Lutonix® Drug Coated Balloon LX35130680Recalls & reported problems
Lutonix, Inc. · LUTONIX® 035 Drug Coated Balloon PTA Catheter .035 130cm 6x80mm · Class 3
All models of LUTONIX Drug Coated Balloon (107)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00801741088643), received between Nov 19, 2015 and Oct 17, 2018 — classified as 12 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem12
- Material Rupture2
Patient problems
- Reocclusion11
- No Consequences Or Impact To Patient2
- Intimal Dissection1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryOct 17, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE IN-STENT TARGET LESION LOCATED IN THE RIGHT DISTAL SUPERFICIAL FEMORAL ARTERY (SFA) AND PROXIMAL POPLITEAL ARTERY. APPROXIMATELY 15 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT DISTAL SFA AND PROXIMAL POPLITEAL VASCULATURE BECAME REPORTEDLY OCCLUDED THROUGH A DUPLEX ULTRASOUND EVALUATION ON (B)(6) 2017. A REVASCULARIZATION WAS PERFORMED ON (B)(6) 2017, INVOLVING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) AND AN ADDITIONAL LUTONIX DCB. THE HEALTH CARE PROFESSIONAL (HCP) DEEMED, IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2018-00227.
Patient problems: Reocclusion
MAUDE report 7974470 ↗InjuryJul 25, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA) AND POPLITEAL ARTERY. APPROXIMATELY 12 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA AND POPLITEAL VESSEL WERE REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2018-00168.
Patient problems: Reocclusion
MAUDE report 7719881 ↗InjuryMay 23, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 4 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Reocclusion
MAUDE report 7537362 ↗InjuryApr 19, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT MID SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 26 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Reocclusion
MAUDE report 7444953 ↗InjuryOct 24, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT PROXIMAL AND DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 13 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT PROXIMAL AND DISTAL SFA VESSEL WAS REPORTEDLY REOCCLUDED. AT THE TIME, NO INTERVENTION WAS PERFORMED. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER ARE 3006513822-2017-00203 AND 3006513822-2017-00204.
Patient problems: Reocclusion
MAUDE report 6973886 ↗InjuryApr 5, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 18 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT PROXIMAL SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Reocclusion
MAUDE report 6461748 ↗MalfunctionDec 9, 2016Material Rupture; Material Rupture
IT WAS REPORTED A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER ALLEGEDLY RUPTURED AT THREE ATMOSPHERES WHILE TREATING A MINIMALLY CALCIFIED INTRA-STENT STENOTIC LESION. THE HEALTH CARE PROFESSIONAL (HCP) PREDILATED THE TARGET LESION PRIOR TO TREATMENT WITH THE LUTONIX DCB. THE LUTONIX DCB WAS REMOVED WITHOUT DIFFICULTY. THE LUTONIX DCB WAS RETURNED FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6158517 ↗InjuryNov 10, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 14 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT'S LEFT DISTAL SFA STENTED AREA WAS REPORTEDLY REOCCLUDED. IT WAS ALSO REPORTED THE LIFESTENT WAS DISLOCATED AND TORQUED WHICH PROVIDES EVIDENCE THAT IT IS BREAKING. A REVASCULARIZATION WAS PERFORMED WHICH WAS UNSUCCESSFUL, BUT BASED ON THE GOOD FORMATION OF COLLATERAL VESSELS DURING THE FIRST INTENSIVE WALK TRAINING, NO ADDITIONAL MEDICATION OR INTERVENTION WAS PLANNED. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE BUT DEFINITELY RELATED TO THE PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBER IS 3006513822-2016-00202.
Patient problems: Reocclusion
MAUDE report 6092325 ↗InjuryMay 27, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 3 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT PROXIMAL SFA STENTED AREA WAS REPORTEDLY REOCCLUDED. A PERCUTANEOUS PROCEDURE WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NOTE, THE LUTONIX 035 DRUG COATED BALLOON PTA CATHETER IS INDICATED FOR PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA), AFTER PRE-DILATATION OF DE NOVO OR RESTENOTIC LESIONS UP TO 150 MM IN LENGTH IN THE NATIVE SUPERFICIAL FEMORAL OR POPLITEAL ARTERIES WITH REFERENCE VESSEL DIAMETERS OF 4-7 MM. THESE LUTONIX DCB'S WERE USED TO TREAT THE STENTED LESION PER THE STUDY REQUIREMENTS. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBERS ARE 3006513822-2016-00122 AND 3006513822-2016-00123.
