Skip to content

Bard® Blakemore Esophageal-Nasogastric Tube 0092300Recalls & reported problems

C. R. Bard, Inc. · Bard® Blakemore Esophageal-Nasogastric Tube · Class 2

All models of Bard Blakemore Esophageal-Nasogastric Tube (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741076831), received between Jun 30, 2023 and Jan 23, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 17, 2025Class IOpen, Classified

    Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.

    Z-1758-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Inflation Problem1
  • Material Rupture1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING PROCEDURE, THE ITEM WAS ALLEGEDLY FOUND TO HAVE A PERFORATION IN THE TUBING OF THE BLAKEMORE TUBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24167144 ↗
  • MalfunctionJan 12, 2024Burst Container or Vessel; Material Rupture; Burst Container or Vessel; Material Rupture

    IT WAS REPORTED THAT THE BALLOON HAD RUPTURED WHILE ON USE ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18503756 ↗
  • MalfunctionJun 30, 2023Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT TWO (2) OF THE BLAKEMORE SPECIALTY TUBES DID NOT INFLATE PRIOR TO BEING INSERTED IN A PATIENT IN THE INTENSIVE CARE UNIT (ICU). IT WAS NOTED THAT THE SPECIALTY TUBES WERE NOT EXPIRED. THE CUSTOMER ADDITIONALLY NOTICED THAT THERE WAS DISCOLORATION WITH THE TWO (2) FAILED TUBES. EACH TUBE WAS A DIFFERENT LOT# AND EXPIRATION DATE. AS PER FOLLOW UP INFORMATION RECEIVED VIA EMAIL ON 12-JUNE-2023, THE CLINICAL EDUCATION SPECIALIST REPORTED THAT THE PATIENT WAS DECEASED. THE PATIENT WAS BEING CODED WHEN THE ENDOSCOPY TEAM ARRIVED TO SET UP THE TUBE, AND THE CODE WAS ANNOUNCED WHILE THE ENDOSCOPY TEAM WAS SETTING UP. THE TEAM HAD THE BLAKEMORE KIT SET UP, AND THEY ATTEMPTED TO TEST THE BALLOONS BEFORE USE. THE BALLOONS IN THE KIT WOULD NOT STAY INFLATED, ANOTHER KIT OBTAINED FROM THE ICU, BUT THE BALLOONS WOULD NOT STAY INFLATED. IT WAS ALSO NOTED THAT BOTH OF THESE FAILED DEVICES WERE DISCOLORED. AN ADDITIONAL TUBE WAS OBTAINED FROM THE OPERATING ROOM (OR). THERE WAS NO DISCOLORATION NOTED AND THE BALLOONS WERE ABLE TO REMAIN INFLATED. THE CLINICAL EDUCATION SPECIALIST STATED THAT THE ISSUES ENCOUNTERED WITH THE BLAKEMORE TUBES DID NOT DELAY NOR IMPACT THE OUTCOME. PE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17237765 ↗

Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices

All TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices →