mm, 75 cm Catheter · Class 2

All models of Atlas (22)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741062414), received between Oct 25, 2017 and Aug 30, 2024 — classified as 5 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem7
  • Inflation Problem2
  • Difficult to Remove1
  • Failure to Fold1
  • Material Deformation1
  • Material Rupture1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient2
  • Needle Stick/Puncture1
  • No Consequences Or Impact To Patient1
  • Obstruction/Occlusion1
  • Pain1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2024Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON WAS ALLEGEDLY UNABLE TO DEFLATE. IT WAS FURTHER REPORTED THAT THE BALLOON WAS POPPED USING A NEEDLE TO REMOVE IT FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20104947 ↗
  • MalfunctionAug 8, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15181922 ↗
  • MalfunctionJan 25, 2022Deflation Problem; Inflation Problem; Difficult to Remove; Material Deformation; Deflation Problem; Inflation Problem; Difficult to Remove; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY FAILED TO INFLATE. IT WAS FURTHER REPORTED THAT THE DISTAL END OF THE BALLOON WAS FAILED TO EXPAND. REPORTEDLY USER TRIED TO FLAT THE BALLOON, BUT THE BALLOON WAS UNABLE TO BE DEFLATED. THE BALLOON WAS REMOVED FROM PATIENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13340806 ↗
  • MalfunctionJan 24, 2022Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DEVICE ALLEGEDLY HAD INFLATION ISSUE. IT WAS FURTHER REPORTED THAT THE DEVICE ALLEGEDLY HAD DEFLATION ISSUE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13332552 ↗
  • InjuryJan 19, 2022Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE VIA POPLITEAL VEIN, THE PTA BALLOON ALLEGEDLY HAD DEFLATION ISSUE. IT WAS FURTHER REPORTED THAT POST REMOVAL OF THE BALLOON, THE PATIENT'S ILIAC AND FEMORAL VEIN WERE FOUND TO HAVE THROMBOTIC OCCLUSION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Obstruction/Occlusion

    MAUDE report 13289451 ↗
  • InjurySep 17, 2020Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE VIA FEMORAL VEIN, THE PTA BALLOON WAS ALLEGEDLY FAILURE TO DEFLATE. IT WAS FURTHER REPORTED THAT ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10546542 ↗
  • InjuryMay 11, 2018Deflation Problem; Deflation Problem

    MEDICAL RECORDS RECEIVED AND REVIEWED. DURING A RECANALIZATION PROCEDURE, AFTER MULTIPLE INFLATIONS IN THE EXTERNAL ILIAC VEIN, A PTA BALLOON WAS UNABLE TO BE DEFLATED. A NEEDLE WAS INSERTED INTO THE BALLOON UNDER FLUOROSCOPY WHICH RESULTED IN GOOD DECOMPRESSION, AND THE DEVICE WAS EASILY REMOVED. TWO STENTS WERE PLACED AND ADDITIONAL VENOPLASTY WAS PERFORMED TO COMPLETE THE PROCEDURE. THE PATIENT WAS HOSPITALIZED OVERNIGHT FOR OBSERVATION.

    Patient problems: Needle Stick/Puncture; No Known Impact Or Consequence To Patient

    MAUDE report 7509350 ↗
  • InjuryNov 8, 2017Failure to Fold; Retraction Problem; Failure to Fold; Retraction Problem

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THE PTA BALLOON ALLEGEDLY FOLDED ON ITSELF AND THEN THE BALLOON MATERIAL BUNCHED AT THE DISTAL END. REPORTEDLY, THE BUNCHED BALLOON MATERIAL ALLEGEDLY GOT STUCK INSIDE THE SHEATH DURING REMOVAL. IT WAS FURTHER REPORTED THAT THE PATIENT WAS SENT TO THE OR FOR SURGICAL REMOVAL OF THE BALLOON. PATIENT RECOVERED WITHOUT FURTHER INTERVENTION AND WAS DISCHARGED THE NEXT DAY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7011830 ↗
  • InjuryOct 25, 2017Deflation Problem

    WHILE TRYING TO REMOVE AN ATLAS PTA BALLOON DILATION CATHETER, IT WOULD NOT DEFLATE COMPLETELY. THE PATIENT HAD TO BE TAKEN TO THE OPERATING ROOM (OR) FOR SURGICAL REMOVAL. SOME PAIN WAS NOTED IN PATIENT CHART, BUT OVERALL THE PATIENT RECOVERED WITHOUT FURTHER INTERVENTION AND WAS DISCHARGED THE NEXT DAY. OTHER BALLOONS OF THIS TYPE WERE REMOVED FROM THE SHELVES BY THE NURSE MANAGER AND THE REP WAS NOTIFIED TO COME IN. AT THIS TIME, MORE INFORMATION IS BEING COLLECTED ON SPECIFIC DEVICE DETAILS, WHETHER THE DEVICE WAS SAVED, AND WHAT COMMUNICATION HAS OCCURRED BETWEEN THE MANUFACTURER AND THE INTERVENTIONAL RADIOLOGY DEPARTMENT. PER SITE REPORTER: THE MANUFACTURER REP WAS NOTIFIED AND INSTRUCTED TO COME IN BY THE INTERVENTIONAL RADIOLOGY DEPARTMENT. IT IS UNKNOWN AT THIS TIME WHETHER OR NOT THE DEVICE WAS SAVED AND WILL BE PASSED TO THE MANUFACTURER REP.

    Patient problems: Pain

    MAUDE report 6975764 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →