Skip to content

Arctic Sun® Temperature Management System, Strap Kit, Fluid Delivery Line 73490Recalls & reported problems

Medivance, Inc. · Arctic Sun® Temperature Management System, Strap Kit, Fluid Delivery Line · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741080289), received between Dec 30, 2022 and Sep 5, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak1
  • Material Split, Cut or Torn1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 5, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THEY WERE TRYING TO INITIATE THERAPY USING ARCTIC SUN DEVICE BUT WERE GETTING ALERT 02 (LOW FLOW). FLOW RATE (FR) WAS 0LPM, INLET PRESSURE (IP) WAS -0.4PSI, CIRCULATION PUMP COMMAND (CPC) WAS 100 PERCENTAGE, PUMP HOURS WERE 5528, SYSTEM HOURS WERE 6456. HAD NURSE STOP THERAPY, EMPTY PADS, AND DISCONNECT PADS. PLACED DEVICE IN MANUAL MODE, WITH ONLY THE FLUID DELIVERY LINE (FDL) ATTACHED, FLOW RATE (FR) WAS 1.8LPM, INLET PRESSURE (IP) WAS -7.3PSI, CIRCULATION PUMP COMMAND (CPC) WAS 74 PERCENTAGE. CONNECTED LEFT THIGH PAD, FLOW RATE (FR) WAS 2.1LPM, INLET PRESSURE (IP) WAS -4.6PSI. MOVED LEFT THIGH PAD TO ANOTHER VALVE SET, FLOW RATE (FR) WAS 0.8LPM, INLET PRESSURE (IP) WAS -7.1PSI, CIRCULATION PUMP COMMAND (CPC) WAS 45 PERCENTAGE. CONNECTED LEFT CHEST PAD, NURSE MISTAKENLY ATTACHED IT TO THE FIRST VALVE SET LOCATION THEY CONNECTED THE LEFT THIGH PAD TO AT FIRST. 0.1LPM, -0.7PSI. MOVED TO THE OTHER VALVE SET ON THAT SIDE OF THE FLUID DELIVERY LINE (FDL), FLOW RATE (FR) WAS 1.8LPM, INLET PRESSURE (IP) WAS -7.3PSI, CIRCULATION PUMP COMMAND (CPC) WAS 68 PERCENTAGE. CONNECTED RIGHT THIGH PAD, 2.4LPM, -7.3PSI, CIRCULATION PUMP COMMAND (CPC) WAS 84 PERCENTAGE. CONNECTE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20148264 ↗
  • MalfunctionJan 30, 2023Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT WHEN ARCTIC SUN DEVICE WAS CHECKED TO BE DELIVERED, THERE WAS A BURR LEFT IN THE BUTTON OF THE ATTACHED STRAP KIT AND THE FASTENER DID NOT FIT. PER ADDITIONAL INFORMATION RECEIVED VIA FOLLOW-UP ON 25-JAN-2023, IT WAS REPORTED THAT ISSUE WOULD BE RESOLVED BY REPLACING IT WITH A NEW ONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16264310 ↗
  • MalfunctionDec 30, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE STRAP KIT FOR THE FLUID DELIVERY LINE WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16077939 ↗

Other System, thermal regulating devices

All System, thermal regulating devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page