Bard® DigniShield® Stool Management System with ENFit™ ENSMS002Recalls & reported problems
C. R. Bard, Inc. · Bard® DigniShield® Stool Management System with ENFit™ · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Bard DigniShield stool management system is a rectal tube used to divert stool in bedridden patients. The 44 reports mostly describe leaking of fluid or stool around or through the device, the balloon failing to hold it in place, and the material splitting, tearing or breaking apart. Seven reports involve injury, and no deaths are reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00801741113147), received between Sep 20, 2019 and Aug 3, 2026 — classified as 7 injuries, 36 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak11
- Failure to Capture8
- Patient Device Interaction Problem6
- Material Split, Cut or Torn4
- Material Puncture/Hole3
- Adverse Event Without Identified Device or Use Problem2
- Biocompatibility2
- Inadequacy of Device Shape and/or Size2
Patient problems
- No Clinical Signs, Symptoms or Conditions34
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
- Erosion3
- Necrosis3
- Skin Tears3
- Pressure Sore/Ulcer2
- No Consequences Or Impact To Patient1
- Unspecified Infection1
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 3, 2026Material Puncture/Hole
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50110042 ↗MalfunctionJul 27, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50026503 ↗MalfunctionFeb 19, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THE CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY STATED PHYSICIAN INDICATED "DEATH". PRIOR TO DEVICE PLACEMENT AND INDICATED LEAKAGE OF STOOL AROUND THE DEVICE AFTER DEVICE PLACEMENT AND INDICATED THE LEAKAGE OF STOOL AROUND THE DEVICE TO BE DISEASE OR OTHER RELATED AND STATED "DEATH". ADDITIONALLY, THE RESPONDENT INDICATED HOSPICE WHEN ASKED WHAT INDICATION THE DEVICE WAS USED FOR IN RELATION TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24389364 ↗MalfunctionFeb 18, 2026Fluid/Blood Leak
CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY STATED PHYSICIAN INDICATED LEAKAGE OF STOOL AROUND THE DEVICE AFTER PLACEMENT OF THE DEVICE AND INDICATED THE COMPLICATION TO BE DISEASE OR OTHER RELATED AND STATED CRITICAL ILLNESS IN RELATION TO THE DEVICE ENFIT DIGNISHIELD STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24378859 ↗MalfunctionFeb 18, 2026Fitting Problem
IT WAS REPORTED IN AN ONLINE SURVEY BY THE CRITICAL CARE NURSES THAT PHYSICIAN INDICATED TISSUE INJURY (NECROSIS, TEARS, EROSION, PERFORATION, BLEEDING) AFTER PLACEMENT OF THE DEVICE AND INDICATED THE COMPLICATION TO BE DISEASE OR OTHER RELATED AND STATED BEDRIDDEN. PATIENT OBESE BAG WOULD NOT SECURE WELL IN RELATION TO THE DEVICE ENFIT DIGNISHIELD STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002. NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378851 ↗MalfunctionFeb 18, 2026Positioning Failure
IT WAS REPORTED BY THE CRITICAL CARE NURSES IN AN ONLINE SURVEY THAT PHYSICIAN INDICATED LEAKAGE OF STOOL AROUND THE DEVICE AFTER PLACEMENT OF THE DEVICE AND INDICATED THE COMPLICATION TO BE DISEASE OR OTHER RELATED AND STATED POOR DEVICE POSITIONING. ADDITIONALLY, THE RESPONDENT INDICATED THE PATIENT HAD AN INFECTION, BOWEL OBSTRUCTION, AND ABDOMINAL DISTENSION PRIOR TO DEVICE PLACEMENT IN RELATION TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378848 ↗InjuryFeb 18, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT THE CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY STATED PHYSICIAN STATED THAT THE PATIENT EXPERIENCED AN ADVERSE EVENT DIRECTLY ATTRIBUTABLE TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002, AND THE PATIENT EXPERIENCED TISSUE INJURY(NECROSIS, TEARS, EROSION, PERFORATION, BLEEDING) AND IT DID NOT RESULT IN PATIENT INJURY REQUIRING MEDICAL OR SURGICAL INTERVENTION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Erosion; Necrosis; Skin Tears
MAUDE report 24378847 ↗InjuryFeb 18, 2026Patient Device Interaction Problem
IT WAS REPORTED IN AN ONLINE SURVEY BY THE CRITICAL CARE NURSES THAT PHYSICIAN INDICATED TISSUE INJURY (NECROSIS, TEARS, EROSION, PERFORATION, BLEEDING) AFTER PLACEMENT OF THE DEVICE AND INDICATED THE COMPLICATION TO BE DISEASE OR OTHER RELATED AND STATED BEDRIDDEN. PATIENT OBESE BAG WOULD NOT SECURE WELL IN RELATION TO THE DEVICE ENFIT DIGNISHIELD STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002. NO MEDICAL INTERVENTION WAS REPORTED.
