Seeker® SK15018Recalls & reported problems

Bard Peripheral Vascular, Inc. · Seeker® Crossing Support Catheter 0.018" Guidewire compatible, 150cm · Class 2

All models of Seeker (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Bard Seeker crossing support catheter is a thin tube used to help a guidewire pass through blocked leg arteries during recanalization and angioplasty procedures. Most of the 29 reports describe the catheter breaking, snapping apart, or getting stuck and being difficult to remove, sometimes leaving a broken piece inside the patient; some also mention damaged packaging. Two reports involved injuries and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00801741067914), received between Oct 25, 2018 and Sep 2, 2026 — classified as 2 injuries, 28 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component14
  • Packaging Problem9
  • Break7
  • Retraction Problem5
  • Material Deformation3
  • Device-Device Incompatibility2
  • Difficult to Open or Remove Packaging Material2
  • Defective Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • Foreign Body In Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2026Detachment of Device or Device Component

    On (b)(6) 2026, a patient underwent a recanalization procedure using a Seeker Support Catheter via femoral access. During the procedure, it was reported that the catheter ruptured while being advanced over a 0.018-inch guidewire. The catheter separated into two pieces. The procedure was subsequently completed using a competitor's catheter. No patient harm was reported, and there was no reported patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50559113 ↗
  • MalfunctionJul 27, 2026Detachment of Device or Device Component

    On (b)(6)2026, a patient underwent a recanalization procedure using the Seeker catheter. During the procedure, the catheter was allegedly found fractured. There was no reported patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50032306 ↗
  • MalfunctionFeb 17, 2026Retraction Problem; Detachment of Device or Device Component; Device Dislodged or Dislocated; Material Deformation

    ON (B)(6) 2026, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING SEEKER. DURING THE PROCEDURE, THE SEEKER CATHETER WAS ALLEGEDLY UNABLE TO REMOVE AND NEEDED TO REMOVE FORCEFUL AND OBSERVED STRUCTURAL DAMAGE TO THE CATHETER WITH THE RADIOPAQUE RING. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24362861 ↗
  • InjuryJan 13, 2026Detachment of Device or Device Component

    ON (B)(6) 2025, A PATIENT UNDERWENT A ANGIOPLASTY PROCEDURE USING THE SEEKER CATHETER. DURING THE PROCEDURE, THE SEEKER ALLEGEDLY BROKEN COMPLETELY. IT WAS FURTHER REPORTED THAT THE DETACHED PIECE WAS LEFT INSIDE THE PATIENT. THE CURRENT STATUS OF PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 24050667 ↗
  • MalfunctionSep 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ON (B)(6) 2025, THE PATIENT UNDERWENT A RECANALIZATION PROCEDURE IN RIGHT COMMON FEMORAL ARTERY USING THE SEEKER. DURING THE PROCEDURE IT WAS REPORTED THAT THE GUIDEWIRE WAS ALLEGEDLY UNABLE TO PASS THROUGH THE OCCLUDED LESION. IT WAS FURTHER REPORTED THAT THE HEAD END OF THE SUPPORT CATHETER BROKE, DURING THE OPENING PROCESS. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22931079 ↗
  • MalfunctionJul 2, 2025Break; Break

    ON (B)(6) 2025, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE SEEKER CATHETER. PRIOR TO THE PROCEDURE, THE SUPPORT CATHETER SEEKER WAS ALLEGEDLY FOUND TO BE BROKEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22380747 ↗
  • MalfunctionApr 4, 2025Detachment of Device or Device Component; Difficult to Open or Remove Packaging Material; Detachment of Device or Device Component; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT PRIOR TO A RECANALIZATION PROCEDURE, THE CATHETER WAS ALLEGEDLY STUCK AND COULD NOT BE REMOVED WHEN OUTER PACKAGE WAS OPENED. IT WAS FURTHER REPORTED THAT THE CATHETER WAS BROKEN AND DETACHED BY APPLYING A LITTLE FORCE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21774842 ↗
  • InjuryMar 11, 2025Retraction Problem; Detachment of Device or Device Component; Retraction Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE SUPPORT CATHETER WAS ALLEGEDLY DISCONNECTED DURING WITHDRAWAL. IT WAS FURTHER REPORTED THAT THE CATHETER WAS BROKEN AND LAGGING IN THE ARTERY. REPORTEDLY, THE ANTERIOR TIBIAL ARTERY WAS NARROWED, SO THE INTERVENTIONAL DEVICE COULD NOT PASS THROUGH, AND IT COULD ONLY BE REMOVED SURGICALLY, AND THE OPERATION WAS TERMINATED AFTER OPENING THE POSTERIOR TIBIAL AND PERONEAL ARTERIES. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 21570056 ↗
  • MalfunctionNov 22, 2024Break; Detachment of Device or Device Component; Device-Device Incompatibility; Break; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY OF THE LOWER LIMB, THE CATHETER WAS ALLEGEDLY JAMMED DURING ADVANCEMENT. IT WAS FURTHER REPORTED THAT A SNAG OCCURS THE CATHETER CANNOT BE ALLEGEDLY WITHDRAWN SMOOTHLY ALONG THE GUIDEWIRE. REPORTEDLY, THE GUIDEWIRE AND CATHETER WERE WITHDRAWN FROM THE BODY, AND THE BROKEN PART WAS WITHDRAWN TOGETHER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20755301 ↗
  • MalfunctionOct 19, 2024Material Separation; Stretched; Detachment of Device or Device Component; Material Separation; Stretched; Detachment of Device or Device Component

