Skip to content

Site-Rite 9770001Recalls & reported problems

Bard Access Systems, Inc. · SITE-RITE 6 PROBE · Class 2

All models of Site-Rite (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741125744), received between Feb 22, 2018 and May 21, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image3
  • No Display/Image2
  • Crack1
  • Device Operates Differently Than Expected1
  • Display or Visual Feedback Problem1
  • Image Display Error/Artifact1
  • Signal Artifact/Noise1

Patient problems

  • No Known Impact Or Consequence To Patient7
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 21, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THE UNIT IS PRODUCING GRAINY IMAGES/DISPLAY. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19360998 ↗
  • MalfunctionJan 12, 2019No Display/Image

    IT WAS REPORTED, THAT THE IMAGE FROM THE PROBE IS DIFFICULT TO SEE WHEN PLACING A PICC.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8242929 ↗
  • MalfunctionJan 12, 2019No Display/Image

    IT WAS REPORTED, THAT THE UNIT'S PROBE SHOWS A DARK SHADOW ON THE ULTRASOUND SCREEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8242926 ↗
  • MalfunctionDec 7, 2018Poor Quality Image

    BIOMED REPORTED SYSTEM HAS POOR ULTRASOUND IMAGE QUALITY.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8143795 ↗
  • MalfunctionNov 20, 2018Image Display Error/Artifact

    PER BIOMED: THE PROBE IS CASTING SHADOWS IN THE IMAGE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8092456 ↗
  • MalfunctionSep 7, 2018Crack; Poor Quality Image

    BIOMED REPORTED THAT SITE RITE 6 PROBE (B)(4) DISPLAYS A POOR QUALITY IMAGE. PROBE IS PHYSICALLY CRACKED AND DAMAGED AND BIOMED FEELS THAT IS CAUSING THE POOR IMAGE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7858221 ↗
  • MalfunctionMar 14, 2018Signal Artifact/Noise; Display or Visual Feedback Problem

    PER BIOMED: SITE RITE 6 PROBE 10145405 IS CAUSING SOME BLANK SPOTS, ARTIFACT IN THE IMAGE. POSSIBLE DAMAGED CRYSTALS IN THE PROBE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7340080 ↗
  • MalfunctionFeb 22, 2018Device Operates Differently Than Expected

    IT WAS REPORTED BY THE FACILITY THAT THERE IS A "DROP-OUT" ON THE IMAGE WITH THE SITE RITE 6 PROBE (SN:(B)(4)) WITH BLACK LINES DOWN THE SCREEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7289846 ↗

Other Transducer, ultrasonic, diagnostic devices