Power-Trialysis catheter 5615150Recalls & reported problems
Bard Access Systems, Inc. · 13Fr Trialysis PC 15cm Tray · Class 2
All models of Power-Trialysis catheter (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741075094), received between Mar 16, 2016 and Nov 29, 2023 — classified as 2 deaths, 6 malfunctions.
The most frequently reported specific device problem is “Failure to Infuse” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 10, 2026Class IOpen, Classified
this is a downstream recall of class I recall RES 98777
Z-2984-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Infuse2
- Break1
- Complete Blockage1
- Component Incompatible1
- Detachment of Device or Device Component1
- Fluid/Blood Leak1
- Insufficient Device Problem Information1
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Death1
- Exsanguination1
- Insufficient Information1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
DeathNov 29, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
CUSTOMER INQUIRED ON ANY CHANGES IN DESIGN OR MANUFACTURING PROCESS FOR DEVICE. ADDITIONAL INFORMATION RECEIVED 11/6/2023: IT WAS REPORTED BY THE CUSTOMER "THE TWO PORTS AND THEIR RESPECTIVE LINES LEADING TO THE "Y" HUB (ARTERIAL AND VENOUS) BECAME DETACHED FROM THE HUB. THIS WAS A 79 YEAR OLD WHITE MALE PATIENT WEIGHING 77KG WITH A PMH SIGNIFICANT FOR LEUKEMIA AND GI NEUROENDOCRINE TUMOR WHO PRESENTED WITH WEAKNESS AND ENCEPHALOPATHY. HE DEVELOPED AKI WHICH REQUIRED DIALYSIS WITH CATHETER IN QUESTION. THE PATIENT CODED REQUIRING ACLS AND THE PATIENT EXPIRED." ADDITIONAL INFORMATION RECEIVED 11/30/2023: IT WAS REPORTED BY THE CUSTOMER ¿THE PATIENT WAS NOTED TO HAVE BEEN RESTING COMFORTABLY PRIOR TO THE CODE. HE DID HAVE DIALYSIS EARLIER IN THE DAY ~ 8 HOURS PRIOR TO CODE. THE TWO PORTS AND THEIR RESPECTIVE LINES LEADING TO THE "Y" HUB (ARTERIAL AND VENOUS) BECAME DETACHED FROM THE HUB. CAUSE OF DEATH WAS EXSANGUINATION FROM DETACHMENT OF THE DIALYSIS CATHETER. CODE EFFORTS RENDERED IMMEDIATELY UPON LOSS OF PULSES. ROSC NOT ACHIEVED. DEVICE WAS NOTED TO HAVE BEEN WORKING PROPERLY, SITE WAS CLEAN, DRY, AND INTACT. IT WAS NOTED THAT A TEGADERM WAS NOT BEING USED AT THE TIME OF THE EVE
Patient problems: Exsanguination; Insufficient Information
MAUDE report 18229401 ↗MalfunctionMay 2, 2023Component Incompatible; Component Incompatible
PER MEDWATCH: GUIDEWIRE BECAME STUCK IN NEEDLE DURING SHORT TERM TRIAL CATHETER INSERTION. ADDITIONAL INFORMATION RECEIVED 02/28/2023: GUIDEWIRE WAS USED DURING AN INSERTION OF A TRIAL CATHETER FOR A PATIENT, IT BECAME STUCK IN NEEDLE DURING CATHETER INSERTION. NO PATIENT HARM REPORTED. (B)(6) 2023 ONE OF THE COILS OF THE COILED WIRE OF THE GUIDEWIRE WAS FOUND TO BE UNRAVELED NEAR THE DISTAL END OF THE GUIDEWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16856817 ↗MalfunctionJan 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT LUMEN CRACKED APART IN THE CURVE SECTION OF THE LUMEN. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16158171 ↗MalfunctionFeb 16, 2022Failure to Infuse; Failure to Infuse
IT WAS REPORTED A PATIENT HAS CLOTTED OFF TWO LINES IN THE LAST 12 HOURS SUSPECTED DUE TO THE LINE CURVING AND CAUSING A CLOT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13546049 ↗MalfunctionFeb 16, 2022Failure to Infuse; Failure to Infuse
IT WAS REPORTED A PATIENT HAS CLOTTED OFF TWO LINES IN THE LAST 12 HOURS SUSPECTED DUE TO THE LINE CURVING AND CAUSING A CLOT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13546043 ↗MalfunctionMar 25, 2020Material Frayed; Material Frayed
IT WAS REPORTED THAT THE MD WAS PLACING THE TRIALYSIS CATHETER AND WHEN THEY PULLED OUT THE WIRE IT LOOKED LIKE A SPRING AND WAS ALL CURLED UP. ADDENDUM PER RECEIVED EMAIL: "I WAS PLACING A DIALYSIS CATHETER FOR THIS PATIENT. WHILE PULLING THE WIRE, THE END OF IT BECAME LOOSE AND UNCOILED (LIKE A SPRING)."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9879300 ↗MalfunctionAug 22, 2018Complete Blockage
IT WAS REPORTED VIA MEDWATCH DESCRIPTION THAT THE "DIALYSIS CATHETER WAS BEING EXCHANGED BECAUSE IT HAD CLOTTED OFF. WHEN REMOVING THE CATHETER A BULGE WAS NOTED IN THE BLUE LUMEN LINE. IT APPEARS TO BE A WEAK POINT IN THE CATHETER LUMEN. POTENTIAL FOR MASSIVE BLEEDING IF LINE HAD RUPTURED."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7807809 ↗DeathMar 16, 2016Insufficient Device Problem Information
IT WAS REPORTED PER PHYSICIAN THAT THERE WAS A PATIENT DEATH WHILE USING THE POWER TRIALYSIS PRODUCT. THE PHYSICIAN WITNESSED THE DEATH ASSOCIATED WITH THE INSERTION OF A POWERTRIALYSIS AND FEELS THAT THE DILATORS ARE TOO LONG.
Patient problems: Death
MAUDE report 5504030 ↗
Other CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED devices
- Power-Trialysis catheter 5605150Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls13 reports
- Power-Trialysis catheter 5605240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls8 reports
- MAHURKAR 8888101002HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls1 death8 reports
- Power-Trialysis Slim-Cath catheter 5855240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- Power-Trialysis catheter 5615200Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888101003HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888102003HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888102004HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page