Bard® Blakemore Esophageal-Nasogastric Tube 0092300Recalls & reported problems
C. R. Bard, Inc. · Bard® Blakemore Esophageal-Nasogastric Tube · Class 2
All models of Bard Blakemore Esophageal-Nasogastric Tube (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741076831), received between Jun 30, 2023 and Jan 23, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Apr 17, 2025Class IOpen, Classified
Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.
Z-1758-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Burst Container or Vessel1
- Inflation Problem1
- Material Rupture1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT DURING PROCEDURE, THE ITEM WAS ALLEGEDLY FOUND TO HAVE A PERFORATION IN THE TUBING OF THE BLAKEMORE TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24167144 ↗MalfunctionJan 12, 2024Burst Container or Vessel; Material Rupture; Burst Container or Vessel; Material Rupture
IT WAS REPORTED THAT THE BALLOON HAD RUPTURED WHILE ON USE ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18503756 ↗MalfunctionJun 30, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT TWO (2) OF THE BLAKEMORE SPECIALTY TUBES DID NOT INFLATE PRIOR TO BEING INSERTED IN A PATIENT IN THE INTENSIVE CARE UNIT (ICU). IT WAS NOTED THAT THE SPECIALTY TUBES WERE NOT EXPIRED. THE CUSTOMER ADDITIONALLY NOTICED THAT THERE WAS DISCOLORATION WITH THE TWO (2) FAILED TUBES. EACH TUBE WAS A DIFFERENT LOT# AND EXPIRATION DATE. AS PER FOLLOW UP INFORMATION RECEIVED VIA EMAIL ON 12-JUNE-2023, THE CLINICAL EDUCATION SPECIALIST REPORTED THAT THE PATIENT WAS DECEASED. THE PATIENT WAS BEING CODED WHEN THE ENDOSCOPY TEAM ARRIVED TO SET UP THE TUBE, AND THE CODE WAS ANNOUNCED WHILE THE ENDOSCOPY TEAM WAS SETTING UP. THE TEAM HAD THE BLAKEMORE KIT SET UP, AND THEY ATTEMPTED TO TEST THE BALLOONS BEFORE USE. THE BALLOONS IN THE KIT WOULD NOT STAY INFLATED, ANOTHER KIT OBTAINED FROM THE ICU, BUT THE BALLOONS WOULD NOT STAY INFLATED. IT WAS ALSO NOTED THAT BOTH OF THESE FAILED DEVICES WERE DISCOLORED. AN ADDITIONAL TUBE WAS OBTAINED FROM THE OPERATING ROOM (OR). THERE WAS NO DISCOLORATION NOTED AND THE BALLOONS WERE ABLE TO REMAIN INFLATED. THE CLINICAL EDUCATION SPECIALIST STATED THAT THE ISSUES ENCOUNTERED WITH THE BLAKEMORE TUBES DID NOT DELAY NOR IMPACT THE OUTCOME. PE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17237765 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page