Skip to content

Wideband™ Self-Adhering Male External Catheter 36101Recalls & reported problems

C. R. Bard, Inc. · Wideband™ Self-Adhering Male External Catheter · Class 1

All models of Wideband Self-Adhering Male External Catheter (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741070976), received between Nov 6, 2017 and Nov 10, 2025 — classified as 2 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Nonstandard Device” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Nonstandard Device6
  • Product Quality Problem2
  • Expiration Date Error1
  • Inadequacy of Device Shape and/or Size1
  • Insufficient Device Problem Information1
  • Material Deformation1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient2
  • No Patient Involvement2
  • Pain1
  • Pressure Sore/Ulcer1
  • Skin Tears1
  • Urinary Tract Infection1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 10, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE WOUND CARE NURSE AT THE (B)(6) CENTER. THEY CURRENTLY RECEIVED BD CONDOM CATHETERS FROM THEIR DISTRIBUTORS FOR USE ON THEIR VETERANS BUT DO NOT SEEM TO HAVE SIZERS FOR PROPER MEASURING TO APPLY THE CORRECT SIZE PER PATIENT. THEY HAVE BEEN EXPERIENCING AN INCREASE IN THE NUMBER OF MEDICAL DEVICE RELATED PRESSURE INJURIES LATELY WITH THE USE OF CONDOM CATHETERS AND BELIEVE IT WAS IN RELATION TO NOT SECURING TO CORRECT SIZE. NO MEDICAL INTERVENTION WAS REPORTED. CUSTOMER EMAIL RESPONSE RECEIVED ON 04NOV2025. IT WAS REPORTED THAT, "THE LOT NUMBERS FOR THE CONDOM CATHETERS THAT WERE USED MOST OFTEN ARE AS FOLLOWS: 29MM REF 36102 LOT NUMBER: JUKQ1552, 25MM REF 36101 LOT NUMBER: JUKV1982. THERE WERE NO DEFECTS IN THE PRODUCTS. CUSTOMER CAN ATTRIBUTE THE MINOR SKIN INJURIES TO USE OF THE INCORRECT SIZE DEVICE FOR SELECTED PATIENTS. IT WAS BROUGHT TO CUSTOMER'S ATTENTION THAT THE NURSING STAFF WERE NOT PROVIDED WITH ACCURATE SIZING GUIDES TO ASSIST THEIR CHOICES. THESE INJURIES WERE CHARACTERIZED AS MEDICAL DEVICE RELATED PRESSURE INJURIES FOR PATIENTS, NOW HEALED AND LONG DISCHARGED.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 23518896 ↗
  • MalfunctionAug 4, 2022Expiration Date Error; Expiration Date Error

