Arctic Sun® Temperature Management System, Power Cord 73300Recalls & reported problems
Medivance, Inc. · Arctic Sun® Temperature Management System, Power Cord · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741080258), received between Jul 5, 2023 and Jun 3, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Power Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Power Problem3
- Circuit Failure2
- Defective Component2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 3, 2025Defective Component; Defective Component
IT WAS REPORTED THAT THE ARCTIC SUN DEVICE WAS NOT COOLING. A CHECK OF THE DIAGNOSTICS SHOWED THAT THE CHILLER TEMPERATURE (T4) WAS AT 23C. THEY STATED THAT MORE THAN 600 ML DRAINED FROM THE CHILLER TANK. THEY DISCUSSED THIS WAS INDICATIVE OF A FAILED CHILLER. PER FOLLOW UP INFORMATION REC EIVED VIA TASK ON 02APR2025, NO PATIENT INVOLVED AT THE TIME OF THE MALFUNCTION. PER SAMPLE EVALUATION RESULTS RECEIVED VIA TASK ON 08MAY2025, THE DEVICE FAILED AN ELECTRICAL SAFETY TEST STEP 2(3) MAIN CABLE CHECK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22132773 ↗MalfunctionMar 27, 2025Circuit Failure; Circuit Failure
IT WAS REPORTED THAT THE SCREEN STAYED BLACK IN AN ARCTIC SUN DEVICE. PER REVIEW OF WO ON (B)(6) 2025, THERE WAS NO PATIENT INVOLVEMENT. PER FOLLOW UP INFORMATION RECEIVED VIA TASK ON 12MAR2025, THE DEVICE WOULD BE SENT TO TASK MANAGEMENT FOR REPAIR. DEVICE HAS NOT BEEN SERVICED YET. PER SAMPLE EVALUATION RESULTS RECEIVED VIA TASK ON 25MAR2025, IT WAS REPORTED THAT THERE WAS AN ELECTRICAL SAFETY TEST FAILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21709691 ↗MalfunctionNov 30, 2023Circuit Failure; Power Problem; Circuit Failure; Power Problem
IT WAS REPORTED THAT THE WATER TEMPERATURE DROPPED OUT AND PATIENT TEMPERATURE PLUMMETED ON THE BACK OF THIS IN AN ARCTIC SUN DEVICE. THIS HAPPENED TWICE DURING THE REWARM PHASE OF THERAPY. PER SAMPLE EVALUATION RESULTS ON 13NOV2023, IT WAS REPORTED THAT THE ARCTIC SUN DEVICE FAILED THE PROTECTIVE EARTH DURING EST DUE TO HIGH RESISTANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18237367 ↗MalfunctionJul 18, 2023Defective Component; Power Problem; Defective Component; Power Problem
IT WAS REPORTED THAT IN FIRST CALL, NURSE STATED THAT THEY HAD A PATIENT ON ARCTIC SUN DEVICE AND WHEN THEY RETURNED TO THE ROOM THE DEVICE WAS POWERED OFF. IT HAD NOT BEEN UNPLUGGED OR TOUCHED. THEY HAVE SWITCHED OUTLETS, BUT IT WOULD NOT TURN ON. VERIFIED DEVICE WAS NOT PLUGGED INTO BED. NOTED IF THE DEVICE WOULD NOT POWER ON, THEN THEY WOULD NEED TO SWAP OUT FOR NEW DEVICE AND SEND THIS ONE TO BIOMED FOR EVALUATION. IN SECOND CALL, BIOMED BEDSIDE IN ICU WITH NURSE. DEVICE HAD BEEN ALERTING PATIENT LINE OPEN AND LOW FLOW. NURSE STATED THAT THE PADS HAD BEEN ON FOR LESS THAN ONE SHIFT, BUT THEY HAD ISSUES SINCE THE START OF