Lutonix® Drug Coated Balloon LX35130660Recalls & reported problems

Lutonix, Inc. · LUTONIX® 035 Drug Coated Balloon PTA Catheter .035 130cm 6x60mm · Class 3

All models of LUTONIX Drug Coated Balloon (107)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Lutonix drug-coated balloon catheter is used to open narrowed arteries, typically in the legs. The 23 reports mostly describe the treated artery becoming blocked again months after the procedure, sometimes with leg pain while walking and need for another procedure, with 21 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
21
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00801741088605), received between Aug 26, 2015 and May 6, 2019 — classified as 21 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem21
  • Material Rupture2

Patient problems

  • Reocclusion20
  • Claudication3
  • No Consequences Or Impact To Patient2
  • Intimal Dissection1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 6, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN RIGHT PROXIMAL POPLITEAL ARTERY. APPROXIMATELY 10 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT PROXIMAL POPLITEAL ARTERY WAS REPORTEDLY REOCCLUDED VIA DUPLEX ULTRASOUND (DUS) IMAGING. APPROXIMATELY 28 MONTHS AFTER THE INDEX PROCEDURE, A REVASCULARIZATION WAS PERFORMED AND THE HEALTH CARE PROFESSIONAL (HCP) DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR TO THE PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 8584350 ↗
  • InjuryFeb 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, ONE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT MID SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 12 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT MID SFA VESSEL WAS REPORTEDLY REOCCLUDED. A CLINICALLY DRIVEN REINTERVENTION WAS PERFORMED APPROXIMATELY 27 MONTHS AFTER THE INDEX PROCEDURE DUE TO ASSOCIATED RIGHT LEG CLAUDICATION. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Claudication; Reocclusion

    MAUDE report 8326001 ↗
  • InjuryDec 26, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 6 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT ALLEGEDLY EXPERIENCED LIFESTYLE LIMITING CLAUDICATION AT THE TARGET LIMB DUE TO THE REOCCLUSION OF THE TARGET LESION AND VESSEL. REPORTEDLY, THE PATIENT HAD A REINTERVENTION INVOLVING THE TARGET LESION AND ADDITIONAL VESSELS ON (B)(6) 2016, BUT IT IS UNKNOWN WHAT TREATMENT WAS PERFORMED. A TARGET LESION REINTERVENTION WAS PERFORMED ON (B)(6) 2017, REPORTEDLY AFTER THE PATIENT EXPERIENCED LIFESTYLE LIMITING CLAUDICATION. THE INVESTIGATOR ASSESSED THAT THE LIFESTYLE LIMITING CLAUDICATION WAS POSSIBLY RELATED TO THE STUDY DEVICE USED DURING THE INDEX PROCEDURE AS WELL AS THE INDEX PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Claudication; Reocclusion

    MAUDE report 8196011 ↗
  • InjuryJun 1, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT DISTAL SUPERFICIAL FEMORAL ARTERY (SFA) AND POPLITEAL ARTERY. APPROXIMATELY 24 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S VESSELS CONTAINING THE TARGET LESION WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 7560307 ↗
  • InjuryApr 2, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT MID SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 15 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VASCULATURE WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2018-00074.

    Patient problems: Reocclusion

    MAUDE report 7389419 ↗
  • InjuryJan 2, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 12 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 7158118 ↗
  • InjuryNov 1, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE ENTIRE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 14 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER ARE 3006513822-2017-00245 AND 3006513822-2017-00246.

    Patient problems: Reocclusion

    MAUDE report 6993076 ↗
  • InjuryOct 27, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 7 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT SFA VASCULATURE WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2017-00227.

    Patient problems: Reocclusion

    MAUDE report 6984117 ↗
  • InjuryOct 26, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT POPLITEAL ARTERY. APPROXIMATELY 5 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT POPLITEAL VASCULATURE WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6980993 ↗
  • InjuryOct 24, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT PROXIMAL AND DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 13 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT PROXIMAL AND DISTAL SFA VESSEL WAS REPORTEDLY REOCCLUDED. AT THE TIME, NO INTERVENTION WAS PERFORMED. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER ARE 3006513822-2017-00202 AND 3006513822-2017-00203.

    Patient problems: Reocclusion

    MAUDE report 6973998 ↗
  • InjuryOct 24, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT PROXIMAL AND DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 13 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT PROXIMAL AND DISTAL SFA VESSEL WAS REPORTEDLY REOCCLUDED. AT THE TIME, NO INTERVENTION WAS PERFORMED. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER ARE 3006513822-2017-00202 AND 3006513822-2017-00204.

    Patient problems: Reocclusion

    MAUDE report 6973939 ↗
  • InjuryOct 11, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 9 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT EXPERIENCED SYMPTOMS OF CLAUDICATION AND A REVASCULARIZATION PROCEDURE WAS PERFORMED REVEALING THE RIGHT PROXIMAL SFA VASCULATURE WAS REPORTEDLY REOCCLUDED. THE HCP DEEMED THE REINTERVENTION TO BE SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE SYMPTOMS OF CLAUDICATION WERE POSSIBLY RELATED TO THE STUDY DEVICE, AND DEFINITELY RELATED TO THE PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Claudication; Reocclusion

    MAUDE report 6940219 ↗
  • InjuryAug 10, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT MID SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 8 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT MID SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBER IS 3006513822-2017-00156.

    Patient problems: Reocclusion

    MAUDE report 6786351 ↗
  • InjuryApr 27, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT POPLITEAL ARTERY. APPROXIMATELY 12 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT POPLITEAL VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6525821 ↗
  • InjuryMar 31, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 8 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS DEFINITELY RELATED TO THE STUDY DEVICE AND POSSIBLY RELATED TO THE PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6449694 ↗
  • InjuryMar 22, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 12 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT PROXIMAL SFA VESSEL WAS REPORTEDLY REOCCLUDED. DURING THE ORIGINAL INDEX PROCEDURE, THE PHYSICIAN DID NOT PREDILATE THE TARGET LESION PER CLINICAL PROTOCOL. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS POSSIBLY RELATED TO THE STUDY DEVICE AND DEFINITELY RELATED TO THE PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6425679 ↗
  • InjuryFeb 8, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 16 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6311488 ↗
  • InjuryFeb 8, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 13 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6311330 ↗
  • InjuryAug 24, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA) AND POPLITEAL ARTERY. APPROXIMATELY 18 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT'S LEFT SFA AND POPLITEAL VESSELS WERE REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED INVOLVING NORMAL PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) AND STENT PLACEMENT, WHICH THE HCP DEEMED SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBER IS 3006513822-2016-00172.

    Patient problems: Reocclusion

    MAUDE report 5901708 ↗
  • InjuryAug 16, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A GRADE E FLOW LIMITING DISSECTION WAS ALLEGEDLY PRESENT AFTER TREATMENT WITH A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER. THE HEALTH CARE PROFESSIONAL (HCP) GAINED ACCESS TO TREAT THE MILDLY CALCIFIED POPLITEAL ARTERY. THE HCP PERFORMED A SUCCESSFUL ATHERECTOMY AND PREDILATION OF THE ARTERY. THE HCP PREPARED THE LUTONIX DCB IN THE NORMAL FASHION AND TREATED THE TARGET LESION, WHICH RESULTED IN A GRADE E FLOW LIMITING DISSECTION. THE HCP USED A SUPRA STENT TO TREAT THE GRADE E FLOW LIMITING DISSECTION. THE INVESTIGATOR ASSESSED THE EVENT WAS POSSIBLY RELATED TO THE STUDY DEVICE AND DEFINITELY RELATED TO THE PROCEDURE. THE LUTONIX DCB'S WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR RETURN. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Intimal Dissection

    MAUDE report 5880467 ↗
  • InjuryApr 27, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA) AND POPLITEAL ARTERY. APPROXIMATELY 18 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT'S RIGHT SFA AND POPLITEAL ARTERY WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED INVOLVING AN ADDITIONAL DRUG COATED BALLOON AND STENT PLACEMENT, WHICH THE HCP DEEMED WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBERS ARE 3006513822-2016-00104 AND 3006513822-2016-00105. ADDITIONAL INFORMATION WAS RECEIVED ON AUGUST 16TH, 2016, STATING THE (B)(6) ADJUDICATED THE RELATIONSHIP TO THE DEVICE AS "POSSIBLE" AND PROCEDURE RELATEDNESS AS "NOT RELATED."

    Patient problems: Reocclusion

    MAUDE report 5612762 ↗
  • MalfunctionFeb 17, 2016Material Rupture; Material Rupture

    IT WAS REPORTED A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER ALLEGEDLY RUPTURED DURING TREATMENT OF A CALCIFIED, NON-TORTUOUS RIGHT POPLITEAL ARTERY. THE HEALTH CARE PROFESSIONAL (HCP) REPORTEDLY ADVANCED THE LUTONIX DCB TO THE TARGET LESION WITHIN THE POPLITEAL ARTERY AND INCREMENTALLY INFLATED THE BALLOON TO 7 ATMOSPHERES, AT WHICH POINT THE BALLOON ALLEGEDLY RUPTURED. THE LUTONIX DCB WAS REMOVED COMPLETELY WITHOUT ISSUES AND ANOTHER LUTONIX DCB WAS USED TO COMPLETE THE PROCEDURE. THE LUTONIX DCB WAS RETURNED FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5442839 ↗
  • MalfunctionAug 26, 2015Material Rupture

    THE DESCRIPTION OF THE TRACKING PATHWAY AND LESION WITH REGARDS TO CALCIFICATION AND TORTUOSITY ARE UNKNOWN AT THIS TIME.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5032583 ↗