Bard Peripheral Vascular, Inc. · Seeker® Crossing Support Catheter 0.014" Guidewire compatible, 135cm · Class 2

All models of Seeker (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741067877), received between Jul 11, 2019 and Sep 2, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Break1
  • Device Markings/Labelling Problem1
  • Device-Device Incompatibility1
  • Failure to Advance1
  • Material Deformation1
  • Material Split, Cut or Torn1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2026Peeled/Delaminated; Device-Device Incompatibility

    On (b)(6) 2026, a patient underwent a recanalization procedure using a Seeker Crossing Support Catheter for the treatment of a calcified superficial femoral artery (SFA) lesion via a contralateral approach. During the procedure, it was reported that the guidewire could not be advanced through the distal tip of the catheter. After multiple attempts, the guidewire was successfully advanced; however, difficulty was subsequently encountered during withdrawal of the guidewire through the distal tip. Upon removal, flaking of the guidewire's polymer coating was observed. There was no reported patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50559365 ↗
  • MalfunctionApr 8, 2024Retraction Problem; Failure to Advance; Material Deformation; Retraction Problem; Failure to Advance; Material Deformation

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA THE CONTRALATERAL APPROACH, BECAUSE THE BIFURCATION ANGLE WAS TOO TIGHT, THE CATHETER HAD ALLEGEDLY KINKED. IT WAS FURTHER REPORTED THAT THE CATHETER ALLEGEDLY COULD NOT ABLE TO CROSS THE LESION. FURTHERMORE, THE DEVICE ALLEGEDLY HAD RETRACTION PROBLEM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19057609 ↗
  • MalfunctionJan 23, 2024Detachment of Device or Device Component; Material Split, Cut or Torn; Detachment of Device or Device Component; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE SUPPORT CATHETER ALLEGEDLY TORN OFF IN A SECTION OF THE CENTIMETER TIP WITHOUT BEING ABLE TO REACH THE INTRODUCER SHEATH. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18565644 ↗
  • MalfunctionDec 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE TIP OF THE DEVICE WAS ALLEGEDLY WRENCHED OFF. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18333990 ↗
  • MalfunctionJan 6, 2022Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT PRIOR TO A RECANALIZATION PROCEDURE, THE PHYSICIAN ALLEGEDLY NOTED THAT THE POSITION OR THE LENGTH OF THE MARKER SEEMED TO BE DIFFERENT. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13173902 ↗
  • MalfunctionJul 11, 2019Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES ONE (1) MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL SK13514 SEEKER CROSSING SUPPORT CATHETER ALLEGEDLY EXPERIENCED A BREAK AND TIP DETACHMENT. THIS INFORMATION WAS RECEIVED FROM * ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. PATIENT AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient; No Patient Involvement

    MAUDE report 8783027 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →