Skip to content

LUTONIX ® 035 Drug Coated Balloon PTA Catheter LX3575680VRecalls & reported problems

Lutonix, Inc. · Lutonix® AV Drug Coated Balloon PTA Catheter 6x80x75 · Class 3

All models of LUTONIX 035 Drug Coated Balloon PTA Catheter (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741135408), received between Feb 2, 2021 and May 9, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Material Rupture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Restenosis2
  • Stenosis1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL THAT ONE MONTH POST INDEX PROCEDURE USING A DRUG-COATED BALLOON CATHETER IN THE CEPHALIC VEIN, THE SUBJECT HAD AN ADVERSE EVENT OF RECURRENT STENOSIS IN THE DRAINAGE VEIN AND JUXTA-ANASTOMOSIS REGION AND REQUIRED AN ARTERIOVENOUS ACCESS CIRCUIT REINTERVENTION. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS IN THE UPPER ARM WITH FIFTY FOUR PERCENT INITIAL STENOSIS AND ZERO PERCENT FINAL RESIDUAL STENOSIS. REPORTEDLY, STANDARD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY WAS USED FOR TREATMENT. THE RE-INTERVENTION WAS SUCCESSFUL, AND THE OUTCOME OF THE ADVERSE EVENT WAS RESOLVED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Restenosis; Stenosis; Thrombosis/Thrombus

    MAUDE report 19275999 ↗
  • InjuryOct 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL THAT TWO MONTHS AND EIGHT DAYS POST AN INDEX PROCEDURE COMPLETION USING A DRUG-COATED BALLOON CATHETER IN THE CEPHALIC VEIN, THE SUBJECT DEVELOPED A SERIOUS ADVERSE EVENT OF ANASTOMOTIC STENOSIS ON THE CEPHALIC VEIN DUE TO DECREASED ACCESS BLOOD FLOW WHICH REQUIRED AN ARTERIOVENOUS ACCESS CIRCUIT REINTERVENTION. REPORTEDLY, STANDARD PERCUTANEOUS TRANSLUMINAL BALLOON ANGIOPLASTY AND DRUG COATED BALLOON CATHETER PROCEDURES WERE PERFORMED TO TREAT DECREASED ACCESS BLOOD FLOW. THE TREATMENT RE-INTERVENTION WAS SUCCESSFUL, NO NEW ACCESS CREATED, AND THE OUTCOME OF THE SERIOUS ADVERSE EVENT WAS RESOLVED. IT WAS FURTHER REPORTED THAT FIVE MONTHS AND TWENTY-TWO DAYS POST AN INDEX PROCEDURE COMPLETION, THE SUBJECT DEVELOPED A SERIOUS ADVERSE EVENT OF TWO STENOSIS ONE IN THE CEPHALIC VEIN AND THE OTHER ON THE SUBCLAVIAN VEIN DUE TO DECREASED ACCESS BLOOD FLOW WHICH REQUIRED AN ARTERIOVENOUS ACCESS CIRCUIT REINTERVENTION. REPORTEDLY, STANDARD PERCUTANEOUS TRANSLUMINAL BALLOON ANGIOPLASTY AND DRUG COATED BALLOON CATHETER PROCEDURES WERE PERFORMED TO TREAT DECREASED ACCESS BLOOD FLOW. THE TREATMENT RE-INTERVENTION WAS SUCCESSFUL, NO NEW AC

    Patient problems: Restenosis

    MAUDE report 17894225 ↗
  • MalfunctionFeb 18, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE VIA CEPHALIC VEIN FOR TREATING STENOSIS AND HIGH VENOUS PRESSURE, THE BALLOON ALLEGEDLY RUPTURED AT NOMINAL PRESSURE. IT WAS FURTHER REPORTED THAT THE BALLOON WAS RETRIEVED WITHOUT ANY ISSUE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13556386 ↗
  • MalfunctionFeb 2, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THROUGH THE RIGHT UPPER ARM, THE DRUG COATED DEVICE ALLEGEDLY RUPTURED AT 12 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11264466 ↗

Other Drug-coated peripheral transluminal angioplasty catheter devices

All Drug-coated peripheral transluminal angioplasty catheter devices →

Sources and updates