Patient problems: Erosion; Necrosis; Skin Tears
MAUDE report 24378475 ↗MalfunctionNov 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE BARD RECTAL TUBE MALFUNCTION: TWO OF THEM THE IRRIGATION PORT BROKE AND STOOL LEAKED EVERYWHERE. THE LAST ONE ACTUALLY HAD A SMALL TEAR WHERE THE TUBING COMES OUT OF THE RECTUM, ALONG THE SEAM RESULTING IN LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23522784 ↗MalfunctionNov 11, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE BARD RECTAL TUBE MALFUNCTION: TWO OF THEM THE IRRIGATION PORT BROKE AND STOOL LEAKED EVERYWHERE. THE LAST ONE ACTUALLY HAD A SMALL TEAR WHERE THE TUBING COMES OUT OF THE RECTUM, ALONG THE SEAM RESULTING IN LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23522780 ↗MalfunctionNov 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE BARD RECTAL TUBE MALFUNCTION: TWO OF THEM THE IRRIGATION PORT BROKE AND STOOL LEAKED EVERYWHERE. THE LAST ONE ACTUALLY HAD A SMALL TEAR WHERE THE TUBING COMES OUT OF THE RECTUM, ALONG THE SEAM RESULTING IN LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23522770 ↗MalfunctionNov 11, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE BARD RECTAL TUBE MALFUNCTION: TWO OF THEM THE IRRIGATION PORT BROKE AND STOOL LEAKED EVERYWHERE. THE LAST ONE ACTUALLY HAD A SMALL TEAR WHERE THE TUBING COMES OUT OF THE RECTUM, ALONG THE SEAM RESULTING IN LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23522767 ↗MalfunctionJun 4, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE CRITICAL CARE NURSE STATED THAT IN THE ONLINE SURVEY A PHYSICIAN STATED THAT DIGNISHILED WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR PATIENT IN RELATION TO THE DEVICE ENFIT DIGNISHIELD STOOL MANAGEMENT SYSTEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22140354 ↗MalfunctionJun 3, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE PRODUCT WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR PATIENT BECAUSE IN RELATION TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22133977 ↗MalfunctionJun 3, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE PRODUCT WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR PATIENT BECAUSE IN RELATION TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22133972 ↗MalfunctionJun 2, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE PRODUCT WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR PATIENT BECAUSE THE PATIENT WAS VERY AGITATED AND RESTLESS IN BED AND PRODUCT CAME OUT A COUPLE OF TIMES IN RELATION TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22124542 ↗MalfunctionMay 30, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT DIGNISHIELD WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR PATIENT " BECAUSE " NOT SURE IF" IN RELATION TO THE DEVICE "ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM" WITH PRODUCT CODE " ENSMS002".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22114158 ↗MalfunctionMay 30, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE PRODUCT WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR PATIENT BECAUSE THE PATIENT HAD SEVERE LEAKAGE AROUND THE RECTUM AND IT DID NOT FULLY DIVERT STOOL INTO THE COLLECTION BAG. THE PATIENT WAS INTUBATED AND VERY SICK AND WAS DIFFICULT TO CLEAN UP FREQUENTLY IN RELATION TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22114145 ↗MalfunctionMay 30, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE CRITICAL CARE NURSE STATED THAT IN THE ONLINE SURVEY A PHYSICIAN STATED THAT DIGNISHILED WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR PATIENT BECAUSE IN RELATION TO THE DEVICE "ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM" WITH PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22114110 ↗MalfunctionMay 30, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT CRITICAL CARE NURSE REPORTED THAT IN THE ONLINE SURVEY A PHYSICIAN STATED THAT "NO, THE PRODUCT WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR YOUR PATIENT" IN RELATION TO THE DEVICE "ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM" WITH PRODUCT CODE "ENSMS002".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22114092 ↗MalfunctionMay 30, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE PRODUCT WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR PATIENT IN RELATION TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE ENSMS002.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22114088 ↗InjuryMay 30, 2025Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT THE CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT THE PATIENT EXPERIENCED AN ADVERSE EVENT DIRECTLY ATTRIBUTABLE TO THE DEVICE ENFIT DIGNISHIELD® STOOL MANAGEMENT SYSTEM WITH PRODUCT CODE: ENSMS002, AND THE PATIENT EXPERIENCED TISSUE INJURY (NECROSIS, TEARS, EROSION, PERFORATION, BLEEDING) AND IT DID NOT RESULT IN PATIENT INJURY REQUIRING MEDICAL OR SURGICAL INTERVENTION.
Patient problems: Erosion; Necrosis; Skin Tears
MAUDE report 22113611 ↗MalfunctionMay 29, 2025Partial Blockage; Partial Blockage
IT WAS REPORTED THAT THE CRITICAL CARE NURSE STATED THAT IN THE ONLINE SURVEY A PHYSICIAN STATED THAT THEY WERE NOT ABLE TO SUCCESSFULLY COLLECT A SAMPLE VIA THE SAMPLE PORT WHEN USING THE DIGNISHIELD DEVICE ON THE PATIENT BECAUSE BLOCKED PORT IN RELATION TO THE DEVICE ENFIT DIGNISHIELD STOOL MANAGEMENT SYSTEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22107222 ↗MalfunctionMay 29, 2025Inability to Irrigate; Inability to Irrigate
IT WAS REPORTED THAT THE CRITICAL CARE NURSES REPORTED IN AN ONLINE SURVEY THAT CUSTOMER WAS NOT ABLE TO SUCCESSFULLY DELIVER MEDICATION TO THE PATIENT WHEN USING THE DIGNISHIELD DEVICE BECAUSE IT WAS NOT CLEARING WITH THE C DIFF IN RELATION TO THE DEVICE ENFIT DIGNISHIELD STOOL MANAGEMENT SYSTEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22105999 ↗MalfunctionMay 29, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE DIGNISHIELD WAS NOT SUCCESSFULLY ABLE TO DIVERT AND COLLECT LIQUID OR SEMI-LIQUID STOOL FOR THE SPECIFIED DURATION OF USE FOR YOUR PATIENT BECAUSE UNSURE ABOUT WHY, BUT STOOL WAS LEAKING SIGNIFICANTLY AROUND THE TUBE. AFTER TROUBLESHOOTING UNSUCCESSFULLY, THEY ENDED UP REPLACING THE TUBE IN RELATION TO THE DEVICE ENFIT DIGNISHIELD STOOL MANAGEMENT SYSTEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22105992 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page