Bard Access Systems, Inc. · Ponsky PEG Safety System · Class 2 · Implant

All models of Ponsky PEG Safety System (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Bard Access Systems Ponsky PEG Safety System is a gastrostomy feeding tube placed through the abdomen. The 32 reports mostly describe the internal bumper or tube breaking, detaching, or separating, sometimes leaving a fragment inside the stomach that had to be removed by endoscope, plus leaking fluid or blood. Two injuries were reported and no deaths. No recalls name this device, and there were no reports in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00801741081187), received between Mar 10, 2017 and Feb 16, 2022 — classified as 2 injuries, 30 malfunctions.

The most frequently reported specific device problem is “Fracture” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture7
  • Break6
  • Defective Component5
  • Fluid/Blood Leak5
  • Material Separation4
  • Detachment of Device or Device Component3
  • Component Missing2
  • Detachment Of Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • No Consequences Or Impact To Patient9
  • No Known Impact Or Consequence To Patient5
  • Device Embedded In Tissue or Plaque2
  • No Information2
  • Bowel Perforation1
  • Fistula1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 16, 2022Fracture; Material Separation; Fracture; Material Separation

    IT WAS REPORTED THAT SOMETIME POST FEEDING DEVICE PLACEMENT PROCEDURE, THE INTERNAL BUMPER OF THE DEVICE ALLEGEDLY DETACHED AND THE TUBE CAME OFF WHEN THE DEVICE WAS TRIED TO BE USED, THE INTERNAL BUMPER WAS REPORTED TO HAVE BEEN EXCRETED WITH THE STOOL. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13539955 ↗
  • MalfunctionNov 1, 2021Fracture; Material Separation; Fracture; Material Separation

    IT WAS REPORTED THAT SIX MONTHS POST DEVICE IMPLANT, THE INTERNAL BUMPER WAS ALLEGEDLY BROKEN AND WAS FOUND TO BE INSIDE THE STOMACH. IT WAS FURTHER REPORTED THAT THE BUMPER WAS REMOVED USING AN ENDOSCOPY PROCEDURE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12726032 ↗
  • MalfunctionSep 20, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT POST DEVICE PLACEMENT, THE FEEDING ADAPTER OF THE TUBE ALLEGEDLY CAME OFF. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12493713 ↗
  • MalfunctionAug 31, 2021Blocked Connection; Blocked Connection

    IT WAS REPORTED THAT APPROXIMATELY ONE MONTH POST CATHETER IMPLANT, THE FEEDING ADAPTER ALLEGEDLY HAD DIFFICULTY CONNECTING WITH THE ENTERAL NUTRITION. THE DEVICE WAS REPORTED TO BE EXPLANTED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12390247 ↗
  • MalfunctionJul 30, 2021Loss of or Failure to Bond; Fracture; Material Separation; Detachment of Device or Device Component; Loss of or Failure to Bond; Fracture; Material Separation; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE APPROXIMATELY SEVEN MONTHS POST FEEDING DEVICE IMPLANT, THE INTERNAL BUMPER ALLEGEDLY DETACHED. IT WAS FURTHER REPORTED THAT THE INTERNAL BUMPER WAS REMOVED USING AN ENDOSCOPE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12253338 ↗
  • MalfunctionJun 7, 2021Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING THE DEVICE REMOVAL PROCEDURE, THE DEVICE WAS ALLEGEDLY BROKEN AND REMAINED IN THE STOMACH. IT WAS FURTHER REPORTED PART WAS REMOVED USING AN ENDOSCOPE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11945014 ↗
  • InjuryJun 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE FEEDING TUBE PLACEMENT PROCEDURE, THE TRANSVERSE COLON WAS ALLEGEDLY PERFORATED AND RESULTING IN TRANSVERSE COLON FISTULA. IT WAS FURTHER REPORTED THAT THE CATHETER WAS REMOVED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Bowel Perforation; Fistula

    MAUDE report 11918983 ↗
  • MalfunctionMar 24, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT SOME TIME POST FEEDING TUBE PLACEMENT DURING EXCHANGING THE CATHETER, THE BUMPER ALLEGEDLY CAME OFF. IT WAS FURTHER REPORTED THAT THE BUMPER LEFT IN THE PATIENT BODY WAS REMOVED USING ENDOSCOPE. THE HEALTH INJURY TO THE PATIENT WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11559287 ↗
  • MalfunctionDec 24, 2020Fracture; Malposition of Device; Fracture; Malposition of Device

    IT WAS REPORTED THAT APPROXIMATELY 34 DAYS POST FEEDING TUBE PLACEMENT, THE CATHETER WAS ALLEGEDLY FOUND PENETRATED TRANSVERSE COLON. IT WAS FURTHER REPORTED THAT DEVICE WAS USED MULTIPLE TIMES. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11070916 ↗
  • MalfunctionDec 1, 2020Peeled/Delaminated; Deformation Due to Compressive Stress; Peeled/Delaminated; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT DURING A FEEDING DEVICE PLACEMENT PROCEDURE, THE SURFACE OF THE WIRE WAS ALLEGEDLY FOUND PEELED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10926749 ↗
  • MalfunctionNov 26, 2020Component Missing; Component Missing

    IT WAS REPORTED THAT THE UPON OPENING THE PORT DEVICE PACKAGE, THE FORCEPS WAS ALLEGEDLY MISSING IN THE KIT. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10910778 ↗
  • MalfunctionNov 12, 2020Fracture; Fracture

