Lutonix® Drug Coated Balloon LX351304100Recalls & reported problems

Lutonix, Inc. · LUTONIX® 035 Drug Coated Balloon PTA Catheter .035 130cm 4x100mm · Class 3

All models of LUTONIX Drug Coated Balloon (107)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Lutonix drug-coated balloon catheter is used to open narrowed arteries in the leg, including the thigh and behind the knee arteries. The 36 reports mostly describe the treated artery narrowing or closing again months after the procedure, often requiring a later intervention to reopen it; a few mention tears in the artery wall and one mentions the balloon not deflating. There are 35 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
0
Injuries reported
35
Recalls naming it
0

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 00801741088667), received between Dec 7, 2015 and May 16, 2019 — classified as 35 injuries, 1 malfunction.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem35
  • Deflation Problem1

Patient problems

  • Reocclusion32
  • Intimal Dissection3
  • Claudication2
  • Necrosis1
  • No Consequences Or Impact To Patient1
  • Pain1
  • Thrombosis1

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 16, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESIONS LOCATED IN THE LEFT MID SUPERFICIAL FEMORAL ARTERY (SFA) AND LEFT MID POPLITEAL ARTERY. APPROXIMATELY 13 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S TARGET LESIONS WERE REPORTEDLY REOCCLUDED VIA DUPLEX ULTRASOUND (DUS) IMAGING. THE REGISTRY'S INDEPENDENT EVALUATOR ASSESSED THAT BOTH TARGET LESIONS WERE PATENT AT THIS TIME. APPROXIMATELY 29 MONTHS AFTER THE INDEX PROCEDURE, A REVASCULARIZATION WAS PERFORMED ON TARGET LESION 1, LEFT MID POPLITEAL ARTERY. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICES OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2018-00266.

    Patient problems: Reocclusion

    MAUDE report 8618005 ↗
  • InjuryApr 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT POPLITEAL ARTERY. APPROXIMATELY 6 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT POPLITEAL VESSEL WAS REPORTEDLY REOCCLUDED. A CLINICALLY DRIVEN REVASCULARIZATION WAS PERFORMED AT THIS TIME. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 8555579 ↗
  • InjuryNov 5, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESIONS LOCATED IN THE RIGHT DISTAL SUPERFICIAL FEMORAL ARTERY (SFA) AND RIGHT POPLITEAL ARTERY. APPROXIMATELY 18 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT SFA VASCULATURE AND POPLITEAL VASCULATURE WERE REPORTEDLY REOCCLUDED VIA DUPLEX ULTRASOUND (DUS) IMAGING. A REVASCULARIZATION WAS PERFORMED AND THE HEALTH CARE PROFESSIONAL (HCP) DEEMED IT SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICES OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2018-00250.

    Patient problems: Reocclusion

    MAUDE report 8039245 ↗
  • InjurySep 17, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TWO SEPARATE TARGET LESIONS. TARGET LESION 1, LOCATED IN THE RIGHT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA), WAS TREATED WITH ONE LUTONIX DCB. TARGET LESION 2, LOCATED IN THE RIGHT DISTAL SFA AND RIGHT POPLITEAL ARTERY, WAS TREATED WITH TWO LUTONIX DCBS. APPROXIMATELY 32 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT SFA AND RIGHT POPLITEAL ARTERIES WERE REPORTEDLY REOCCLUDED VIA DUPLEX ULTRASOUND (DUS) IMAGING. A REVASCULARIZATION WAS PERFORMED ON TARGET LESION 1 AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICES OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 7880820 ↗
  • InjurySep 11, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT MID SUPERFICIAL FEMORAL ARTERY (SFA). DURING THE INDEX PROCEDURE, THE PATIENT'S LEFT SFA WAS OBSERVED TO HAVE A GRADE B, NON-FLOW LIMITING, DISSECTION AFTER PRE-DILATION. THE HEATH CARE PROFESSIONAL (HCP) PROCEEDED TO TREAT THE DISEASED VESSEL WITH THE TWO LUTONIX DCBS. IT WAS OBSERVED THAT THE PATIENT HAD A GRADE C, FLOW-LIMITING, DISSECTION AFTER TREATMENT WITH THE LUTONIX DEVICES. THE HCP CHOSE TO USE THE PRE-DILATION DEVICE, BARD RIVAL BALLOON, TO DO A POST-DILATION TO RESOLVE THE DISSECTION. THE HCP DEEMED THE PROCEDURE TO BE SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICES OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2018-00191.

    Patient problems: Intimal Dissection

    MAUDE report 7864140 ↗
  • InjuryAug 29, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT POPLITEAL ARTERY. APPROXIMATELY 9 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT POPLITEAL VASCULATURE WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION INVOLVING STENT PLACEMENT WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS LIKELY NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 7826582 ↗
  • InjuryMay 30, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT POPLITEAL ARTERY. APPROXIMATELY 24 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT POPLITEAL ARTERY WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 7554007 ↗
  • InjuryMar 22, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESIONS LOCATED IN THE ENTIRE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA) AND ENTIRE RIGHT POPLITEAL ARTERY. APPROXIMATELY 13 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT POPLITEAL VESSEL WAS REPORTEDLY REOCCLUDED, BUT NO ADDITIONAL INTERVENTION WAS PERFORMED. APPROXIMATELY 26 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE REOCCLUSION EVENTS WERE NOT RELATED TO THE STUDY DEVICES OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBERS ARE 3006513822-2018-00058 AND 3006513822-2018-00059.

