Skip to content

Arctic Sun® Temperature Management System, Transmission Interface Module Kit 76000Recalls & reported problems

Medivance, Inc. · Arctic Sun® Temperature Management System, Transmission Interface Module Kit · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741127762), received between Sep 16, 2022 and Sep 16, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Computer Software Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2022Computer Software Problem; Computer Software Problem

    IT WAS REPORTED THAT THE BIOMED WAS WITH THE INTEGRATION SYSTEM TEAM AT THE HOSPITAL AND THIS ACCOUNT WAS USING THE TRANSDUCER INTERFACE MODULE TO OUTPUT DATA TO EPIC, BIOMED FOUND THAT IF THEY START A THERAPY WITH A TARGET TEMPERATURE AND CHANGE THAT TARGET TEMPERATURE IN MID THERAPY THE DEVICE RESPONDS NORMAL BUT THE DATA COMING OUT OF THE TRANSDUCER INTERFACE MODULE DID NOT UPDATE AND CONTINUES TO SHOW THE ORIGINAL TARGET TEMPERATURE. BIOMED TRIED A SECOND DEVICE AND IT DID THE SAME THING , THEY ARE REMOVING THIS VALUE FROM THE DATA THEY COLLECT BUT WOULD LIKE TO GET THE ISSUE CORRECTED SO THEY COULD CONTINUE TO COLLECT THE TARGET TEMPERATURE DATA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15437527 ↗

Other System, thermal regulating devices

All System, thermal regulating devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page