THERAPY. BIOMED CHECKED CONNECTIONS AND CHANGED OUT THE FLUID DELIVERY LINE (FDL) BECAUSE IT APPEARED WORN. WALKED THROUGH STOP THERAPY, DRAIN AND DISCONNECT. RECONNECTED HOLDING NOWHERE NEAR CLEAR CONNECTORS. VERIFIED AUDIBLE CLICKS WITH EACH CONNECTION. ALL LINES STRAIGHT WITH NO BENDS OR KINKS AND VERIFIED THAT FLUID DELIVERY LINE (FDL) WAS SECURE IN THE LOCK POSITION. RESTARTED THERAPY AND DEVICE WOULD PRIME TO WATER FLOW RATE (WFR) WAS 0 L/M AND ALERTING LOW AIR LEAK AND PATIENT LINE OPEN. SYSTEM DIAGNOSTICS WERE OUTLET MONITOR TEMPERATURE (T1) WAS 7.3C, O
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17341908 ↗MalfunctionJul 5, 2023Power Problem; Power Problem
IT WAS REPORTED THAT THE RESPIRATORY CAME IN WHILE REWARMING PATIENT AND UNPLUGGED ARCTIC SUN DEVICE FROM THE WALL. AT SOME POINT SPARK FROM WALL OCCURRED IN THE ROOM. THEY WERE MOVING PATIENT AND DEVICE TO A NEW ROOM. WANTED TO KNOW HOW TO CONTINUE THERAPY. TARGETED TEMPERATURE (TT) WAS 37C AND PATIENT TEMPERATURE (PT) WAS 36.8C. ADVISED TO PLUG IN DEVICE IN NEW ROOM. AT BOOT UP SELECT CONTINUE CURRENT PATIENT AND START UNDER REWARM WHERE THEY LEFT OFF. AFTER SPARK OCCURRED IN OTHER ROOM THEY MOVED PATIENT TO NEW ROOM AND HAVE PLUGGED IN THE DEVICE TO RESUME THERAPY. DEVICE WOULD NOT POWER ON. ADVISED THEY WOULD NEED TO SEND THIS DEVICE TO BIOMED FOR EVALUATION. ELECTRICAL ISSUE IN OTHER ROOM POSSIBLY DAMAGED THE DEVICE. CONFIRMED THEY DID NOT HAVE ANOTHER DEVICE AVAILABLE FOR USE. WILL NEED TO CONSULT PROTOCOL ON ALTERNATE METHOD FOR THERAPY. BIOMED CALL. A NURSE PLUGGED IN DEVICE AND IT SPARKED THE NIGHT BEFORE. BIOMED HAD WORKED ON THE DEVICE AND REPLACED THE POWER CORD. THEY WOULD LIKE TO DO A FUNCTIONAL CHECK AND MAKE SURE EVERYTHING WAS WORKING APPROPRIATELY. WARM TRANSFERRED FOR PROPER HANDLING. PER FOLLOW UP INFORMATION RECEIVED VIA EMAIL ON 20JUL2023, (SEE ATTACHMENT). IT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17263198 ↗
Other System, thermal regulating devices
- Arctic Sun® 5000 50000000Medivance, Inc. · System, thermal regulating1 recall3 deaths6,903 reports
- Arctic Sun® 5000 50000000EMedivance, Inc. · System, thermal regulating1 recall2 deaths4,413 reports
- Arctic Sun Stat 60000000Medivance, Inc. · System, thermal regulating1 recall1,613 reports
- Arctic Sun® 5000 50000000LMedivance, Inc. · System, thermal regulating1 recall388 reports
- Arctic Sun® Temperature Management System, ArcticGel™ Pads 31802Medivance, Inc. · System, thermal regulating1 recall385 reports
- Altrix 8001Stryker Corporation · System, thermal regulating1 recall160 reports
- Arctic Sun Stat 60000000LMedivance, Inc. · System, thermal regulating1 recall104 reports
- HotDog Temperature Management System WC77Augustine Temperature Management LLC · System, Thermal Regulating2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.