    Patient problems: Reocclusion

    MAUDE report 7361435 ↗
  • InjuryFeb 14, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESIONS LOCATED IN THE RIGHT POPLITEAL ARTERY. APPROXIMATELY 9 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT POPLITEAL ARTERY WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED, BY AN OUTSIDE PROVIDER OF THE CLINICAL STUDY. THE OUTSIDE PHYSICIAN USED AN ADMIRAL DEB FOLLOWED BY PTA ABOUT A WEEK LATER. APPROXIMATELY 19 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT HAD AN ABOVE-THE-KNEE AMPUTATION, WHICH REMOVED THE TARGET LESIONS, DUE TO A NON-HEALING WOUND OF A BELOW-THE-KNEE AMPUTATION. THE PRINCIPAL INVESTIGATOR OF THE CLINICAL STUDY ASSESSED THAT NEITHER OF THE AMPUTATIONS WERE RELATED TO THE STUDY DEVICE OR TO THE PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 7269700 ↗
  • InjuryJan 30, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT MID AND DISTALSUPERFICIAL FEMORAL ARTERY (SFA) AND RIGHT POPLITEAL ARTERY. APPROXIMATELY EIGHT DAYS AFTER THE INDEX PROCEDURE, THE PATIENT REPORTEDLY EXPERIENCED SYMPTOMS OF NUMBNESS AND PAIN, AND THE PATIENT WAS HOSPITALIZED, IN WHICH IT WAS FOUND THAT THE PATIENT'S RIGHT DISTAL SFA AND POPLITEAL VESSELS HAD RE-OCCLUDED. A THROMBOLYSIS FOLLOWED BY A BALLOON ANGIOPLASTY A WEEK LATER WERE PERFORMED AND THE HCP DEEMED THEY WERE SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2018-00017.

    Patient problems: Pain; Reocclusion; Thrombosis

    MAUDE report 7230008 ↗
  • InjuryJan 30, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). AFTER THE PRE-DILATION DEVICE WAS USED, A NON FLOW-LIMITING GRADE B DISSECTION WAS NOTICED. UPON EVALUATION OF THE DISSECTION AFTER THE TREATMENT WITH LUTONIX DCB BY THE CLINICAL REGISTRY'S INDEPENDENT INVESTIGATOR, IT WAS DEEMED THE DISSECTION CHANGED TO A FLOW-LIMITING GRADE D DISSECTION. THE HCP PERFORMED A POST-DILATION TREATMENT. THE CLINICAL REGISTRY'S INDEPENDENT INVESTIGATOR EVALUATED THE DISSECTION TO BE A NON FLOW-LIMITING GRADE C DISSECTION AFTER THE POST-DILATION THE INVESTIGATOR ASSESSED THAT THE EVENT WAS LIKELY NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR FURTHER EVALUATION. THE PATIENT DEVELOPED REOCCLUSION IN 7 MONTHS AFTER THE INDEX PROCEDURE, IN THE RIGHT SFA.

    Patient problems: Intimal Dissection

    MAUDE report 7229998 ↗
  • InjuryJan 23, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 7 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT SFA VESSEL WAS REPORTEDLY REOCCLUDED. THE PATIENT WAS HOSPITALIZED, BUT NO ADDITIONAL INTERVENTION HAS BEEN PERFORMED AT THIS TIME. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS POSSIBLY RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 7213442 ↗
  • InjuryDec 15, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT MID SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 6 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT'S VESSEL CONTAINING THE TARGET LESION WAS REPORTEDLY REOCCLUDED. A REINTERVENTION WAS PERFORMED INVOLVING TWO STENTS PLACED IN THE LEFT MID SFA. THE HCP DEEMED THE TREATMENT TO BE SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS UNLIKELY RELATED TO THE STUDY DEVICES AND PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2017-00304.

    Patient problems: Reocclusion

    MAUDE report 7121218 ↗
  • InjuryNov 1, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 1 MONTH AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VASCULATURE WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2017-00265.

    Patient problems: Reocclusion

    MAUDE report 6995139 ↗
  • InjuryOct 27, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 18 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VASCULATURE WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2017-00223.

    Patient problems: Reocclusion

    MAUDE report 6983444 ↗
  • InjuryOct 27, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 18 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT SFA VASCULATURE WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURER¿S REPORT NUMBER IS 3006513822-2017-00224.

    Patient problems: Reocclusion

    MAUDE report 6983365 ↗
  • InjuryOct 26, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE RIGHT POPLITEAL ARTERY. APPROXIMATELY 6 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT POPLITEAL VASCULATURE WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6981322 ↗
  • InjuryOct 6, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESIONS LOCATED IN THE RIGHT DISTAL SUPERFICIAL FEMORAL ARTERY (SFA), AND ALL BRANCHES OF THE POPLITEAL ARTERY. APPROXIMATELY 12 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S RIGHT DISTAL SFA AND BRANCHES OF THE POPLITEAL ARTERIES WERE REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICES OR PROCEDURE. THE SAMPLES WERE DISCARDED BY THE USER FACILITY AND ARE NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. NEW INFORMATION: IT WAS REPORTED THAT APPROXIMATELY 30 MONTHS POST INDEX PROCEDURE, REOCCLUSION WAS IDENTIFIED VIA IMAGING.

    Patient problems: Reocclusion

    MAUDE report 6921916 ↗
  • InjurySep 26, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT POPLITEAL ARTERY. APPROXIMATELY 1 MONTH AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT POPLITEAL VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. APPROXIMATELY 10 MONTHS AFTER THE REINTERVENTION ON (B)(6) 2016, A SECOND REINTERVENTION, ON (B)(6) 2017 WAS PERFORMED INVOLVING THE TARGET LESION SEGMENT. APPROXIMATELY 1 MONTH LATER ON (B)(6) 2017, THE PATIENT'S LEFT LEG WAS AMPUTATED ABOVE THE KNEE DUE TO CRITICAL LIMB ISCHEMIA. THE INVESTIGATOR ASSESSED THE REOCCLUSION/AMPUTATION WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO FURTHER ADVERSE PATIENT OUTCOMES WERE REPORTED AND THE SUBJECT WITHDREW FROM THE CLINICAL STUDY.

    Patient problems: Necrosis; Reocclusion

    MAUDE report 6895511 ↗
  • InjuryAug 29, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT POPLITEAL ARTERY. APPROXIMATELY 19 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT POPLITEAL VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6826286 ↗
  • InjuryJul 6, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT POPLITEAL ARTERY. APPROXIMATELY 11 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT'S LEFT POPLITEAL VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6691395 ↗
  • InjuryJun 26, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, TWO LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT MID SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 6 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT WAS EXPERIENCING CLAUDICATION IN THE TARGET LIMB. AFTER A DUPLEX ULTRASOUND, IT WAS EVIDENT THAT THE SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED INVOLVING THE PLACEMENT OF TWO STENTS AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS POSSIBLY RELATED TO THE STUDY DEVICE AND NOT RELATED TO THE PROCEDURE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBER IS 3006513822-2017-00128.

    Patient problems: Claudication; Reocclusion

    MAUDE report 6666300 ↗
  • InjuryJun 2, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT PROXIMAL-MID SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 21 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT PROXIMAL-MID SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. AN INDEPENDENT EVALUATOR HAS ADJUDICATED THE EVENT AS POSSIBLY RELATED TO THE DEVICE WITH THE RATIONALE THAT THE DRUG EFFECT CANNOT BE EXCLUDED. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBER ARE 3006513822-2017-00112 AND 3006513822-2017-00113.

    Patient problems: Reocclusion

    MAUDE report 6608936 ↗
  • InjuryJun 2, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, THREE LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETERS WERE USED TO TREAT THE TARGET LESION LOCATED IN THE LEFT PROXIMAL-MID SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 21 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S LEFT PROXIMAL-MID SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE. AN INDEPENDENT EVALUATOR HAS ADJUDICATED THE EVENT AS POSSIBLY RELATED TO THE DEVICE WITH THE RATIONALE THAT THE DRUG EFFECT CANNOT BE EXCLUDED. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND IS NOT AVAILABLE FOR EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS IS ONE OF THREE PRODUCTS INVOLVED WITH THE REPORTED EVENT AND THE ASSOCIATED MANUFACTURERS REPORT NUMBER ARE 3006513822-2017-00112 AND 3006513822-2017-00114.

    Patient problems: Reocclusion

    MAUDE report 6608802 ↗
  • InjuryJun 1, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH A CLINICAL REGISTRY THAT DURING THE INDEX PROCEDURE, A LUTONIX PERCUTANEOUS TRANSLUMINAL DRUG COATED BALLOON (DCB) DILATATION CATHETER WAS USED TO TREAT THE TARGET LESION LOCATED IN THE MID LEFT SUPERFICIAL FEMORAL ARTERY (SFA). APPROXIMATELY 14 MONTHS AFTER THE INDEX PROCEDURE, THE PATIENT¿S MID LEFT SFA VESSEL WAS REPORTEDLY REOCCLUDED. A REVASCULARIZATION WAS PERFORMED AND THE HCP DEEMED IT WAS SUCCESSFUL. THE INVESTIGATOR ASSESSED THAT THE EVENT WAS NOT RELATED TO THE STUDY DEVICE OR PROCEDURE, BUT THE INDEPENDENT REVIEWER ASSESSED THE EVENT AS RELATED TO THE DEVICE. THE SAMPLE WAS DISCARDED BY THE USER FACILITY AND NOT AVAILABLE FOR FURTHER EVALUATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Reocclusion

    MAUDE report 6604194 ↗

Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices

All Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices →