Sleek OTW PTA Dilatation Catheter 426-2022XRecalls & reported problems
Clearstream Technologies Limited · Sleek OTW PTA Dilatation Catheter 2.0mm x 220mm balloon, 150cm shaft · Class 2
All models of Sleek OTW PTA Dilatation Catheter (24)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741133626), received between Feb 9, 2018 and Dec 8, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Material Rupture3
- Stretched2
- Break1
- Device-Device Incompatibility1
- Difficult to Open or Remove Packaging Material1
- Failure to Advance1
- Retraction Problem1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2025Break; Stretched; Difficult to Open or Remove Packaging Material; Break; Stretched; Difficult to Open or Remove Packaging Material
ON (B)(6) 2025, A PATIENT UNDERWENT A PROCEDURE USING THE SLEEK OTW PTA CATHETER. UPON OPENING, THE SLEEVE DID NOT SEPARATE FROM THE BALLOON. HE TRIED TO PULL IT, WHICH RESULTED IN DAMAGE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23738527 ↗MalfunctionOct 21, 2020Material Rupture; Stretched; Detachment of Device or Device Component; Material Rupture; Stretched; Detachment of Device or Device Component
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 426-2022X PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED RUPTURE AND DETACHMENT. THIS REPORT WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED A PATIENT WITH NO CONSEQUENCES. THE PATIENT'S AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10715048 ↗MalfunctionJul 30, 2020Retraction Problem; Device-Device Incompatibility; Retraction Problem; Device-Device Incompatibility
IT WAS REPORTED THAT DURING THE PROCEDURE, THE DEVICE ALLEGEDLY DRAGGED AND STUCK WHEN TRYING TO REMOVE THE CATHETER. IT WAS FURTHER REPORTED THAT THE DOCTOR LOST WIRE PLACEMENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10342611 ↗MalfunctionJul 17, 2018Failure to Advance; Detachment of Device or Device Component; Failure to Advance; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PTA BALLOON CATHETER DELIVERY SYSTEM WAS ALLEGEDLY UNABLE TO CROSS THE LESION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7697140 ↗MalfunctionMar 27, 2018Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PTA BALLOON ALLEGEDLY RUPTURED DURING THE SECOND INFLATION. REPORTEDLY, THE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT INCIDENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7374109 ↗MalfunctionFeb 9, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT THE BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7260357 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page