    IT WAS REPORTED THAT THE MALE EXTERNAL CATHETER SLIPS OFF BECAUSE IT WAS NOT ADHESIVE ENOUGH. STATED THAT THE CATHETER APPEARED THINNER. ALSO STATED THAT THE CUSTOMER RECEIVED THE ORDER FROM BYRAM 30 DAYS AGO, BUT THE CATHETER HAD AN EXPIRATION DATE OF 12/20/2016. PER FOLLOW UP VIA PHONE ON (B)(6) 2022, IT WAS REPORTED THAT THE SITUATION HAS NOT BEEN RESOLVED. PATIENT CONFINED TO A WHEELCHAIR SINCE PARTIALLY PARALYZED. PATIENT COULD NOT MOVE AROUND AT NIGHT WHEN IN BED AND WOKE UP ONE MORNING COMPLETELY DRENCHED IN URINE. THIS HAS OCCURRED WHILE THEY USE WHEELCHAIR TOO. STATED THAT THE CATHETERS APPEAR DEFECTIVE RIGHT OUT OF THE BOX. CUSTOMER PROVIDED TWO CATALOG NUMBERS WERE PCN-36101 WITH LOT# JUGN9178- AND PCN-36301-PROVIDED THROUGH BYRAM HEALTHCARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15166713 ↗
  • MalfunctionMay 11, 2022Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT THE CUSTOMER HAVE BEEN EXPERIENCING AN ADHESIVENESS ISSUE WITH THE MALE EXTERNAL CATHETER. THIS ISSUE HAS OCCURRED MULTIPLE TIMES AND HAS CAUSED BLEEDING. ADDITIONALLY, THE CUSTOMER THOUGHT THE LAST REPLACEMENTS THEY RECEIVED MAY HAVE BEEN EXPIRED AND CONTRIBUTED TO THIS ISSUE. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 15APR2022, CUSTOMER RETURNED ADDITIONAL MALE EXTERNAL CATHETER FROM DIFFERENT LOT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14359091 ↗
  • MalfunctionMar 8, 2022Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCING AN ADHESIVENESS ISSUE WITH THE MALE EXTERNAL CATHETER. THIS ISSUE HAS OCCURRED MULTIPLE TIMES AND CAUSED BLEEDING. ADDITIONALLY, THE CUSTOMER THOUGHT THE LAST REPLACEMENTS THEY RECEIVED MAY HAVE BEEN EXPIRED AND CONTRIBUTED TO THIS ISSUE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13694178 ↗
  • MalfunctionNov 3, 2021Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT THE CUSTOMER HAD ISSUES WITH THE MALE EXTERNAL CONDOM CATHETER AND HAD DIFFICULTY WHILE REMOVING THE CATHETER. ALSO STATED THAT THE RESIDUE STAYED ON THE PATIENT¿S ANATOMY AND WAS HARD TO REMOVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12745356 ↗
  • MalfunctionJun 17, 2021Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT MALE EXTERNAL CATHETER WAS STICKY AND THE GLUE ON THE CONDOM HAD TURNED WHITE. WHEN CUSTOMER REMOVED THE CATHETER, THE GLUE STAYED STUCK ON THE SKIN AND CAUSED MINOR TEARING. NO MEDICAL INTERVENTION WAS REPORTED. PER FOLLOW UP CALL ON (B)(6) 2021, INITIAL REPORTER STATED THE CUSTOMER HAD TO GENTLY REMOVE USING ALCOHOL OR OTHER SOLUTION TO ASSIST WITH REMOVAL. THE CUSTOMER HAD NEVER SEEN THE GLUE GET WHITE LIKE IT WAS. NO MEDICAL CARE HAS BEEN REPORTED. PER CUSTOMER VIA PHONE ON (B)(6) 2021, THE CUSTOMER STATED THAT THE WHITE SUBSTANCE WAS BEING LEFT ON THE PENIS, AND IT WAS SOMEWHAT STICKY LIKE THE GLUE WAS LEFT BEHIND. ALSO, THE CUSTOMER FELT LIKE THE CATHETERS WERE OLD BECAUSE THEY HAD SMALL PRINT ON THE PACKAGING, BUT THE BETTER ONES HAVE LARGE PRINT. IT WAS STATED THAT WHEN TRIED TO REMOVE THE WHITE POWER THAT WAS LEFT ON THE PENIS, THE RUBBING CAUSED THE SKIN TO COME OFF AND CAUSED SOME BLEEDING. THE CUSTOMER DID NOT GO TO THE DOCTOR AND REPORTED THAT THE CUSTOMER WAS HEALING SLOWLY. PER INVESTIGATOR NOTIFICATION RECEIVED ON (B)(6) 2021, DURING SAMPLE EVALUATION A WEAK ADHESIVE WAS FOUND ON THE CATHETER. PER CUSTOMER VIA PHONE ON (B)(6) 2021, THE CUSTOMER

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Skin Tears

    MAUDE report 12014285 ↗
  • MalfunctionMar 27, 2021Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT MALE EXTERNAL CATHETER HAD TOO ADHESIVE WHICH CAUSED THE PATIENT PAIN WHILE REMOVING THE CATHETER . THE CUSTOMER PREFERRED NEXT SIZE 27 MM, BUT STILL PATIENT FEARED IF THE ADHESIVE FOR THAT ONE WAS JUST AS STRONG AS, IT MIGHT END UP INJURING. THE CUSTOMER ASKED IF THEY COULD APPLY ANYTHING PRIOR USING THE CATHETER. NO MEDICAL INTERVENTION WAS REPORTED. PER CUSTOMER VIA EMAIL ON 04JUN2021, IT WAS STATED THAT CUSTOMER WAS USING ADAPT, BY HOLLISTER 7731 AND 7737 MEDICAL ADHESIVE REMOVER AND CUSTOMER BELIEVED THE DIFFERENT PRODUCT NUMBERS WERE ONLY DUE TO THE LARGE AND SMALL SIZE SPRAY BOTTLES. IT WAS ALSO STATED THAT NOW THE PRODUCT WORKED GREAT AND WELL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain

    MAUDE report 11578806 ↗
  • MalfunctionJun 19, 2019Product Quality Problem; Product Quality Problem; Material Deformation

    IT WAS REPORTED THAT THE CATHETERS WERE TOO LUBRICIOUS MAKING THEM DIFFICULT TO UNROLL AND UNUSABLE.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 8713920 ↗
  • MalfunctionAug 9, 2018Nonstandard Device; Product Quality Problem; Nonstandard Device

    IT WAS REPORTED THAT THE MALE EXTERNAL CATHETER ADHESIVE WAS TOO STRONG.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 7768234 ↗
  • InjuryNov 6, 2017Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT ALLEGEDLY EXPERIENCED AN UTI AND WAS TREATED WITH ANTIBIOTICS.

    Patient problems: Urinary Tract Infection

    MAUDE report 7002747 ↗

Other DEVICE, INCONTINENCE, UROSHEATH TYPE, NON-STERILE devices

All DEVICE, INCONTINENCE, UROSHEATH TYPE, NON-STERILE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page