Patient problems: Reocclusion
MAUDE report 5685247 ↗InjuryMay 27, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 3 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT'S RIGHT PROXIMAL SFA STENTED AREA WAS REPORTEDLY REOCCLUDED. A PERCUTANEOUS PROCEDURE WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NOTE, THE LUTONIX 035 DRUG COATED BALLOON PTA CATHETER IS INDICATED FOR PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA), AFTER PRE-DILATATION OF DE NOVO OR RESTENOTIC LESIONS UP TO 150 MM IN LENGTH IN THE NATIVE SUPERFICIAL FEMORAL OR POPLITEAL ARTERIES WITH REFERENCE VESSEL DIAMETERS OF 4-7 MM. THESE LUTONIX DCB'S WERE USED TO TREAT THE STENTED LESION PER THE STUDY REQUIREMENTS. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBERS ARE 3006513822-2016-00123 AND 3006513822-2016-00124.
Patient problems: Reocclusion
MAUDE report 5685193 ↗InjuryApr 7, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 15 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT'S LEFT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Reocclusion
MAUDE report 5555873 ↗InjuryMar 17, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 11 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT'S RIGHT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A THORACIC EPIDURAL ANALGESIA (TEA) WITH BYPASS SURGERY WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE, DRUG OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND ARE ASSOCIATED MANUFACTURERS REPORT NUMBERS 3006513822-2016-00034.
Patient problems: Reocclusion
MAUDE report 5507703 ↗InjuryMar 16, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A GRADE D DISSECTION WAS ALLEGEDLY PRESENT AFTER TREATMENT WITH A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER. THE HEALTH CARE PROFESSIONAL (HCP) GAINED ACCESS TO TREAT THE MILDLY CALCIFIED DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). THE HCP PREDILATED THE TARGET LESION WITH A BOSTON SCIENTIFIC STERLING PTA BALLOON. THE HCP THEN TREATED THE TARGET LESION WITH THE LUTONIX DCB FOR 185 SECONDS AT NINE ATMOSPHERES OF PRESSURE, WHICH RESULTED IN A GRADE D DISSECTION OF THE LESION. THE HCP USED AN INTACT VASCULAR TACK ENDOVASCULAR SYSTEM TO SUCCESSFULLY TREATED THE GRADE D DISSECTION. THE LUTONIX DCB WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR RETURN. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Intimal Dissection
MAUDE report 5504307 ↗MalfunctionNov 19, 2015Material Rupture
IT WAS REPORTED A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER ALLEGEDLY FAILED TO INFLATE AT THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA) TREATMENT SITE. THE HEALTH CARE PROFESSIONAL (HCP) GAINED PATIENT ACCESS ON THE LEFT SIDE AND SUCCESSFULLY PREDILATED THE TARGET LESION WITH A NORMAL PTA BALLOON. REPORTEDLY, THE LUTONIX DCB WAS ADVANCED TO THE SFA; HOWEVER, AS THE BALLOON WAS INCREMENTALLY INFLATED, THE BALLOON ALLEGEDLY WOULD NOT STAY INFLATED. THE DEVICE WAS SUCCESSFULLY REMOVED. ANOTHER LUTONIX DCB OF THE SAME SIZE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5236757 ↗
Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices
- LUTONIX® Drug Coated Balloon LX351305150Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter117 reports
- LUTONIX® Drug Coated Balloon LX351306150Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter75 reports
- Ranger™ H74939419502010Boston Scientific Corporation · Drug-Eluting Peripheral Transluminal Angioplasty Catheter71 reports
- Ranger™ H74939419602010Boston Scientific Corporation · Drug-Eluting Peripheral Transluminal Angioplasty Catheter71 reports
- Lutonix® Drug Coated Balloon LX351305100Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter66 reports
- IN.PACT™ Admiral™ ADM06004013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall12 reports
- IN.PACT™ Admiral™ ADM06015013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall11 reports
- IN.PACT™ Admiral™ ADM04015013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page