Skip to content

Arctic Sun® Temperature Management System, Temperature Cable Phono 73505Recalls & reported problems

Medivance, Inc. · Arctic Sun® Temperature Management System, Temperature Cable Phono · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741080326), received between Jul 22, 2022 and Apr 4, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Sensing Problem1
  • Disconnection1
  • Material Fragmentation1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 4, 2023Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE TEMPERATURE WAS FLUCTUATING ON THE ARCTIC SUN DEVICE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 16673340 ↗
  • MalfunctionJul 22, 2022Disconnection; Material Fragmentation; Disconnection; Material Fragmentation

    IT WAS REPORTED THAT THE PATIENT'S TEMPERATURE SUDDENLY STOPPED BEING DISPLAYED WHILE USING ARCTIC SUN-5000 FOR COOLING. WHEN THE TEMPERATURE PROBE CONNECTION CABLE WAS SHAKEN, THE BODY TEMPERATURE WAS DISPLAYED. THEREFORE IT WAS DETERMINED THAT THE CONNECTION CABLE WAS DISCONNECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15083984 ↗