    IT WAS REPORTED THAT POST IMPLANTED PROCEDURE APPROXIMATELY, THE DEVICE ALLEGEDLY BROKEN. IT WAS FURTHER REPORTED THAT DEVICE WAS REMOVED FROM THE PATIENT. THERE WAS NO REPORTED ON PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10827394 ↗
  • MalfunctionNov 2, 2020Illegible Information; Illegible Information

    IT WAS REPORTED THAT TWENTY-FOUR DAYS POST GASTROSTOMY FEEDING TUBE PLACEMENT, THE DEPTH MARKINGS ALLEGEDLY DISAPPEARED ON THE TUBE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10770719 ↗
  • MalfunctionOct 26, 2020Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT APPROXIMATELY ONE DAY POST FEEDING TUBE PLACEMENT, THE LINEAR EXTERNAL BOLSTER ALLEGEDLY DAMAGED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10734370 ↗
  • MalfunctionApr 8, 2020Fluid/Blood Leak; Nonstandard Device; Fitting Problem; Fluid/Blood Leak; Nonstandard Device; Fitting Problem

    IT WAS REPORTED THAT SOME TIME POST GASTROSTOMY FEEDING TUBE PLACEMENT, THE CAP OF THE FEEDING TUBE ADAPTER WAS ALLEGEDLY UNABLE TO CLOSE PROPERLY DUE TO FINS REMAINED IN THE CAP CAUSED AN ALLEGED LEAK. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9941125 ↗
  • MalfunctionFeb 14, 2020Component Missing; Component Missing

    IT WAS REPORTED THAT UPON UNPACKING THE FEEDING DEVICE PACKAGE, THE GUIDEWIRE WAS ALLEGEDLY FOUND MISSING IN THE PACKAGE PRIOR TO USE ON PATIENT. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Patient Involvement

    MAUDE report 9712144 ↗
  • MalfunctionDec 2, 2019Defective Component; Defective Component

    IT WAS REPORTED THAT THE DEPTH MARKINGS ALLEGEDLY WERE NOT PRINTED ON THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9398259 ↗
  • MalfunctionJul 17, 2019Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES TWO MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 006035, FEEDING DEVICE ALLEGEDLY EXPERIENCED A FLUID LEAK. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. THE TWO MALFUNCTIONS INVOLVED PATIENTS WITH NO REPORTED CONSEQUENCE. THE PATIENTS¿ AGE, WEIGHT, AND SEX WERE NOT REPORTED FOR EITHER MALFUNCTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8798941 ↗
  • MalfunctionApr 19, 2019Crack; Defective Component; Crack; Defective Component

    THIS REPORT SUMMARIZES ONE(1) MALFUNCTION EVENT. A REVIEW OF THE EVENTS INDICATED THAT MODEL 006035 FEEDING DEVICE ALLEGEDLY EXPERIENCED A CRACK. THIS REPORT WAS RECEIVED FROM ONE SOURCE. THIS EVENT INVOLVED ONE MALE PATIENT, AGE AND WEIGHT WERE NOT PROVIDED. PATIENT HAD NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8533417 ↗
  • MalfunctionDec 27, 2018Defective Component; Defective Component

    IT WAS REPORTED THAT AFTER THE PERCUTANEOUS ENDOSCOPIC GASTROSTOMY TUBE WAS PLACED, THE TUBE WAS FOUND DAMAGED. IT WAS FURTHER REPORTED THAT THE TUBE WAS REPAIRED AND REMAINS IN THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8200595 ↗
  • MalfunctionDec 27, 2018Break; Defective Component; Break; Defective Component

    IT WAS REPORTED THAT AFTER THE PERCUTANEOUS ENDOSCOPIC GASTROSTOMY TUBE WAS PLACED, THE TUBE WAS FOUND DAMAGED. IT WAS FURTHER REPORTED THAT THE TUBE WAS REPAIRED AND REMAINS IN THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8199991 ↗
  • MalfunctionOct 23, 2018Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT NUTRITION LEAKAGE OCCURRED FROM THE MIDDLE OF THE CATHETER AFTER THE CONSTRUCTED GASTROSTOMY. THE DAMAGED PART OF CATHETER WAS CUT OFF. THE REMAINING PIECE OF THE CATHETER IS CURRENTLY IN USE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7997492 ↗
  • MalfunctionMay 9, 2018Defective Component; Defective Component

    IT WAS REPORTED THAT THE DEPTH MARKS ON THE CATHETER SHAFT HAD ALLEGEDLY DISAPPEARED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7500678 ↗
  • MalfunctionMar 26, 2018Detachment Of Device Component

    IT WAS REPORTED THAT THE INTERNAL BUMPER THAT IS ATTACHED TO THE CATHETER MOVED FROM THE PATIENT'S STOMACH TO THE OUTSIDE OF THE STOMACH WALL. OBSERVING THE PATIENT'S SKIN SWELLING, THE CATHETER AND INTERNAL BUMPER WERE REMOVED FROM THE PATIENT. THE INTERNAL BUMPER DID NOT MIGRATE INSIDE THE STOMACH. NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7370191 ↗
  • MalfunctionMar 20, 2018Fluid/Blood Leak; Hole In Material

    IT WAS REPORTED THAT A HOLE IN THE CATHETER WAS FOUND ON THE OUTSIDE OF THE PATIENT'S BODY AND LEAKED TPN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7354690 ↗

Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices

All TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices →