LIFESTENT® 5F051503CSRecalls & reported problems
Bard Peripheral Vascular, Inc. · LIFESTENT® 5F Vascular Stent System, 5 mm x 150 mm (135 cm delivery catheter) · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741120664), received between Jan 9, 2020 and May 13, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation5
- Break3
- Detachment of Device or Device Component2
- Difficult to Remove2
- Fracture2
- Misfire2
- Device-Device Incompatibility1
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
- Stenosis1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2026Break; Detachment of Device or Device Component; Difficult to Advance
ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING A LIFESTENT DEVICE. IT WAS REPORTED THAT THE NOSE CONE OF THE STENT DETACHED AFTER DEPLOYMENT. THE CATHETER FRAGMENTS WERE SUBSEQUENTLY RETRIEVED FROM THE PATIENT¿S BODY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25157848 ↗MalfunctionNov 3, 2025Material Deformation; Material Deformation
ON (B)(6) 2025, A PATIENT UNDERWENT A STENT PLACEMENT WITH LIFESTENT VASCULAR STENT. DURING THE PROCEDURE WHEN IT WAS PLACED IN THE PATIENT IT WAS FOUND THAT THE LENGTH OF THE STENT WAS NOT ACCURATE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23447441 ↗InjuryNov 19, 2024Fracture; Obstruction of Flow; Fracture; Obstruction of Flow
IT WAS REPORTED THAT APPROXIMATELY ONE YEAR, NINE MONTHS AND SIXTEEN DAYS STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY FRACTURED AND PATIENT ALLEGEDLY EXPERIENCED RESTENOSIS. HOWEVER, THE CURRENT STATUS OF PATIENT IS UNKNOWN.
Patient problems: Stenosis
MAUDE report 20721700 ↗MalfunctionApr 11, 2024Fracture; Difficult to Remove; Misfire; Material Deformation; Fracture; Difficult to Remove; Misfire; Material Deformation
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE RIGHT POPLITEAL ARTERY VIA THE LEFT COMMON FEMORAL ARTERY, THE STENT ALLEGEDLY DIDN'T FULLY DEPLOY AS ONLY ONE TO TWO CENTIMETERS OF THE PROXIMAL END OF THE STENT GOT DEPLOYED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19085880 ↗MalfunctionMar 25, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, WHILE REMOVING THE CATHETER, THE CATHETER ALLEGEDLY GOT SNAPPED IN HALF INSIDE THE PATIENT. IT WAS FURTHER REPORTED THAT THE PIECE THAT HOLDS THE CATHETER TO THE STENT WAS ALLEGEDLY BROKEN. REPORTEDLY, THE CATHETER WAS PULLED OUT WITH THE SHEATH. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18966715 ↗MalfunctionAug 31, 2023Failure to Advance; Material Deformation; Failure to Advance; Material Deformation
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY WOULD NOT GO THROUGH THE INTRODUCER SHEATH FROM A PEDAL ARTERY ACCESS SITE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17659090 ↗MalfunctionJul 27, 2022Material Deformation; Positioning Problem; Material Deformation; Positioning Problem
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY HAD A POSITIONING PROBLEM. IT WAS FURTHER REPORTED THAT CATHETER ALLEGEDLY BUCKLED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15110782 ↗MalfunctionApr 13, 2022Break; Difficult to Remove; Device-Device Incompatibility; Break; Difficult to Remove; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE GUIDEWIRE WAS ALLEGEDLY STUCK IN THE DELIVERY SYSTEM AND THEN GUIDEWIRE WAS ALLEGEDLY NOT ABLE TO BE REMOVED, STENT COULDN'T DEPLOY. IT WAS FURTHER REPORTED THAT THE DELIVERY SYSTEM HAD THE THUMB WHEEL WAS ALLEGEDLY BROKEN. REPORTEDLY, THE HEALTH CARE PROVIDED HAD TO REMOVE THE GUIDEWIRE AND THE DELIVERY SYSTEM AS A SINGLE UNIT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14098056 ↗MalfunctionJan 9, 2020Misfire; Material Deformation; Misfire; Material Deformation
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 5F051503CS VASCULAR STENT ALLEGEDLY EXPERIENCED MISFIRE AND MATERIAL DEFORMATION. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THERE WERE NO CONSEQUENCES TO THE ONE PATIENT INVOLVED. THE PATIENT WAS REPORTED AS A (B)(6) YEAR OLD MALE WEIGHING (B)(6) POUNDS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9570157 ↗
Other STENT, SUPERFICIAL FEMORAL ARTERY devices
- EverFlex™ EVD35-06-150-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 death110 reports
- SUPERA S-60-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY108 reports
- SUPERA S-55-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY88 reports
- LifeStent® Solo Vascular Stent EX062003CLBard Peripheral Vascular, Inc. · STENT, SUPERFICIAL FEMORAL ARTERY80 reports
- BioMimics 3D Vascular Stent System 144700-12Veryan Medical Limited · Stent, Superficial Femoral Artery74 reports
- SUPERA S-50-120-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY55 reports
- EverFlex™ EVD35-07-120-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
- EverFlex™ EVD35-07-060-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page