Bardia® Foley Catheter Silicone Coated 123622ARecalls & reported problems

C. R. Bard, Inc. · Bardia® Foley Catheter, Silicone Coated, 30cc Balloon · Class 2

All models of Bardia Foley Catheter Silicone Coated (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00801741095221), received between May 10, 2017 and Jun 26, 2024 — classified as 2 injuries, 17 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel5
  • Adverse Event Without Identified Device or Use Problem2
  • Failure to Infuse2
  • Material Fragmentation2
  • Partial Blockage2
  • Patient-Device Incompatibility2
  • Blocked Connection1
  • Deflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient6
  • No Known Impact Or Consequence To Patient4
  • No Information2
  • Bruise/Contusion1
  • Hemorrhage/Blood Loss/Bleeding1
  • Irritation1
  • Pressure Sore/Ulcer1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2024Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT THE CATHETER BALLOON WAS LEAKING. EMERGENCY ROOM LIGHT WAS ON AND IT LEAKED AT NIGHT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19613751 ↗
  • MalfunctionJun 24, 2024Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT THE STERILE WATER COULD NOT BE DRAINED FROM THE FOLEY CATHETER BALLOON DURING A PRETEST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19594616 ↗
  • MalfunctionMay 26, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING CHEMO MEDICATION IT WAS SEEN POWER PICC CATHETER 5FR DL LUMEN CAP IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17009560 ↗
  • MalfunctionJun 2, 2022Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT THE FOLEY CATHETER SLIPPED OFF AND THE NURSE HAD CHECKED FOR ANY DAMAGE IN THE BALLOON. AND THE NURSE HAD CHANGED A NEW CATHETER TO THE PATIENT. PER FOLLOW UP INFORMATION RECEIVED ON 19MAY2022, IT WAS STATED THAT THE CATHETER SLIPPED OFF DUE TO THE BALLOON BURST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14579761 ↗
  • MalfunctionMay 10, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE FOLEY CATHETER BUSTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14347205 ↗
  • MalfunctionSep 23, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT SOMETIMES THE FOLEY CATHETERS DID NOT LAST AS LONG AS OTHERS INCLUDING THE LAST ONE THAT PATIENT HAD USED AND NO SPECIFICS WERE GIVEN. PER FOLLOW UP VIA PHONE ON 01SEP2021, CUSTOMER STATED THAT THE ONLY ISSUE WITH THE FOLEY CATHETERS WAS THAT OCCASIONALLY LEAKED AROUND WHERE THEY INSERTED THE CATHETER AND HAVE TO CHANGE THE CATHETER PREMATURELY, BUT CUSTOMER SAID IT MIGHT BE DUE TO THEIR BODY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12519094 ↗
  • MalfunctionNov 26, 2020Partial Blockage; Partial Blockage

    IT WAS REPORTED THAT THE LAST FOLEY CATHETER GOT CLOGGED UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10908560 ↗
  • MalfunctionNov 18, 2020Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT WHEN ASKED ABOUT THE CHALLENGES THE HCP FACED WITH THE DEVICE, THE HCP MENTIONED "BALLONNET QUI A EXPLOSÉ SANS DÉPASSER LE NOMBRE LIMITE DE GONFLEMENT" MEANING "BALLOON THAT EXPLODED WITHOUT EXCEEDING THE LIMIT NUMBER OF INFLATION".

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10863915 ↗
  • InjuryFeb 21, 2020Patient-Device Incompatibility; Adverse Event Without Identified Device or Use Problem; Patient-Device Incompatibility

