Skip to content

PowerPort ClearVUE, ChronoFlex 1606382Recalls & reported problems

Bard Access Systems, Inc. · PowerPort ClearVUE isp, 6F ChronoFlex Catheter, Intermediate NAUTILUS DELTA Kit · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741136719), received between Jul 13, 2018 and Jul 13, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unable to Obtain Readings” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Unable to Obtain Readings1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 13, 2018Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED DURING INSERTION OF THE ISP CLEARVUE NAUTILUS PORT. THE METAL STEM CONNECTOR OF THE LUER LOCK TO THE PORT CATHETER WAS SCRUBBED USING A SOAKED 4X4 WITH STERILE ISOPROPYL ALCOHOL. THE ALLIGATOR CLIP WAS ATTACHED AND SCORING METHOD WAS PERFORMED, "SCORING" THE METAL STEM TO IMPROVE CONDUCTIVITY. DESPITE THESE EFFORTS NO DISCERNIBLE WAVE FORM WAS ESTABLISHED, "FLATLINE". FLUOROSCOPY WAS OBTAINED TO VERIFY THE PLACEMENT OF THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7688181 ↗

Other PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices

All PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page