    IT WAS REPORTED THAT PRIOR TO A RECANALIZATION PROCEDURE, THE CATHETER HAS BEEN ALLEGEDLY BROKEN. IT WAS FURTHER REPORTED THAT THE CONE HEAD FALLS OFF AND SEPARATES FROM THE CATHETER. FURTHERMORE, THE CONE HEAD STILL CONNECTED TO THE CORE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20488942 ↗
  • MalfunctionSep 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT PRIOR TO A RECANALIZATION PROCEDURE IN THE LOWER LIMB ARTERY OPENING, THE SUPPORT CATHETER WAS REMOVED FROM THE PROTECTIVE SLEEVE AND THE TIP WAS ALLEGEDLY FOUND TO BE FULLY SEPARATED FROM THE CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20115796 ↗
  • MalfunctionJun 28, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE IN THE OCCLUDED LOWER LIMB ARTERY, THE CATHETER ALLEGEDLY BROKE DURING THE FLUSHING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19636875 ↗
  • MalfunctionFeb 15, 2024Break; Material Puncture/Hole; Detachment of Device or Device Component; Device-Device Incompatibility; Break; Material Puncture/Hole; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE IN RIGHT FEMORAL ARTERY, THE GUIDEWIRE ALLEGEDLY FAILED TO PASS THROUGH THE SUPPORT CATHETER. IT WAS FURTHER REPORTED THAT THE GUIDEWIRE ALLEGEDLY POKED OUT AT THE HALF-SECTION OF THE SUPPORT CATHETER. REPORTEDLY, THE SUPPORT CATHETER WAS ALLEGEDLY BROKEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18708711 ↗
  • MalfunctionSep 27, 2023Break; Difficult to Open or Remove Packaging Material; Material Deformation; Break; Difficult to Open or Remove Packaging Material; Material Deformation

    IT WAS REPORTED THAT PRIOR TO A RECANALIZATION PROCEDURE, THE CATHETER WAS ALLEGEDLY COULD NOT BE REMOVED FROM THE PROTECTOR. IT WAS FURTHER REPORTED THAT AFTER REMOVAL, THE CATHETER WAS FOUND TO BE BROKEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17824731 ↗
  • MalfunctionMay 30, 2023Defective Component; Device Contamination with Chemical or Other Material; Defective Component; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT DURING PREPARATION FOR A RECANALIZATION PROCEDURE, RED SLIM SUBSTANCE WAS ALLEGEDLY NOTICED AS IT WAS REMOVED FROM THE LOOP. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17017780 ↗
  • MalfunctionDec 16, 2022Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, CATHETER ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT RUPTURED CATHETER TIP WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15993801 ↗
  • MalfunctionFeb 10, 2022Retraction Problem; Material Twisted/Bent; Retraction Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE DEVICE ALLEGEDLY BENT. IT WAS FURTHER REPORTED THAT THE DEVICE WAS STUCK INSIDE THE SHEATH AND THERE WAS DIFFICULTY REMOVING. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13504016 ↗
  • MalfunctionSep 8, 2021Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING WAS ALLEGEDLY DAMAGED AND THE BOXES ARE DENTED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12437543 ↗
  • MalfunctionSep 8, 2021Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING OF THE DEVICE WAS ALLEGEDLY DAMAGED. IT WAS FURTHER REPORTED THAT DURING TRANSPORTATION, IT WAS FOUND THAT THE PRODUCTS WERE NOT PACKED IN BOXES WHICH CAUSED THE BOXES TO BE DENTED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12437538 ↗
  • MalfunctionSep 7, 2021Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING OF THE DEVICE WAS ALLEGEDLY DAMAGED. IT WAS FURTHER REPORTED THAT DURING TRANSPORTATION, IT WAS FOUND THAT THE PRODUCTS WERE NOT PACKED IN BOXES WHICH CAUSED THE BOXES TO BE DENTED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12434743 ↗
  • MalfunctionSep 7, 2021Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING OF THE DEVICE WAS ALLEGEDLY DAMAGED. IT WAS FURTHER REPORTED THAT DURING TRANSPORTATION, IT WAS FOUND THAT THE PRODUCTS WERE NOT PACKED IN BOXES WHICH CAUSED THE BOXES TO BE DENTED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431932 ↗
  • MalfunctionSep 7, 2021Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING OF THE DEVICE WAS ALLEGEDLY DAMAGED. IT WAS FURTHER REPORTED THAT DURING TRANSPORTATION, IT WAS FOUND THAT THE PRODUCTS WERE NOT PACKED IN BOXES WHICH CAUSED THE BOXES TO BE DENTED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431883 ↗
  • MalfunctionSep 7, 2021Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING OF THE DEVICE WAS ALLEGEDLY DAMAGED. IT WAS FURTHER REPORTED THAT DURING TRANSPORTATION, IT WAS FOUND THAT THE PRODUCTS WERE NOT PACKED IN BOXES WHICH CAUSED THE BOXES TO BE DENTED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431877 ↗
  • MalfunctionSep 7, 2021Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING OF THE DEVICE WAS ALLEGEDLY DAMAGED. IT WAS FURTHER REPORTED THAT, THE PRODUCTS WERE NOT PACKED IN BOXES WHICH CAUSED THE BOXES TO BE DENTED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431875 ↗
  • MalfunctionSep 7, 2021Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE PACKAGING OF THE DEVICE WAS ALLEGEDLY DAMAGED. IT WAS FURTHER REPORTED THAT DURING TRANSPORTATION, IT WAS FOUND THAT THE PRODUCTS WERE NOT PACKED IN BOXES WHICH CAUSED THE BOXES TO BE DENTED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12431783 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page