BrachyStar™ 918201Recalls & reported problems
Bard Brachytherapy, Inc. · BrachyStar Brachytherapy Applicator Seed Implant Needle with Lubricious Coating and cm Markings, 18g x 20cm Length · Class 1
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741075957), received between Nov 21, 2018 and Aug 25, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent4
- Material Deformation3
- Difficult to Insert1
- Incorrect Measurement1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 25, 2025Incorrect Measurement; Material Deformation; Incorrect Measurement; Material Deformation
ON 30-JUL-2025, A PATIENT WAS UNDERGOING A BRACHYTHERAPY PROCEDURE. BEFORE USE OF THE NEEDLE, IT WAS REPORTED THAT THE PINK PART OF THE BRACHY NEEDLE STYLET OR CANNULA HAD A SHAPE DEFECT. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22882008 ↗MalfunctionAug 25, 2025Material Deformation; Material Deformation
ON (B)(6) 2025, A PATIENT WAS UNDERGOING A BRACHYTHERAPY PROCEDURE. BEFORE USE OF THE NEEDLE, IT WAS REPORTED THAT THE PINK PART OF THE BRACHY NEEDLE STYLET OR CANNULA HAD A SHAPE DEFECT. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22881939 ↗MalfunctionAug 25, 2025Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent
ON (B)(6) 2025, A PATIENT WAS UNDERGOING A BRACHYTHERAPY PROCEDURE. BEFORE USE OF THE NEEDLE, IT WAS REPORTED THAT THE PINK PART OF THE BRACHY NEEDLE STYLET OR CANNULA HAD A SHAPE DEFECT. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22881933 ↗MalfunctionFeb 22, 2022Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT PRIOR TO A BRACHYTHERAPY PROCEDURE, THE NEEDLE WAS ALLEGEDLY NOTED TO BE CURVED UPON OPENING THE PACKAGE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13576928 ↗MalfunctionAug 25, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT IMMEDIATELY AFTER OPENING THE PACKAGE, THE NEEDLE WAS ALLEGEDLY FOUND TO BE BENT. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12364903 ↗MalfunctionAug 25, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT IMMEDIATELY AFTER OPENING THE PACKAGE, THE NEEDLE WAS ALLEGEDLY FOUND TO BE BENT. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12364900 ↗MalfunctionAug 6, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, THE INNER NEEDLE AND THE CANNULA WERE ALLEGEDLY NARROW AND NOT SMOOTH. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12286580 ↗MalfunctionNov 21, 2018Difficult to Insert
IT WAS REPORTED THAT THE USER FOUND IT DIFFICULT TO INSERT THE NEEDLE THUS, ANOTHER NEEDLE WAS USED.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8097127 ↗
Other System, applicator, radionuclide, manual devices
- Theragenics CPPS-MN-1820-5Theragenics Corporation · System, Applicator, Radionuclide, Manual1 recall0 reports
- QuickLink™ 70310QCA1Bard Brachytherapy, Inc. · System, applicator, radionuclide, manual16 reports
- SIROS Delivery Set SIR-10300Sirtex Medical Inc. · System, Applicator, Radionuclide, Manual9 reports
- BrachyStar™ 918205Bard Brachytherapy, Inc. · System, applicator, radionuclide, manual8 reports
- BrachyStar™, FastFill™ 818205Bard Brachytherapy, Inc. · System, applicator, radionuclide, manual8 reports
- Prostate Stabilization Set 500118200Argon Medical Devices, Inc. · System, Applicator, Radionuclide, Manual1 report
- Cradle, BK 8848 6211Best Nomos · System, Applicator, Radionuclide, Manual1 report
- TheraSphere™ M00139553CEA0Boston Scientific Corporation · System, applicator, radionuclide, manual1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.