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED BLEEDING FOR 2 OR 3 DAYS AFTER USING THE LATEX FOLEY CATHETER. PATIENT'S WIFE STATED THAT THE PATIENT EXPERIENCED BLEEDING AFTER USING THE LATEX CATHETER, WHICH RESULTED IN SWELLING OF PENIS/TISSUE, SORES AND BRUISES. PATIENT'S WIFE NOTICED THAT WHEN HER HUSBAND USED THE SILICONE COATED CATHETER PROVIDED BY THE HOSPITAL, HE DID NOT EXPERIENCE THE ISSUES. PATIENT WAS FOLLOWING UP WITH HIS DOCTOR. PER FOLLOW UP ON 19FEB2020, PATIENT WAS ADMITTED INTO THE HOSPITAL OVER THE WEEKEND REGARDING THE IRRITATION AND BLEEDING ( BLOOD CLOTS) DURING URINATION. THE PRIMARY CARE DOCTOR SWITCHED THE PATIENTS CATHETERS TO SILICONE DUE TO THE LATEX IRRITATION. PATIENT WAS STILL EXPERIENCING A LITTLE IRRITATION AND BLEEDING. PATIENT WAS CURRENTLY ON ANTIBIOTICS AND HAD BEEN RELEASED FROM THE HOSPITAL. ADDITIONAL INFORMATION REQUESTED REGARDING THE MEATUS TEAR A YEAR AGO, STATED THAT THE PERVIOUS CATHETER USED ON PATIENT WAS NOT A BD PRODUCT.

    Patient problems: Bruise/Contusion; Hemorrhage/Blood Loss/Bleeding; Irritation; Pressure Sore/Ulcer; Swelling

    MAUDE report 9740191 ↗
  • MalfunctionNov 26, 2019Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT THE BALLOON POPPED WHILE INFLATING THE CATHETER. NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9378555 ↗
  • MalfunctionNov 26, 2019Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT THE BALLOON POPPED WHILE INFLATING THE CATHETER. NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9378547 ↗
  • MalfunctionNov 26, 2019Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT THE BALLOON POPPED WHILE INFLATING THE CATHETER. NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9374995 ↗
  • MalfunctionAug 30, 2019Inaccurate Flow Rate; Inaccurate Flow Rate

    IT WAS REPORTED THAT THE CATHETER WAS NOT DRAINING DIRECTLY INTO THE DRAIN BAG AND AS A RESULT ALLOWED FOR URINE LEAKAGE. NO MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8952227 ↗
  • InjuryFeb 15, 2019Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT THE CATHETERS WERE "BROWN" COMPARED TO THE OTHER CATHETERS THE PATIENT RECEIVED FROM ANOTHER COMPANY. THE PATIENT BELIEVES THE COLORING ON THE CATHETERS CAUSED A UTI. THE PATIENT WAS SEEN BY DOCTOR WHO PRESCRIBED ANTIBIOTICS.

    Patient problems: Urinary Tract Infection

    MAUDE report 8342684 ↗
  • MalfunctionDec 18, 2018Partial Blockage; Partial Blockage; Blocked Connection

    IT WAS REPORTED THAT SOME OF THE PATIENT'S FOLEY CATHETERS KEPT GETTING PLUGGED UP. ADDITIONAL INFORMATION WAS LATER RECEIVED THAT THE DEVICE WAS IN ABOUT 10 HOURS BEFORE THE CATHETER BEGAN TO CLOG. THE PATIENT INDICATED THAT THE CATHETER WAS BEING CLOGGED BY SEDIMENT FROM THE URINE. THE PATIENT HAD AN UPCOMING APPOINTMENT WITH HIS DOCTOR.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8173358 ↗
  • MalfunctionAug 2, 2017No Flow

    IT WAS REPORTED THAT THE CATHETER STOPPED DRAINING.

    Patient problems: No Information

    MAUDE report 6761925 ↗
  • MalfunctionAug 1, 2017Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, UPON INSERTION OF THE FOLEY CATHETER, NO URINE WOULD DRAIN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6758500 ↗
  • MalfunctionJul 17, 2017Failure to Infuse

    IT WAS REPORTED THAT THE CATHETER FAILED TO INFUSE WHEN THE PATIENT ATTEMPTED TO INFLATE THE BALLOON AFTER INSERTION. A NEW CATHETER WAS OBTAINED AND USED WITHOUT IMPACT TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6720266 ↗
  • MalfunctionMay 10, 2017Failure to Infuse

    IT WAS REPORTED THAT THE CATHETER FAILED TO INFUSE.

    Patient problems: No Information

    MAUDE report 6556889 ↗

Other CATHETER, UROLOGICAL devices

All CATHETER, UROLOGICAL devices →